INSTRUMENTAL CONTRIBUTION TO UNCERTAINTY MEASUREMENT IN FLAME ATOMIC ABSORPTION SPECTROMETRY

被引:0
|
作者
Petrov, Alexander [1 ]
Gentscheva, Galia [1 ]
Havezov, Ivan [1 ]
Ivanova, Elisaveta [1 ]
机构
[1] Bulgarian Acad Sci, Inst Gen & Inorgan Chem, BU-1113 Sofia, Bulgaria
来源
关键词
flame atomic absorption spectrometry; instrumental contribution to uncertainty; copper; cobalt; cadmium; nickel;
D O I
暂无
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
The sources of uncertainty in the flame atomic absorption determination of copper, cobalt, cadmium and nickel, originating from the spectrometer, were studied. They were identified and divided in three main groups according to their origin - from the flame, from the light source (hollow-cathode lamp) and from the optics. The combined uncertainty of the flame atomic absorption determination was experimentally determined.
引用
收藏
页码:1235 / 1240
页数:6
相关论文
共 50 条
  • [41] ATOMIC-ABSORPTION, ATOMIC EMISSION, AND FLAME EMISSION-SPECTROMETRY
    JACKSON, KW
    QIAO, H
    ANALYTICAL CHEMISTRY, 1992, 64 (12) : R50 - R66
  • [42] OPTIMIZATION OF INSTRUMENTAL PARAMETERS IN FLAMELESS ATOMIC-ABSORPTION SPECTROMETRY
    POSMA, FD
    SMIT, HC
    ROOZE, AF
    ANALYTICAL CHEMISTRY, 1975, 47 (13) : 2087 - 2091
  • [43] ONLINE PRE-CONCENTRATION FOR THE MEASUREMENT OF CADMIUM IN URINE BY FLAME ATOMIC-ABSORPTION SPECTROMETRY
    KARAKAYA, A
    TAYLOR, A
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1989, 4 (03) : 261 - 263
  • [44] Beam injection flame furnace atomic absorption spectrometry:: A new flame method
    Gáspár, A
    Berndt, H
    ANALYTICAL CHEMISTRY, 2000, 72 (01) : 240 - 246
  • [45] Quantifying uncertainty in the measurement of arsenic in suspended particulate matter by Atomic Absorption Spectrometry with hydride generator
    Nahar Singh
    VN Ojha
    Nijhuma Kayal
    Tarushee Ahuja
    Prabhat K Gupta
    Chemistry Central Journal, 5
  • [46] Quantifying uncertainty in the measurement of arsenic in suspended particulate matter by Atomic Absorption Spectrometry with hydride generator
    Singh, Nahar
    Ojha, V. N.
    Kayal, Nijhuma
    Ahuja, Tarushee
    Gupta, Prabhat K.
    CHEMISTRY CENTRAL JOURNAL, 2011, 5
  • [47] MEASUREMENT OF LEAD IN BLOOD BY ATOMIC ABSORPTION SPECTROMETRY
    ELSER, RC
    SAVORY, J
    MUSHAK, P
    CLINICAL CHEMISTRY, 1969, 15 (08) : 788 - &
  • [48] Comparison of the analytical performance of flame atomic magneto-optic rotation spectrometry in the Faraday configuration with that of flame atomic absorption spectrometry
    Ince, AT
    Dawson, JB
    Snook, RD
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1996, 11 (10) : 967 - 972
  • [49] ATOMIC-ABSORPTION, ATOMIC-EMISSION, AND FLAME EMISSION-SPECTROMETRY
    JACKSON, KW
    MAHMOOD, TM
    ANALYTICAL CHEMISTRY, 1994, 66 (12) : R252 - R279
  • [50] Automation of the standard additions method in flame atomic absorption spectrometry
    López-García, I
    Viñas, P
    Gonzálvez, J
    Hernández-Córdoba, M
    TALANTA, 2002, 56 (04) : 787 - 796