Accuracy of the autorefractor PowerRefractor in clinical work - A comparative study

被引:5
|
作者
Schittkowski, M
Hucks-Sievers, S
Krentz, H
Guthoff, R
机构
[1] Univ Augenklin, D-18057 Rostock, Germany
[2] Univ Rostock, Inst Med Informat & Biometrie, D-2500 Rostock 1, Germany
关键词
PowerRefractor; Retinomax; RK-5; videorefraction; photorefraction; objective refraction; retinoscopy;
D O I
10.1055/s-2005-858545
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background: The aim of this prospective study was to investigate accuracy and efficiency of the autorefractometer PowerRefractor compared to established autorefractometers and retinoscopy as standard method. Patients and Methods: 150 patients (300 eyes) were examined. Patients were initially (pupil not influenced) analyzed with the Power Refractor (PRneutral). After that a examination followed under standardized Cyclopentolat cycloplegia with the Power Refractor (PRcyclo), the hand-held Retinomax K-plus (Nikon), the table-top mounted RK-5 (Canon) and retinoscopy. Results: Deviation from retinoscopy (percentage of values with more than 1 dpt difference/maximal deviation): Sphere: PRneutral:66,3%/8,5 dpt; PRcyclo: 48%/5 dpt, RK-5: 26,7%/4,75 dpt; Retinomax K-plus: 6,7%/1,25 dpt; Cylinder: PRneutral: 12%/6,25 dpt; PRcyclo: 21%/4,75 dpt; RK-5: 1,3%/2,75 dpt; Retinomax K-plus: 2,3%/2,5 dpt; spherical equivalent: PRneutral: 44,7%/7,1 dpt; PRcyclo: 35,3%/5,4 dpt; RK-5: 9%/4,5 dpt; Retinomax K-plus: 1,3%/1,4 dpt; error of entire refraction: PRneutral: 62,7%/7,8 dpt; PRcyclo: 61,3%/7,3 dpt; RK-5: 24%/4,5 dpt; Retinomax K-plus: 14,7%/3,6 dpt Conclusions: The best accuracy compared to retinoscopy had Retinomax K-plus followed by RK-5. PowerRefractor had the biggest deviations of the tested devices. PowerRefractor may be a usefull device for screening small children and handicapped people because of the one meter observing distance. For a precise refraction especially in children a retinoscopy under cycloplegic conditions is still necessary.
引用
收藏
页码:983 / 992
页数:10
相关论文
共 50 条
  • [31] Comparative study of the accuracy for OpenFOAM solvers
    Bondarev, A. E.
    Kuvshinnikov, A. E.
    2017 IVANNIKOV ISPRAS OPEN CONFERENCE (ISPRAS), 2017, : 132 - 136
  • [32] Microlaparoscopy: A comparative study of diagnostic accuracy
    Faber, BM
    Coddington, CC
    FERTILITY AND STERILITY, 1997, 67 (05) : 952 - 954
  • [33] A comparative study on the accuracy of noninvasive thermometers
    Fadzil, Fadzlin Mohd
    Choon, David
    Arumugam, Kulenthran
    AUSTRALIAN FAMILY PHYSICIAN, 2010, 39 (04) : 237 - 239
  • [34] Accuracy of Two Autorefractors - Pediatric Autorefractor plusoptiX and Retinomax - in Cycloplegic Children in Comparison to Retinoscopy
    Schmidt-Bacher, A. E.
    Kahlert, C.
    Kolling, G.
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2010, 227 (10) : 792 - 797
  • [35] Accuracy of digital and conventional dental implant impressions for fixed partial dentures: A comparative clinical study
    Gedrimiene, Agne
    Adaskevicius, Rimas
    Rutkunas, Vygandas
    JOURNAL OF ADVANCED PROSTHODONTICS, 2019, 11 (05): : 271 - 279
  • [36] Comparative Accuracy of Stationary and Smartphone-Based Photogrammetry in Oral and Maxillofacial Surgery: A Clinical Study
    Seifert, Lukas B.
    Beyer, Michel
    Czok, Vanessa
    Aigner, Alexander
    Abazi, Sead
    Thieringer, Florian M.
    Sader, Robert
    JOURNAL OF CLINICAL MEDICINE, 2024, 13 (22)
  • [37] The ideal of accuracy in clinical work its importance, its limitations
    Cabot, RC
    BOSTON MEDICAL AND SURGICAL JOURNAL, 1904, 154 : 557 - 560
  • [38] Comparative Study of Road Profilers' Accuracy and Precision
    Wang, Hao
    Flintsch, Gerardo W.
    JOURNAL OF TESTING AND EVALUATION, 2010, 38 (02) : 188 - 194
  • [39] Comparative study on the accuracy of layout of transition curves
    Abdel-Malek, Magued N.
    Allakany, Abdel-Hamid Mohamed
    1600, Alexandria University, Alexandria, Egypt (34):
  • [40] Comparative study of accuracy of Doppler and grating velocimeters
    Popov, IA
    Veselov, LM
    Hanson, SG
    Gluzmann, A
    FOURTH INTERNATIONAL CONFERENCE ON VIBRATION MEASUREMENTS BY LASER TECHNIQUES: ADVANCES AND APPLICATIONS, 2000, 4072 : 222 - 227