Validation of a Monte Carlo modelled gamma camera for Lutetium-177 imaging

被引:10
|
作者
Ramonaheng, K. [1 ]
van Staden, J. A. [1 ]
du Raan, H. [1 ]
机构
[1] Univ Free State, Dept Med Phys, POB 339, ZA-9300 Bloemfontein, South Africa
基金
新加坡国家研究基金会;
关键词
Validation; Modelling; Lu-177; Monte Carlo; SIMIND; Simulations; Experiments; Planar; SPECT; QUANTITATIVE SPECT; SCATTER CORRECTION; SIMULATION; DOSIMETRY; QUANTIFICATION; RECONSTRUCTION; THERAPY; ACCURACY; CT;
D O I
10.1016/j.apradiso.2020.109200
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
This study validated a model of the Siemens Symbia T16 dual-head SPECT/CT gamma camera created using the Monte Carlo program SIMIND for Lu-177. The validation was done by comparing experimental and simulated gamma camera performance criteria tests for the Lu-177 208 keV photopeak with a medium-energy collimator. Results showed good agreement between the experimental and simulated values. These results illustrated that SIMIND could emulate the Symbia T16 successfully and therefore, can be used with confidence to model (177) Lu images.
引用
收藏
页数:9
相关论文
共 50 条
  • [31] Gamma Camera Imaging with Rotating Multi-Pinhole Collimator. A Monte Carlo Feasibility Study
    Ilisie, Victor
    Moliner, Laura
    Morera, Constantino
    Nuyts, Johan
    Benlloch, Jose Maria
    SENSORS, 2021, 21 (10)
  • [32] Radiation Exposure of the Nuclear Medicine Personnel during new therapy procedures, the Lutetium-177 DOTATATE Production, Administration and Imaging
    Hoving, B. G.
    de Wit-van der Veen, B. J.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2016, 43 : S67 - S67
  • [33] Monte-Carlo Simulation and Experimental Validation of a Gamma-Camera Prototype Featuring a Novel Collimation Technique
    Meinderink, T.
    Osterloh, P.
    Osterloh, K.
    Goetz, T. I.
    Kuwert, T.
    Ritt, P.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2018, 45 : S18 - S18
  • [34] Ratiometric relationship between γH2AX imaging and Lutetium-177 is predictive of survival following radionuclide therapy in vivo
    O'Neill, E.
    Dias, G. M.
    Cornelissen, B.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2020, 47 (SUPPL 1) : S54 - S55
  • [35] Monte Carlo study on the Imaging resolution and sensitivity of a Comoton camera
    Lee, J. H.
    Lee, C. S.
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2006, 49 (01) : 97 - 102
  • [36] Determination of the DQE of gamma camera systems using the Monte Carlo technique
    Eriksson, I.
    Starck, S.
    Bath, M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2012, 39 : S393 - S393
  • [37] Monte Carlo simulation of radiation dose from terbium-161 and lutetium 177 in single tumor cells
    Alcocer-Avila, Mario
    Champion, Christophe
    Morgat, Clement
    Hindiae, Elif
    JOURNAL OF NUCLEAR MEDICINE, 2019, 60
  • [39] Monte Carlo Optimization of a Three-Stage Compton Camera for Prompt Gamma Imaging and Spectroscopy: Materials and Geometry
    Robertson, D.
    Peterson, S.
    Polf, J.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [40] Monte-Carlo simulation of pinhole collimator of a small field of view gamma camera for small animal imaging
    Zhu Jie
    Ma Wenyan
    Zhu Yufeng
    Ma Hongguang
    Wu Yuelei
    Hu Huasi
    Zhang Boping
    Huo Yonggang
    Liu Silu
    Han Bin
    Wang Zhaomin
    NUCLEAR SCIENCE AND TECHNIQUES, 2009, 20 (01) : 37 - 41