Effect of room illuminance on monitor black level luminance and monitor calibration

被引:26
|
作者
Chakrabarti, K
Kaczmarek, RV
Thomas, JA
Romanyukha, A
机构
[1] US FDA, Radiol Devices Branch,DRARD, ODE, CDRH, Rockville, MD 20850 USA
[2] OHIP, Div Mammog Qual & Radiat Program, CDRH, Rockville, MD 20850 USA
[3] Uniformed Serv Univ Hlth Sci, Dept Radiol, Bethesda, MD 20814 USA
关键词
black level; luminance; illuminance; image contrast; diffuse reflection; image surround;
D O I
10.1007/s10278-003-1720-5
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this article we demonstrate the effect of room illuminance and surrounding monitor black level luminance on image quality for soft copy interpretation. Luminance values of a 10% central target and image quality evaluations and observer performance using a contrast-detail mammography (CDMAM) phantom demonstrate these effects. Our results indicate that high room illuminance has a more damaging effect on image quality when the surrounding monitor luminance is 0% to 5% of the maximum monitor luminance. The effect of room illuminance is less obvious when the surrounding monitor luminance is 20% of the maximum.
引用
收藏
页码:350 / 355
页数:6
相关论文
共 50 条
  • [41] PNEUMOCARDIOGRAM - A POTENTIAL MONITOR FOR OPERATING ROOM
    SMITH, NT
    REITAN, JA
    ANESTHESIA AND ANALGESIA CURRENT RESEARCHES, 1970, 49 (05): : 781 - &
  • [42] SIMPLE MONITOR FOR OPERATING ROOM PHOTOGRAPHY
    BUTLER, LC
    PLASTIC AND RECONSTRUCTIVE SURGERY, 1971, 47 (06) : 600 - &
  • [43] Detector-Based Calibration for Illuminance and Luminance Meters-Experimental Results
    Fiorentin, Pietro
    Scroccaro, Alessandro
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2010, 59 (05) : 1375 - 1381
  • [44] Corrosive media level monitor
    不详
    BRITISH CORROSION JOURNAL, 1996, 31 (01): : 12 - 13
  • [45] MICROWAVE SURFACE LEVEL MONITOR
    STUCHLY, SS
    HAMID, MAK
    ANDRES, A
    IEEE TRANSACTIONS ON INDUSTRIAL ELECTRONICS AND CONTROL INSTRUMENTATION, 1971, IE18 (03): : 85 - &
  • [46] Luminance level of a monitor: influence on detectability and detection rate of breast cancer in 2D mammography
    Bemelmans, Frederic
    Rashidnasab, Alaleh
    Chesterman, Frederique
    Kimpe, Tom
    Bosmans, Hilde
    MEDICAL IMAGING 2016: IMAGE PERCEPTION, OBSERVER PERFORMANCE, AND TECHNOLOGY ASSESSMENT, 2016, 9787
  • [47] The Clinical Utility of the new FIVAFlow™ Monitor to Monitor the Administration of Intravenous Fluid in the Operating Room
    Hung, Christopher R.
    Sparrow, Kathryn
    Hung, Orlando
    ANESTHESIA AND ANALGESIA, 2024, 139 (06): : 2556 - 2557
  • [48] Influence of monitor luminance and tone scale on observer detection performance
    Krupinski, E
    Roehrig, H
    Furukawa, T
    Tang, CK
    IMAGE PERCEPTION: MEDICAL IMAGING 1998, 1998, 3340 : 99 - 104
  • [49] An instrument to monitor the luminance of pattern visual evoked potential stimuli
    Collins, A.D.
    Journal of Medical Engineering and Technology, 1985, 9 (03): : 123 - 124
  • [50] Impact of room illuminance on black level luminance and contrast detection for off-axis viewing on high resolution normal and high-bright flat panel displays
    Thomas, J
    Romanyukha, A
    Chakrabarti, K
    Kaczmarek, R
    MEDICAL PHYSICS, 2005, 32 (06) : 2110 - 2111