Refinement of MLC modeling improves commercial QA dosimetry system for SRS and SBRT patient-specific QA

被引:21
|
作者
Hillman, Yair [1 ]
Kim, Josh [2 ]
Chetty, Indrin [2 ]
Wen, Ning [2 ]
机构
[1] Karmanos Canc Inst McLaren Macomb, Dept Radiat Oncol, Mt Clemens, MI USA
[2] Henry Ford Hosp, Dept Radiat Oncol, Detroit, MI 48202 USA
关键词
Patient-specific QA; quality assurance; small fields; SRS and SBRT; stereotactic radiosurgery; QUALITY-ASSURANCE; SCINTILLATOR DETECTOR; MONTE-CARLO; LOG FILES; RADIOTHERAPY; ERRORS; IMRT; FIELDS;
D O I
10.1002/mp.12808
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeMobius 3D (M3D) provides a volumetric dose verification of the treatment planning system's calculated dose using an independent beam model and a collapsed cone convolution superposition algorithm. However, there is a lack of investigation into M3D's accuracy and effectiveness for stereotactic radiosurgery (SRS) and stereotactic body radiotherapy (SBRT) quality assurance (QA). Here, we collaborated with the vendor to develop a revised M3D beam model for SRS/SBRT cases treated with a 6X flattening filter-free (FFF) beam and high-definition multiple leaf collimator (HDMLC) on an Edge linear accelerator. MethodsEighty SRS/SBRT cases, planned with AAA dose algorithm and validated with Gafchromic film, were compared to M3D dose calculations using 3D gamma analysis with 2%/2 mm gamma criteria and a 10% threshold. A revised beam model was developed by refining the HD-MLC model in M3D to improve small field dose calculation accuracy and beam profile agreement. All cases were reanalyzed using the revised beam model. The impact of heterogeneity corrections for lung cases was investigated by applying lung density overrides to five cases. ResultsFor the standard and revised beam models, respectively, the mean gamma passing rates were 94.6% [standard deviation (SD): 6.1%] and 98.0% [SD: 1.7%] (for the overall patient), 88.2% [SD: 17.3%] and 93.8% [SD: 6.8%] (for the brain PTV), 71.4% [SD: 18.4%] and 81.5% [SD: 14.3%] (for the lung PTV), 83.3% [SD: 16.7%] and 67.9% [SD: 23.0%] (for the spine PTV), and 78.6% [SD: 14.0%] and 86.8% [SD: 12.5%] (for the PTV of all other sites). The lung PTV mean gamma passing rates improved from 74.1% [SD: 7.5%] to 89.3% [SD: 7.2%] with the lung density overridden. The revised beam model achieved an output factor within 3% of plastic scintillator measurements for 2 x 2 cm(2) MLC field size, but larger discrepancies are still seen for smaller field sizes which necessitate further improvement of the beam model. ConclusionSpecial attention needs to be paid to small field dosimetry, MLC modeling, and inhomogeneity corrections in the beam model for SRS/SBRT QA. The improvements noted in this study, and further collaborations between clinical physicists and the vendor to refine the M3D beam model could enable M3D to become a premier SRS/SBRT QA tool.
引用
收藏
页码:1351 / 1359
页数:9
相关论文
共 50 条
  • [1] Evaluation of QA Dosimetry System for SRS and SBRT Patient-Specific QA
    Hillman, Y.
    Kim, J.
    Chetty, I.
    Wen, N.
    MEDICAL PHYSICS, 2017, 44 (06) : 3273 - 3273
  • [2] SRS/SBRT patient-specific QA device comparison: Electronic QA devices vs film dosimetry
    Shacham, Tamir
    Epstein, Dan
    Levin, Daphne
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S3317 - S3318
  • [3] SRS/SBRT Patient-Specific QA Device Comparison
    Epstein, D.
    Shacham, T.
    Levin, D.
    MEDICAL PHYSICS, 2022, 49 (06) : E948 - E948
  • [4] Applications of Deep Learning in Patient-Specific QA of MLC-Equipped Robotic SBRT and SRS
    Han, B.
    Xing, L.
    Wang, L.
    MEDICAL PHYSICS, 2019, 46 (06) : E648 - E648
  • [5] Rotational Film Dosimetry Using a Simple Phantom for Patient-Specific QA of SBRT/SRS Plans
    Grams, M.
    Deisher, A.
    de Los Santos, L. Fong
    MEDICAL PHYSICS, 2017, 44 (06)
  • [6] Planar EPID-Based Dosimetry for SRS and SRT Patient-Specific QA
    Thongsawad, Sangutid
    Chanton, Tadchapong
    Saiyo, Nipon
    Udee, Nuntawat
    LIFE-BASEL, 2021, 11 (11):
  • [7] Evaluation of SRS MapCHECK™ for SABR patient-specific QA
    Miao, Y.
    Kidane, G.
    Ifthaker, A.
    Crees, L.
    Almond, E.
    RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S743 - S743
  • [8] Evaluation of the SRS MapCHECK for CyberKnife Patient-Specific QA
    Davis, S. D.
    Acha, A.
    George, S.
    Romaguera, W.
    Tirpak, L.
    Mishra, V.
    Gutierrez, A.
    MEDICAL PHYSICS, 2019, 46 (06) : E647 - E647
  • [9] Small Field SRS/SBRT Patient Specific QA at Extended SSD
    Stathakis, S.
    Mavroidis, P.
    Esquivel, C.
    Myers, P.
    Li, Y.
    Crownover, R.
    Papanikolaou, N.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [10] On the sensitivity of patient-specific IMRT QA to MLC positioning errors
    Yan, Guanghua
    Liu, Chihray
    Simon, Thomas A.
    Peng, Lee-Cheng
    Fox, Christopher
    Li, Jonathan G.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2009, 10 (01): : 120 - 128