Accommodative intraocular lens (IOL) tilting, the so-called Z syndrome, is a rare complication of Crystalens (Bausch and Lomb) implantation. We report a significant IOL tilting and subsequent high lenticular astigmatism due to posterior capsular fibrosis 2 months after uncomplicated cataract surgery and Crystalens (AT50AO) implantation. The attempt to correct IOL position with neodymium-yttrium-aluminum garnet laser was unsuccessful, and Crystalens exchange with in-the-sulcus, three-piece monofocal IOL was performed. Accommodative IOL tilting could occur early after the surgery. Laser capsulotomy may be the first intervention to restore IOL position, but patients with a higher amount of lenticular astigmatism may require surgical intervention and IOL exchange.
引用
收藏
页码:46 / 48
页数:3
相关论文
共 50 条
  • [21] Evaluation of an accommodative Intraocular lens in human and rabbit eyes
    Pandey, SK
    Werner, L
    Apple, DJ
    Izak, AM
    Trivedi, RH
    McLeod, SD
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2002, 43 : U88 - U88
  • [22] Intraocular lens exchange due to incorrect lens power
    Jin, George J. C.
    Crandall, Alan S.
    Jones, Jason J.
    OPHTHALMOLOGY, 2007, 114 (03) : 417 - 424
  • [23] Refractive Lens Exchange with Foldable Toric Intraocular Lens
    Ruiz-Mesa, Ramon
    Carrasco-Sanchez, Daniel
    Diaz-Alvarez, Sara B.
    Angeles Ruiz-Mateos, M.
    Ferrer-Blasco, Teresa
    Montes-Mico, Robert
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2009, 147 (06) : 990 - 996
  • [24] Refractive lens exchange with the Array multifocal intraocular lens
    Packer, M
    Fine, IH
    Hoffman, RS
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (03): : 421 - 424
  • [25] Refractive lens exchange with an array multifocal intraocular lens
    Dick, HB
    Gross, S
    Tehrani, M
    Eisenmann, D
    Pfeiffer, N
    JOURNAL OF REFRACTIVE SURGERY, 2002, 18 (05) : 509 - 518
  • [26] Objective accommodative amplitude and dynamics with the 1CU accommodative intraocular lens
    Wolffsohn, JS
    Hunt, OA
    Naroo, S
    Gilmartin, B
    Shah, S
    Cunliffe, IA
    Benson, MT
    Mantry, S
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2006, 47 (03) : 1230 - 1235
  • [27] Optimized multielement accommodative intraocular lens with a four-freeform-surface Alvarez lens and a separate aspheric lens
    Wang, Liang-Yu
    Li, Bai-Cheng
    Sheng, Bin
    Xu, Bang-Lian
    Huang, Yuan-Shen
    Ni, Zheng-Ji
    Zhang, Da-Wei
    APPLIED OPTICS, 2019, 58 (27) : 7609 - 7614
  • [28] Management of intraoperative tilting of the scleral-fixated intraocular lens in classical aniridia
    Arvind, Hemamalini
    Hunyor, Alex
    McClellan, Kathy
    Billson, Francis
    Grigg, John
    Jamieson, Robyn
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2007, 91 (09) : 1247 - 1248
  • [29] Late intraocular lens exchange in dissatisfied patients with multifocal intraocular lens implantation
    Kuo, Yu-Wei
    Hou, Yu-Chih
    TAIWAN JOURNAL OF OPHTHALMOLOGY, 2022, 12 (01) : 109 - 112
  • [30] Multifocal intraocular lens exchange to monofocal for the management of neuroadaptation failure
    Al-Shymali, Olena
    Alio del Barrio, Jorge L.
    McAlinden, Colm
    Canto, Mario
    Primavera, Laura
    Alio, Jorge L.
    EYE AND VISION, 2022, 9 (01)