Design and performance of a multi-pinhole collimation device for small animal imaging with clinical SPECT and SPECT-CT scanners

被引:52
|
作者
DiFilippo, Frank P. [1 ]
机构
[1] Cleveland Clin, Dept Nucl Med, Cleveland, OH 44195 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2008年 / 53卷 / 15期
关键词
D O I
10.1088/0031-9155/53/15/012
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A multi-pinhole collimation device is developed that uses the gamma camera detectors of a clinical SPECT or SPECT-CT scanner to produce high-resolution SPECT images. The device consists of a rotating cylindrical collimator having 22 tungsten pinholes with 0.9 mm diameter apertures and an animal bed inside the collimator that moves linearly to provide helical or ordered-subsets axial sampling. CT images also may be acquired on a SPECT-CT scanner for purposes of image co-registration and SPECT attenuation correction. The device is placed on the patient table of the scanner without attaching to the detectors or scanner gantry. The system geometry is calibrated in-place from point source data and is then used during image reconstruction. The SPECT imaging performance of the device is evaluated with test phantom scans. Spatial resolution from reconstructed point source images is measured to be 0.6 mm full width at half maximum or better. Micro-Derenzo phantom images demonstrate the ability to resolve 0.7 mm diameter rod patterns. The axial slabs of a Micro-Defrise phantom are visualized well. Collimator efficiency exceeds 0.05% at the center of the field of view, and images of a uniform phantom show acceptable uniformity and minimal artifact. The overall simplicity and relatively good imaging performance of the device make it an interesting low-cost alternative to dedicated small animal scanners.
引用
收藏
页码:4185 / 4201
页数:17
相关论文
共 50 条
  • [31] Multi-pinhole SPECT imaging with silicon strip detectors
    Peterson, Todd E.
    Shokouhi, Sepideh
    Furenlid, Lars R.
    Wilson, Donald W.
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 2752 - 2756
  • [32] Performance evaluation of a novel multi-pinhole SPECT system
    Jiang, Nianming
    Liu, Hui
    Xue, Mengle
    Li, Cheng
    Gao, Lilei
    Liu, Fan
    Wu, Jing
    Liu, Yaqiang
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2024, 1059
  • [33] Performance Evaluation on reconstructions in a Stationary Multi-pinhole SPECT
    Hsieh, Ho-Hui
    Lin, Kun-Ju
    Hsu, Chin-Han
    Hsiao, Ing-Tsung
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 3777 - +
  • [34] Preliminary Brain SPECT Multi-Pinhole Collimator Mechanical Design for DaTscan Imaging
    He, Yulun
    Konik, Arda
    De Beenhouwer, Jan
    Fromme, Timothy J.
    Lindsay, Clifford
    Kalluri, Kesava
    Li, Xin
    Momsen, Neil C.
    Furenlid, Lars R.
    King, Michael A.
    2017 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2017,
  • [35] Strategies to Reduce Artifacts and Improve Accuracy in Multiplexed Multi-Pinhole Small Animal SPECT
    DiFilippo, Frank P.
    Patel, Sagar
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 3151 - 3154
  • [36] On the development of a high-resolution multi-pinhole SPECT/CT
    Lackas, C
    Schramm, NU
    Hoppin, JW
    Halling, H
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 3893 - 3895
  • [37] Validation of CT-based Attenuation Correction for Multi-Pinhole PSF Reconstruction for Small-Animal SPECT
    Austin, Derek W.
    Feng, Bing
    Mintzer, Robert A.
    Chen, Mu
    Gregor, Jens
    Stuckey, Alan C.
    Wall, Jonathan S.
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 3339 - 3342
  • [38] Performance of a Novel Small-Animal SPECT System with Two Stationary Detectors: A Comparison of Multi-Pinhole Collimators for Mouse Imaging
    Hoffmann, J.
    Janssen, J.
    Kanno, T.
    Higuchi, T.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2020, 47 (SUPPL 1) : S453 - S453
  • [39] System Matrix Based on Sensitivity Model for Small Animal Multi-pinhole SPECT System
    Hsieh, Ho-Hui
    Hsu, Ching-Han
    Mok, Greta S. P.
    Tsai, Yu-Jung
    Chang, Shi-Ing
    Hsiao, Ing-Tsung
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 3308 - 3311
  • [40] Multi-pinhole Dynamic SPECT Imaging: Simulation and System Optimization
    Ma, Dan
    Clough, Anne V.
    Schmidt, Taly Gilat
    MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622