A microfluidic lab-on-chip derivatisation technique for the measurement of gas phase formaldehyde

被引:22
|
作者
Pang, Xiaobing [2 ]
Lewis, Alastair C. [1 ]
机构
[1] Univ York, Natl Ctr Atmospher Sci, York YO10 5DD, N Yorkshire, England
[2] Univ York, Dept Chem, York YO10 5DD, N Yorkshire, England
基金
英国工程与自然科学研究理事会; 英国自然环境研究理事会;
关键词
TOTAL ANALYSIS SYSTEMS; CARBONYL-COMPOUNDS; ATMOSPHERIC FORMALDEHYDE; AIRBORNE CARBONYLS; CHROMATOGRAPHY; AIR; MICROEXTRACTION; PHOTOOXIDATION; SCRUBBER; SORBENT;
D O I
10.1039/c2ay25028d
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A microfluidic lab-on-chip derivatisation technique has been optimized to achieve a rapid, automated and sensitive determination of ambient gaseous formaldehyde when used in combination with GC-MS. The method used a Pyrex micro-reactor comprising three inlets and one outlet, gas and fluid splitting and combining channels, mixing junctions, and a 2.0 m long, 620 mm internal diameter reaction micro-channel. The micro-reactor integrated three functions, that of: (1) mixer and reactor, (2) heater, and (3) preconcentrator. The flow rates of the gas sample and derivatisation solution and the temperature of the micro-reactor were optimized to achieve a near real-time measurement with a rapid and high efficiency derivatisation step following gas sampling. The enhanced phase contact area-to-volume ratio and the high heat transfer rate in the micro-reactor resulted in a fast and high efficiency derivatisation reaction. Calibration showed good linearity in the range of 26 to 331 ppb with correlation coefficients R-2 = 0.988 and 0.997 for PFPH and PFBHA derivatives. For low gas phase formaldehyde mixing ratios (<26 ppb) the derivatisation solution could be repeatedly recycled through the chip enabling preconcentration of the derivative - hydrazone. The calibration curves for this recycling approach also showed good linearity from 4.0 to 26 ppb with method detection limits (MDLs) of 2.1 ppb and 1.1 ppb for PFPH and PFBHA derivatives. The feasibility of the technique was assessed using measurements of laboratory ambient air, with formaldehyde the predominant carbonyl compound at a 12.5 ppb level. The proof of principle experiments demonstrated the potential of the approach for on-line measurements of other carbonyls including acetaldehyde, acetone and propionaldehyde.
引用
收藏
页码:2013 / 2020
页数:8
相关论文
共 50 条
  • [41] Optimization of Droplet Routing for an n-Plex Bioassay on a Digital Microfluidic Lab-on-Chip
    Zhao, Yang
    Sturmer, Ryan
    Chakrabarty, Krishnendu
    Pamula, Vamsee K.
    2009 IEEE BIOMEDICAL CIRCUITS AND SYSTEMS CONFERENCE (BIOCAS 2009), 2009, : 276 - +
  • [42] 3D visualization of convection patterns in lab-on-chip with open microfluidic outlet
    Gazzola, D.
    Scarselli, E. Franchi
    Guerrieri, R.
    MICROFLUIDICS AND NANOFLUIDICS, 2009, 7 (05) : 659 - 668
  • [43] Is a real-time quantifiable liquid biopsy achievable using a microfluidic lab-on-chip ?
    Casali, Veronica
    Guithon, Ingrid Clerc
    Sanden, Boudewijn van der
    Stephan, Olivier
    Gredy, Laetitia
    Vilgrain, Isabelle
    Martin, Donald K.
    EUROBIOTECH JOURNAL, 2023, 7 (04): : 189 - 195
  • [44] Integrated 3D Microfluidic Device for Impedance Spectroscopy in Lab-on-Chip Systems
    Buzzin, Alessio
    Iannascoli, Lorenzo
    Muzi, Marco
    Veroli, Andrea
    Caputo, Domenico
    de Cesare, Giampiero
    Maiolo, Luca
    Maita, Francesco
    Ricci, Giulia
    2019 IEEE 8TH INTERNATIONAL WORKSHOP ON ADVANCES IN SENSORS AND INTERFACES (IWASI), 2019, : 224 - 227
  • [45] Optimizing microfabricated liquid planar waveguides for microfluidic lab-on-chip flow cytometry systems
    Kunstmann-Olsen, C.
    Hoyland, J. D.
    Rubahn, H. -G.
    MICRO-STRUCTURED AND SPECIALTY OPTICAL FIBRES II, 2013, 8775
  • [46] On-Line Testing of Lab-on-Chip Using Reconfigurable Digital-Microfluidic Compactors
    Zhao, Yang
    Chakrabarty, Krishnendu
    INTERNATIONAL JOURNAL OF PARALLEL PROGRAMMING, 2009, 37 (04) : 370 - 388
  • [47] On-Line Testing of Lab-on-Chip Using Reconfigurable Digital-Microfluidic Compactors
    Yang Zhao
    Krishnendu Chakrabarty
    International Journal of Parallel Programming, 2009, 37 : 370 - 388
  • [48] 3D visualization of convection patterns in lab-on-chip with open microfluidic outlet
    D. Gazzola
    E. Franchi Scarselli
    R. Guerrieri
    Microfluidics and Nanofluidics, 2009, 7 : 659 - 668
  • [49] All inkjet-printed electroactive polymer actuators for microfluidic lab-on-chip systems
    Pabst, Oliver
    Beckert, Erik
    Perelaer, Jolke
    Schubert, Ulrich S.
    Eberhardt, Ramona
    Tuennermann, Andreas
    ELECTROACTIVE POLYMER ACTUATORS AND DEVICES (EAPAD) 2013, 2013, 8687
  • [50] Capacitance Variation Induced by Microfluidic Two-Phase Flow across Insulated Interdigital Electrodes in Lab-On-Chip Devices
    Dong, Tao
    Barbosa, Catia
    SENSORS, 2015, 15 (02): : 2694 - 2708