Patient-prosthesis mismatch does not affect survival following aortic valve replacement

被引:70
|
作者
Howell, Neil J. [1 ]
Keogh, Bruce E. [1 ]
Barnet, Vivien [1 ]
Bonser, Robert S. [1 ]
Graham, Timothy R. [1 ]
Rooney, Stephen J. [1 ]
Wilson, Ian C. [1 ]
Pagano, Domenico [1 ]
机构
[1] Univ Hosp NHS Fdn Trust, Dept Cardiothorac Surg, Birmingham, W Midlands, England
关键词
aortic valve disease; aortic valve replacement;
D O I
10.1016/j.ejcts.2006.03.046
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: Patient-prosthesis mismatch (PPM) has been reported to increase perioperative mortality and reduce postoperative survival in patients undergoing aortic valve replacement (AVR). We analysed the effect of PPM at values predicting severe mismatch on survival following AVR in our unit. Methods: Prospectively collected data on 1481 consecutive patients who had undergone AVR with or without coronary artery revascularisation between 1997 and 2005 were analysed. Projected in vitro valve effective orifice area (EOA) and geometric prosthesis internal orifice area (GOA) were evaluated and values were indexed to body surface area (cm(2) m(-2)). PPM was defined as EOAi < 0.6 and/or GOAi < 1.1. Long-term survival data were obtained from the National Institute of Statistics. Results: One thousand four hundred and eighteen patients were identified. 67/1418 (4.7%) patients had GOAi < 1.1; 122/1418 (8.6%) had EOAi < 0.6 and 38 (2.6%) patients exhibited both forms of mismatch. One thousand two hundred and sixty-seven patients (89%) demonstrated no mismatch (reference group). There were 75 in-hospital deaths (overall mortality 5.3%) with no significant difference between the mismatch and the reference groups. Survival data were available for up to 8 years (median 36 months, IQR 6-60 months). There were 160 late deaths (13/143 PPM group vs 147/1198 reference group). The 5-year survival estimate was similar for both groups (83% PPM group; 81% reference group; p = 0.47). Cox-hazard analysis identified advanced age as the only predictor of reduced survival (age > 80, RR 2.13, 95% CI 1.38-4.586, p = 0.004). Conclusions: Severe patient-prosthesis mismatch was predicted in 4-10% of patients undergoing AVR but this did not affect in-hospital mortality or mid-term survival. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:10 / 14
页数:5
相关论文
共 50 条
  • [11] Patient-prosthesis mismatch after aortic valve replacement in the elderly
    Ryomoto M.
    Mitsuno M.
    Yamamura M.
    Tanaka H.
    Kobayashi Y.
    Fukui S.
    Tsujiya N.
    Kajiyama T.
    Miyamoto Y.
    General Thoracic and Cardiovascular Surgery, 2008, 56 (7) : 330 - 334
  • [12] Patient-prosthesis mismatch and long-term outcome following aortic valve replacement
    Pingpoh, Clarence
    Khargi, Krishna
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2022, 61 (03) : 635 - 636
  • [13] Patient-prosthesis mismatch following mechanical mitral valve replacement
    Chan, Vincent
    Lam, Buu-Khanh
    Mesana, Thierry
    Ruel, Marc
    Hendry, Paul
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2007, 49 (09) : 314A - 314A
  • [14] Impact of Patient-Prosthesis Mismatch following Aortic Valve Replacement on Long-Term Survival and Quality of Life
    Hoffmann, Grischa
    Abraham-Westphal, Selam
    Attmann, Tim
    Frank, Derk
    Lutter, Georg
    Cremer, Jochen
    Petzina, Rainer
    THORACIC AND CARDIOVASCULAR SURGEON, 2020, 68 (02): : 124 - 130
  • [15] Patient-prosthesis mismatch after minimally invasive aortic valve replacement
    Filip, Grzegorz
    Litwinowicz, Radoslaw
    Kapelak, Boguslaw
    Bryndza, Magdalena
    Bartus, Magdalena
    Konstanty-Kalandyk, Janusz
    Ceranowicz, Piotr
    Brzezinski, Maciej
    Gafoor, Sameer
    Bartus, Krzysztof
    KARDIOLOGIA POLSKA, 2018, 76 (05) : 908 - 910
  • [16] Subvalvular stenosis at repeat aortic valve replacement for patient-prosthesis mismatch
    Girard, SE
    Miller, FA
    Edwards, WD
    Tazelaar, HD
    Mahoney, DW
    Malouf, YF
    Tajik, AJ
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2000, 35 (02) : 534A - 534A
  • [17] Patient-Prosthesis Mismatch and Strategies to Prevent It During Aortic Valve Replacement
    Apostolakis, Efstratios
    Baikoussis, Nikolaos G.
    Papakonstantinou, Nikolaos A.
    Goudevenos, John
    HELLENIC JOURNAL OF CARDIOLOGY, 2011, 52 (01) : 41 - 51
  • [18] Effect of patient-prosthesis mismatch on longterm survival with aortic valve replacement: Assessment to 15 years
    Jamieson, W.
    Pibarot, P.
    Dumesnil, J. G.
    Ye, J.
    Cheung, A.
    Germann, E.
    Chan, F.
    MacNab, J. S.
    Stanford, E. A.
    Bryson, L. A.
    Licthentstein, S. V.
    CANADIAN JOURNAL OF CARDIOLOGY, 2007, 23 : 83C - 83C
  • [19] Does a Minimally Invasive Approach Increase the Incidence of Patient-Prosthesis Mismatch in Aortic Valve Replacement?
    Bakir, Ihsan
    Casselman, Filip P.
    Onan, Burak
    Van Praet, Frank
    Vermeulen, Yvette
    Degrieck, Ivan
    JOURNAL OF HEART VALVE DISEASE, 2014, 23 (02): : 161 - 167
  • [20] Does patient-prosthesis mismatch have a negative impact on outcomes following mitral valve replacement?
    Aboul-Hassan, Sleiman Sebastian
    Stankowski, Tomasz
    Marczak, Jakub
    Cichon, Romuald
    INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2018, 26 (06) : 1027 - 1034