Graphics processing unit-based dispersion encoded full-range frequency-domain optical coherence tomography

被引:8
|
作者
Wang, Ling [2 ,3 ]
Hofer, Bernd [1 ]
Guggenheim, Jeremy A. [2 ]
Povazay, Boris [1 ]
机构
[1] Med Univ Vienna, Ctr Med Phys & Biomed Engn, A-1090 Vienna, Austria
[2] Cardiff Univ, Sch Optometry & Vis Sci, Cardiff CF24 4LU, S Glam, Wales
[3] Katholieke Univ Leuven, Dept Biosyst, Fac Biosci Engn, B-3001 Heverlee, Belgium
关键词
optical coherence tomography; medical imaging; interferometers; graphics processing unit; dispersion; ULTRAHIGH-SPEED; AXIAL SCANS; NM;
D O I
10.1117/1.JBO.17.7.077007
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Dispersion encoded full-range (DEFR) frequency-domain optical coherence tomography (FD-OCT) and its enhanced version, fast DEFR, utilize dispersion mismatch between sample and reference arm to eliminate the ambiguity in OCT signals caused by non-complex valued spectral measurement, thereby numerically doubling the usable information content. By iteratively suppressing asymmetrically dispersed complex conjugate artifacts of OCT-signal pulses the complex valued signal can be recovered without additional measurements, thus doubling the spatial signal range to cover the full positive and negative sampling range. Previously the computational complexity and low processing speed limited application of DEFR to smaller amounts of data and did not allow for interactive operation at high resolution. We report a graphics processing unit (GPU)-based implementation of fast DEFR, which significantly improves reconstruction speed by a factor of more than 90 in respect to CPU-based processing and thereby overcomes these limitations. implemented on a commercial low-cost GPU, a display line rate of similar to 21,000 depth scans/s for 2048 samples/depth scan using 10 iterations of the fast DEFR algorithm has been achieved, sufficient for real-time visualization in situ. (C) 2012 Society of Photo-Optical Instrumentation Engineers (SPIE). [DOI: 10.1117/1.JBO.17.7.077007]
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Graphics processing unit based dispersion encoded full-range frequency domain OCT
    Wang, Ling
    Povazay, Boris
    Guggenheim, Jeremy A.
    Drexler, Wolfgang
    Hofer, Bernd
    OPTICAL COHERENCE TOMOGRAPHY AND COHERENCE DOMAIN OPTICAL METHODS IN BIOMEDICINE XVI, 2012, 8213
  • [2] Dispersion encoded full range frequency domain optical coherence tomography
    Hofer, Bernd
    Povazay, Boris
    Hermann, Boris
    Unterhuber, Angelika
    Matz, Gerald
    Drexler, Wolfgang
    OPTICS EXPRESS, 2009, 17 (01): : 7 - 24
  • [3] An advanced algorithm for dispersion encoded full range frequency domain optical coherence tomography
    Koettig, Felix
    Cimalla, Peter
    Gaertner, Maria
    Koch, Edmund
    OPTICAL COHERENCE TOMOGRAPHY AND COHERENCE TECHNIQUES VI, 2013, 8802
  • [4] An advanced algorithm for dispersion encoded full range frequency domain optical coherence tomography
    Koettig, Felix
    Cimalla, Peter
    Gaertner, Maria
    Koch, Edmund
    OPTICS EXPRESS, 2012, 20 (22): : 24925 - 24948
  • [5] Dual-reference full-range frequency domain optical coherence tomography
    Lee, Kye-Sung
    Meemon, Panomsak
    Hsu, Kevin
    Dallas, William
    Rolland, Jannick P.
    DESIGN AND QUALITY FOR BIOMEDICAL TECHNOLOGIES II, 2009, 7170
  • [6] Full-range spectral domain Doppler optical coherence tomography
    Meemon, P.
    Lee, K. S.
    Rolland, J. P.
    DESIGN AND QUALITY FOR BIOMEDICAL TECHNOLOGIES III, 2010, 7556
  • [7] Graphics Processing Unit-Based Ultrahigh Speed Real-Time Fourier Domain Optical Coherence Tomography
    Zhang, Kang
    Kang, Jin U.
    IEEE JOURNAL OF SELECTED TOPICS IN QUANTUM ELECTRONICS, 2012, 18 (04) : 1270 - 1279
  • [8] Full-field time-encoded frequency-domain optical coherence tomography
    Povazay, Boris
    Unterhuber, Angelika
    Hermann, Boris
    Sattmann, Harald
    Arthaber, Holger
    Drexler, Wolfgang
    OPTICS EXPRESS, 2006, 14 (17): : 7661 - 7669
  • [9] Doppler imaging with dual-detection full-range frequency domain optical coherence tomography
    Meemon, Panomsak
    Lee, Kye-Sung
    Rolland, Jannick P.
    BIOMEDICAL OPTICS EXPRESS, 2010, 1 (02): : 537 - 552
  • [10] Full-range optical coherence refraction tomography
    Huang, Yong
    Qiao, Zhengyu
    Chen, Jingsi
    Zhang, Dan
    Hao, Qun
    OPTICS LETTERS, 2022, 47 (04) : 894 - 897