On the numerical investigation of cardiovascular balloon-expandable stent using finite element method

被引:30
|
作者
Azaouzi, M. [1 ]
Makradi, A. [1 ]
Petit, J. [2 ]
Belouettar, S. [1 ]
Polit, O. [2 ]
机构
[1] Ctr Rech Publ Henri Tudor, Adv Mat & Struct Dept, L-4002 Esch Sur Alzette, Luxembourg
[2] Univ Paris Ouest, LEME, F-92410 Ville Davray, France
关键词
Cardiovascular stent; Finite element analysis; Numerical simulation; Fatigue life prediction; Goodman diagram; FATIGUE; SIZE; COMPONENTS; BEHAVIOR; FRACTURE; DESIGN; STRUTS;
D O I
10.1016/j.commatsci.2013.05.043
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The present paper shows an effective methodology which is based on Finite Element Analysis (FEA) for the fatigue life prediction (FLP) of a Balloon-Expandable (BE) stent made of stainless steel material (AISI316L). BE stent is a tubular, often mesh-like, structure which is expanded inside a diseased (stenosed) artery segment to restore blood flow and keep the vessel open following angioplasty. With design lives of 10-15 years, fatigue life prediction of these devices is critical for the designer, as fatigue failure may occur after deployment inside the blood vessel due to the heart beating (typically 4 x 10(7) cycles/year). Therefore, it is crucial to study the effect of stenting conditions and stent design on the fatigue life of BE stents. The objective of this paper is to provide quantitative measures of the stent's stress amplitude and mean stress which are generated by the cardiac pulse pressure. This allows prediction of the device's life and optimization of stent designs. In this paper, numerical investigations were conducted using an effective approach based on the well-known stress-based Goodman Diagram (GD) for the FLP of a given stent design. It has been demonstrated that the percentage of artery's expansion as well as the dimensions of the struts (the fundamental structural unit that forms the stent), have both an impact on the fatigue behavior after stent deployment. (C) 2013 Elsevier B. V. All rights reserved.
引用
收藏
页码:326 / 335
页数:10
相关论文
共 50 条
  • [21] Design and modeling balloon-expandable coronary stent for manufacturability
    Suryawan, D.
    Suyitno
    MATERIAL CHEMISTRY DEVELOPMENT FOR FUTURE MEDICINE, INDUSTRY, ENVIRONMENTAL AND BIOMATERIAL APPLICATION, 2017, 172
  • [22] Balloon-expandable stent complications during endovascular grafting
    Ombrellaro, MP
    Stevens, SL
    Kerstetter, K
    Freeman, MB
    Goldman, MH
    VASCULAR SURGERY, 1996, 30 (04): : 293 - 299
  • [23] Balloon-Expandable Covered Stent to Superficial Femoral Artery
    Tamura, Hiroto
    Yuba, Kenichiro
    Kurimoto, Shingo
    JOURNAL OF INVASIVE CARDIOLOGY, 2023, 35 (08):
  • [24] THE MECHANICAL BEHAVIOR OF A BALLOON-EXPANDABLE STENT IN A STENOTIC ARTERY
    Imani, Misagh
    Goudarzi, A. M.
    Ganji, D. D.
    Gerdroodbary, M. Barzegar
    PROCEEDINGS OF THE ASME 12TH BIENNIAL CONFERENCE ON ENGINEERING SYSTEMS DESIGN AND ANALYSIS - 2014, VOL 1, 2014,
  • [25] Numerical simulation method of balloon-expandable coronary stents expansion mechanism
    School of Mechanical Engineering, Southeast University, Nanjing 210096, China
    Jixie Gongcheng Xuebao, 2008, 1 (102-108):
  • [26] A Novel Approach to Aortoiliac Bifurcation Stenting Using a Single Balloon-Expandable Stent
    Alomar, Mohammed
    Tayal, Rajiv
    Amponsah, Michael
    Cohen, Marc
    Wasty, Najam
    JOURNAL OF INVASIVE CARDIOLOGY, 2019, 31 (01): : E8 - E8
  • [27] Repair of a splenic artery aneurysm using a novel balloon-expandable covered stent
    Guller, Jeffrey
    Samman, Mahmoud
    Deyer, Timothy W.
    Lookstein, Robert
    Nowakowski, Scott
    VASCULAR MEDICINE, 2006, 11 (02) : 111 - 113
  • [28] Crimping and repositioning of a maldeployed balloon-expandable arterial stent using a gooseneck snare
    Portugaller, HR
    Pabst, E
    Doerfler, OC
    Tauss, J
    Zangrando, M
    Pilger, E
    Klein, GE
    JOURNAL OF ENDOVASCULAR THERAPY, 2005, 12 (02) : 247 - 251
  • [29] Experimental trial of balloon-expandable, metallic Palmaz stent in the trachea
    Fraga, JC
    Filler, RM
    Forte, V
    Bahoric, A
    Smith, C
    ARCHIVES OF OTOLARYNGOLOGY-HEAD & NECK SURGERY, 1997, 123 (05) : 522 - 528
  • [30] Research on the Coupling Expansion Mechanism of Balloon-Expandable Coronary Stent
    Feng, Haiquan
    Guo, Feifei
    Jiang, Xudong
    Han, Qingsong
    ADVANCES IN FUTURE COMPUTER AND CONTROL SYSTEMS, VOL 2, 2012, 160 : 585 - 590