Experimental analysis on calibration of instrument broadening in a lidar system with Fabry-Perot etalon

被引:6
|
作者
Ma, Yong [1 ]
Li, Hao [1 ]
Yu, Yin [1 ]
Yao, Yuan [2 ]
Fang, Yu [1 ]
Zhou, Bo [1 ]
Huang, Jun [1 ]
Min, Jun [1 ]
Liang, Kun [1 ]
机构
[1] Huazhong Univ Sci & Technol, Elect & Informat Engn Dept, Wuhan 430074, Peoples R China
[2] Huazhong Univ Sci & Technol, Coll Publ Adm, Wuhan 430074, Peoples R China
基金
中国国家自然科学基金; 高等学校博士学科点专项科研基金;
关键词
Rayleigh Brillouin scattering; lidar; Fabry-Perot etalon; high resolution spectrum; F-P ETALON; BRILLOUIN-SCATTERING; BULK VISCOSITY; EDGE TECHNIQUE; SOUND SPEED; TEMPERATURE; ATTENUATION; DEPENDENCE; FREQUENCY; LIQUID;
D O I
10.1080/09500340.2013.868552
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
A Brillouin lidar system based on a Fabry-Perot (F-P) etalon and intensified charge coupled device (ICCD) is widely used in measuring ocean physical properties. In previous signal detecting methods, the resolution of the obtained Brillouin spectrum is the major limitation on measurement accuracy of the Brillouin spectrum. The effect due to the instrument of the F-P etalon is relatively minor and typically neglected. However, because methods for obtaining a high resolution spectrum have greatly improved the spectral resolution, the instrument broadening has become the major error source and should be taken into consideration in the measurement. Instrument broadening is a common topic and has been widely researched. Unfortunately, for applications involving a high resolution spectrum, no detailed study has analyzed the performance of the methods for calibrating the broadening with high resolution experimental data. In this paper, the capability and the time costs of methods for reducing the broadening effect are evaluated for a high resolution spectrum obtained from underwater experiments. We also discuss in detail the implementation of the methods. Experimental results show that the three methods considered can effectively calibrate the broadening effect and improve the measurement accuracy from tens of MHz to MHz magnitude. Considering the similar performance of the methods, one method that is sufficient for online measurement is recommended as the optimal method.
引用
收藏
页码:1968 / 1976
页数:9
相关论文
共 50 条
  • [1] Accuracy analysis of the Fabry-Perot etalon based Doppler wind lidar
    Sun, DS
    Zhong, ZQ
    Zhou, J
    Hu, HL
    Kobayashi, T
    OPTICAL REVIEW, 2005, 12 (05) : 409 - 414
  • [2] Analysis of the direct detection wind lidar with a dual Fabry-Perot etalon
    Chi, RL
    Sun, DS
    Zhong, ZQ
    Li, XM
    Zhou, J
    Hu, HL
    Wang, BX
    Xia, HY
    OPTICAL TECHNOLOGIES FOR ATMOSPHERIC, OCEAN, AND ENVIRONMENTAL STUDIES, PTS 1 AND 2, 2005, 5832 : 140 - 147
  • [3] Design and analysis of 532 nm Doppler wind lidar with Fabry-Perot etalon
    Wang G.
    Sun D.
    Du H.
    Kang J.
    Tang L.
    Shu Z.
    Hu D.
    Xu W.
    Dong J.
    Qiangjiguang Yu Lizishu/High Power Laser and Particle Beams, 2011, 23 (04): : 949 - 953
  • [4] Application and analysis of the dual Fabry-Perot etalon in a direct detection wind lidar
    Chi, Ru-Li
    Liu, Dong
    Zhong, Zhi-Qing
    Sun, Dong-Song
    Zhou, Jun
    Hu, Huan-Ling
    Qiangjiguang Yu Lizishu/High Power Laser and Particle Beams, 2007, 19 (02): : 202 - 206
  • [5] Design of a liquid crystal Fabry-Perot etalon and analysis of resonant mode splitting of a birefringent Fabry-Perot etalon
    Jiao, MX
    Qi, RL
    Liu, ZZ
    Xing, JH
    HIGH-POWER LASERS AND APPLICATIONS III, 2004, 5627 : 473 - 480
  • [6] THEORETICAL INSTRUMENT FUNCTION OF A FABRY-PEROT ETALON WITH FINITE APERTURE
    NEUHAUS, H
    NYLEN, P
    OPTICA ACTA, 1971, 18 (08): : 637 - &
  • [7] Modernization of a Fabry-Perot Etalon
    Efimkov, V. F.
    Zubarev, I. G.
    Mikhailov, S. I.
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 2018, 61 (01) : 114 - 115
  • [8] ON THE FINESSE OF THE FABRY-PEROT ETALON
    POPESCU, G
    REVUE ROUMAINE DE PHYSIQUE, 1981, 26 (07): : 719 - 723
  • [9] Fabry-Perot etalon filter
    Wang, QM
    Zhang, YM
    Liu, WH
    ICO20: OPTICAL INFORMATION PROCESSING, PTS 1 AND 2, 2006, 6027
  • [10] SPIE Proceedings: Concept for Calibration of OSIRIS with a Fabry-Perot Etalon
    Lopez, Ronald A.
    Von Raesfeld, Caroline
    Fitzgerald, Michael P.
    Do, Tuan
    GROUND-BASED AND AIRBORNE INSTRUMENTATION FOR ASTRONOMY IX, 2022, 12184