Refinement of MLC modeling improves commercial QA dosimetry system for SRS and SBRT patient-specific QA

被引:21
|
作者
Hillman, Yair [1 ]
Kim, Josh [2 ]
Chetty, Indrin [2 ]
Wen, Ning [2 ]
机构
[1] Karmanos Canc Inst McLaren Macomb, Dept Radiat Oncol, Mt Clemens, MI USA
[2] Henry Ford Hosp, Dept Radiat Oncol, Detroit, MI 48202 USA
关键词
Patient-specific QA; quality assurance; small fields; SRS and SBRT; stereotactic radiosurgery; QUALITY-ASSURANCE; SCINTILLATOR DETECTOR; MONTE-CARLO; LOG FILES; RADIOTHERAPY; ERRORS; IMRT; FIELDS;
D O I
10.1002/mp.12808
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeMobius 3D (M3D) provides a volumetric dose verification of the treatment planning system's calculated dose using an independent beam model and a collapsed cone convolution superposition algorithm. However, there is a lack of investigation into M3D's accuracy and effectiveness for stereotactic radiosurgery (SRS) and stereotactic body radiotherapy (SBRT) quality assurance (QA). Here, we collaborated with the vendor to develop a revised M3D beam model for SRS/SBRT cases treated with a 6X flattening filter-free (FFF) beam and high-definition multiple leaf collimator (HDMLC) on an Edge linear accelerator. MethodsEighty SRS/SBRT cases, planned with AAA dose algorithm and validated with Gafchromic film, were compared to M3D dose calculations using 3D gamma analysis with 2%/2 mm gamma criteria and a 10% threshold. A revised beam model was developed by refining the HD-MLC model in M3D to improve small field dose calculation accuracy and beam profile agreement. All cases were reanalyzed using the revised beam model. The impact of heterogeneity corrections for lung cases was investigated by applying lung density overrides to five cases. ResultsFor the standard and revised beam models, respectively, the mean gamma passing rates were 94.6% [standard deviation (SD): 6.1%] and 98.0% [SD: 1.7%] (for the overall patient), 88.2% [SD: 17.3%] and 93.8% [SD: 6.8%] (for the brain PTV), 71.4% [SD: 18.4%] and 81.5% [SD: 14.3%] (for the lung PTV), 83.3% [SD: 16.7%] and 67.9% [SD: 23.0%] (for the spine PTV), and 78.6% [SD: 14.0%] and 86.8% [SD: 12.5%] (for the PTV of all other sites). The lung PTV mean gamma passing rates improved from 74.1% [SD: 7.5%] to 89.3% [SD: 7.2%] with the lung density overridden. The revised beam model achieved an output factor within 3% of plastic scintillator measurements for 2 x 2 cm(2) MLC field size, but larger discrepancies are still seen for smaller field sizes which necessitate further improvement of the beam model. ConclusionSpecial attention needs to be paid to small field dosimetry, MLC modeling, and inhomogeneity corrections in the beam model for SRS/SBRT QA. The improvements noted in this study, and further collaborations between clinical physicists and the vendor to refine the M3D beam model could enable M3D to become a premier SRS/SBRT QA tool.
引用
收藏
页码:1351 / 1359
页数:9
相关论文
共 50 条
  • [41] Anatomy-based, patient-specific VMAT QA using EPID or MLC log files
    Defoor, Dewayne L.
    Vazquez-Quino, Luis A.
    Mavroidis, Panayiotis
    Papanikolaou, Nikos
    Stathakis, Sotirios
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (03): : 206 - 215
  • [42] Evaluation of a Diode Array for Patient Specific QA of An Automated VMAT SRS System
    Popple, R.
    Sullivan, R.
    Yuan, Y.
    Wu, X.
    MEDICAL PHYSICS, 2020, 47 (06) : E769 - E769
  • [43] A Practical Approach of Small Field Dosimetry Measurement for Patient-Specific IMRT/SRS/SBRT
    Wu, X.
    Popple, R.
    Yuan, Y.
    FIveash, J.
    Shen, S.
    Duan, J.
    Cardan, R.
    Brezovich, I.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [44] Correction method for depth and field size dependence in Octavius 1000 SRS patient-specific QA
    Shin, Han -Back
    Noh, Yu-Yun
    Kim, Heesoo
    Kim, Changhwan
    Sung, Kihoon
    Kim, Dong Wook
    RADIATION PHYSICS AND CHEMISTRY, 2024, 223
  • [45] MLC Consistency and Patient Specific VMAT QA with Log Files
    McDermott, R.
    Godley, A.
    Balik, S.
    MEDICAL PHYSICS, 2017, 44 (06) : 2938 - 2938
  • [46] A Viable Approach to Patient Specific QA for Spine VMAT SRS Using EPID-Based Dosimetry
    Song, Y.
    Zhang, P.
    Li, J.
    MEDICAL PHYSICS, 2012, 39 (06) : 3742 - 3742
  • [47] Validation and clinical use of a commercial Monte Carlo algorithm for Cyberknife patient-specific QA
    Milder, M. T. W.
    Sohn, M.
    Alber, M.
    Hoogeman, M. S.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S954 - S955
  • [48] Application of TG-218 action limits to SRS and SBRT pre-treatment patient specific QA
    Xia, Yuqing
    Adamson, Justus
    Zlateva, Yana
    Giles, Will
    JOURNAL OF RADIOSURGERY AND SBRT, 2020, 7 (02): : 135 - 147
  • [49] Patient-Specific QA for HDR Brachytherapy Skin Treatments
    Palaniswaamy, G.
    Massingill, B.
    Deb, N.
    Mutyala, S.
    Rangaraj, D.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [50] The limitation of patient-specific QA in IMRT with detector arrays
    Huh, S.
    Cho, W.
    Park, Y.
    MEDICAL PHYSICS, 2006, 33 (06) : 2143 - 2143