Microseparation, fluid pressure and flow in failures of metal-on-metal hip resurfacing arthroplasties

被引:4
|
作者
Wroblewski, B. M. [1 ]
Siney, P. D. [1 ]
Fleming, P. A. [1 ]
机构
[1] Wrightington Hosp, John Charnley Res Inst, Appley Bridge WN6 9EP, Wigan, England
来源
BONE & JOINT RESEARCH | 2012年 / 1卷 / 03期
关键词
Metal-on-metal; Hip resurfacing; Microseparation; Abrasive water jet; Tissue damage; Metallic wear particles;
D O I
10.1302/2046-3758.13.2000040
中图分类号
Q813 [细胞工程];
学科分类号
摘要
Objectives Metal-on-metal (MoM) hip resurfacing was introduced into clinical practice because it was perceived to be a better alternative to conventional total hip replacement for young and active patients. However, an increasing number of reports of complications have arisen focusing on design and orientation of the components, the generation of metallic wear particles and serum levels of metallic ions. The procedure introduced a combination of two elements: large-dimension components and hard abrasive particles of metal wear. The objective of our study was to investigate the theory that microseparation of the articular surfaces draws in a high volume of bursal fluid and its contents into the articulation, and at relocation under load would generate high pressures of fluid ejection, resulting in an abrasive water jet. Methods This theoretical concept using MoM resurfacing components (head diameter 55 mm) was modelled mathematically and confirmed experimentally using a material-testing machine that pushed the head into the cup at a rate of 1000 mm/min until fully engaged. Results The mathematical model showed the pattern but not the force of fluid ejection, the highest pressures were expected when the separation of the components was only a fraction of one millimetre. The experimental work confirmed the results; with the mean peak ejection pressure of 43 763 N/m(2) equivalent to 306 mmHg or 5 psi. Conclusions The mechanical effect of the high-pressure abrasive water jet is the likely cause of the spectrum of complications reported with metal-on-metal resurfacing. Investigating serum levels of metallic elements may not be the best method for assessing the local mechanical effects of the abrasive water jet.
引用
收藏
页码:25 / 30
页数:6
相关论文
共 50 条
  • [21] A consensus paper on metal ions in metal-on-metal hip arthroplasties
    MacDonald, SJ
    Brodner, W
    Jacobs, JJ
    JOURNAL OF ARTHROPLASTY, 2004, 19 (08): : 12 - 16
  • [22] Necrotic and inflammatory changes in metal-on-metal resurfacing hip arthroplasties Relation to implant failure and pseudotumor formation
    Mahendra, Gayana
    Pandit, Hemant
    Kliskey, Karolina
    Murray, David
    Gill, Harinderjit Singh
    Athanasou, Nick
    ACTA ORTHOPAEDICA, 2009, 80 (06) : 653 - 659
  • [23] Clinical wear performance of metal-on-metal hip arthroplasties
    Rieker, CB
    Köttig, P
    Schön, R
    Windler, M
    Wyss, UP
    ALTERNATIVE BEARING SURFACES IN TOTAL JOINT REPLACEMENT, 1998, 1346 : 144 - 156
  • [24] Declining Revision Burden of Metal-on-Metal Hip Arthroplasties
    Lainiala, Olli S.
    Reito, Aleksi P.
    Nieminen, Jyrki J.
    Eskelinen, Antti P.
    JOURNAL OF ARTHROPLASTY, 2019, 34 (09): : 2058 - +
  • [25] Cross sectional imaging of metal-on-metal hip arthroplasties
    Robinson, Elizabeth
    Henckel, Johann
    Sabah, Shiraz
    Satchithananda, Keshthra
    Skinner, John
    Hart, Alister
    ACTA ORTHOPAEDICA, 2015, 86 (02) : 273 - 273
  • [26] Metallosis in metal-on-metal PPF total hip arthroplasties
    Legenstein, R.
    Huber, W.
    Boesch, P.
    BIOCERAMICS AND ALTERNATIVE BEARINGS IN JOINT ARTHROPLASTY, 2007, : 193 - 203
  • [27] Clinical wear performance of metal-on-metal hip arthroplasties
    Rieker, Claude B.
    Köttig, Petra
    Schön, Rolf
    Windler, Markus
    Wyss, Urs P.
    ASTM Special Technical Publication, 1999, (1346): : 144 - 156
  • [28] Revision rates for metal-on-metal hip resurfacing and metal-on-metal total hip arthroplasty - a systematic review
    Sorensen, Sofie-Amalie L. Ras
    Jorgensen, Henrik L.
    Sporing, Sune L.
    Lauritzen, Jes B.
    HIP INTERNATIONAL, 2016, 26 (06) : 515 - 521
  • [29] Complications After Metal-on-Metal Hip Resurfacing Arthroplasty
    Amstutz, Harlan C.
    Le Duff, Michel J.
    Campbell, Patricia A.
    Wisk, Lauren E.
    Takamura, Karren M.
    ORTHOPEDIC CLINICS OF NORTH AMERICA, 2011, 42 (02) : 207 - +
  • [30] Metal-on-metal hip resurfacing in osteonecrosis of the femoral head
    Revell, Matthew P.
    McBryde, Callum W.
    Bhatnagar, Sharad
    Pynsent, Paul B.
    Treacy, Ronan B. C.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2006, 88A : 98 - 103