Multicenter evaluation of hemoglobin A1c assay on capillary electrophoresis

被引:25
|
作者
Marinova, Mariela [1 ]
Altinier, Sara [1 ]
Caldini, Anna [2 ]
Passerini, Gabriella [3 ]
Pizzagalli, Giorgio [3 ]
Brogi, Marco [2 ]
Zaninotto, Martina [1 ]
Ceriotti, Ferruccio [3 ]
Plebani, Mario [1 ]
机构
[1] Univ Hosp Padova, Dept Lab Med, I-35128 Padua, Italy
[2] Careggi Univ Hosp, Gen Lab, Lab Diagnost Dept, I-50134 Florence, Italy
[3] Osped San Raffaele, Ist Sci, Diagnost & Ric San Raffaele, I-20123 Milan, Italy
关键词
Capillary electrophoresis; Diabetes; HPLC; HbA(1c); BIOLOGICAL VARIATION; HBA(1C); GLUCOSE; GOALS; A1C;
D O I
10.1016/j.cca.2013.06.014
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Hemoglobin A(1c) (HbA(1c)) measurement is currently used for the routine monitoring of long-term glycemic status, thus playing a fundamental role in the management of this disease. Since this marker has recently been recommended as an additional tool for diagnosing diabetes, it's of the utmost importance to ensure that the precision and accuracy of HbA(1c) methods are satisfactory. Methods: We assessed the analytical performances of the Capillarys 2 Flex Piercing (R) analyzer and compared the results obtained with those from two other widely used HPLC instruments. Furthermore, we evaluated the convenience and ergonomics of the system in authentic routine work conditions in three centers. Results: Within-laboratory (n = 40) and between-laboratory (n = 120) imprecision CV% using four blood samples with different concentrations of HbA(1c) were <3.4% and <3.1% using IFCC units and <2.1% and <2.0% using NGSP units, respectively. The obtained trueness (<3 mmol/mol, <0.3%) was highly satisfactory, nor was HbA(1c) measurement compromised by the presence of the commonly present hemoglobinopathies. The comparison made with established methods revealed excellent agreement (r > 0.985). Conclusions: The evaluated method is precise, accurate and robust, with a high throughput. It also allows the identification of the most frequent Hb variants and therefore may be a valid alternative to other methods currently proposed for routine use in clinical laboratories. (c) 2013 Elsevier B.V. All rights reserved.
引用
收藏
页码:207 / 211
页数:5
相关论文
共 50 条
  • [41] Utilization of Assay Performance Characteristics to Estimate Hemoglobin A1c Result Reliability
    Woodworth, Alison
    Korpi-Steiner, Nichole
    Miller, James J.
    Rao, Lokinendi V.
    Yundt-Pacheco, John
    Kuchipudi, Lakshmi
    Parvin, Curtis A.
    Rhea, Jeanne M.
    Molinaro, Ross
    CLINICAL CHEMISTRY, 2014, 60 (08) : 1073 - 1079
  • [42] Filter paper-dried blood spot assay of hemoglobin A1c
    Kapur, Sonia
    Kapur, Sanjay
    Zava, David
    DIABETES, 2007, 56 : A580 - A580
  • [43] Development of filter paper hemoglobin A1c assay applicable to newborn screening
    Pollock, Allison J.
    Allen, David B.
    Wiebe, Donald
    Eickhoff, Jens
    MacDonald, Michael
    Baker, Mei
    CLINICA CHIMICA ACTA, 2016, 457 : 24 - 26
  • [44] Evaluation of the tosoh HLC-723GHb V A1c 2.2 hemoglobin A1c analyzer
    Chevenne, D
    Marle, N
    Chauffert, M
    Noel, M
    Ducrocq, R
    Trivin, F
    CLINICAL BIOCHEMISTRY, 1999, 32 (06) : 487 - 490
  • [45] Led Astray by Hemoglobin A1c: A Case of Misdiagnosis of Diabetes by Falsely Elevated Hemoglobin A1c
    Chen, Jean
    Diesburg-Stanwood, Amy
    Bodor, Geza
    Rasouli, Neda
    JOURNAL OF INVESTIGATIVE MEDICINE HIGH IMPACT CASE REPORTS, 2016, 4 (01):
  • [46] Capillary electrophoresis of hemoglobin
    Jenkins, M
    Ratnaike, S
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2003, 41 (06) : 747 - 754
  • [47] Presumptive Identification of Clinically Significant Hemoglobin Variants Hb E, Hb S, Hb D in Hb A1c Capillary Electrophoresis
    Puthiyaveettil, Ankitha K.
    Vaz, Glen S.
    Prasad, Sujay R.
    Putchen, Deepalakshmi D.
    JOURNAL OF APPLIED LABORATORY MEDICINE, 2024,
  • [48] Evaluation of antioxidant effect by capillary electrophoresis assay
    Kohge, T
    Takano-Ishikawa, Y
    Kobori, M
    Yamaki, K
    JOURNAL OF THE JAPANESE SOCIETY FOR FOOD SCIENCE AND TECHNOLOGY-NIPPON SHOKUHIN KAGAKU KOGAKU KAISHI, 2003, 50 (02): : 85 - 89
  • [49] Evaluation of hemolysis effect on hemoglobin measurement by capillary electrophoresis
    Armin Mokhtariye
    Abdol-Reza Varasteh
    Amin Alaei
    Sima Marzban
    Fatemeh Keyfi
    Journal of Analytical Science and Technology, 10
  • [50] Evaluation of hemoglobin A1c measurement on the analyzer Cobas® c311
    El Boukhrissi, F.
    Balouch, H.
    Bamou, Y.