Development and application of a miniature mass spectrometer with continuous sub-atmospheric pressure interface and integrated ionization source

被引:7
|
作者
Zhu, Yanping [1 ]
Zhang, Ruina [1 ]
Wang, Kai [1 ]
Lu, Jun [1 ]
Lin, Lin [2 ]
Yang, Lili [3 ]
Zhang, Qian [4 ]
Wang, Xiaohao [1 ]
Yu, Quan [1 ]
机构
[1] Tsinghua Univ, Shenzhen Int Grad Sch, Div Adv Mfg, Shenzhen 518055, Peoples R China
[2] Southern Univ Sci & Technol, Sustech Core Res Facil, Shenzhen 518055, Peoples R China
[3] Yanshan Univ, Sch Elect Engn, Qinhuangdao 066004, Hebei, Peoples R China
[4] China Acad Informat & Commun Technol, Informatizat & Ind Integrat Res Inst, Beijing 100191, Peoples R China
基金
中国国家自然科学基金;
关键词
PHOTOIONIZATION; SIMULATION;
D O I
10.1016/j.talanta.2022.123994
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
For the miniature mass spectrometer (MS) with a continuous atmospheric pressure interface (CAPI), the gas in the multi-stage chambers directly affects the performance of the instrument. In this study, a sealed ionization chamber is designed to couple with a conventional mini CAPI-MS. In this configuration, the gas environment in the first ionization chamber can be flexibly changed to regulate the gas conditions throughout the entire instrument. By studying the effect of gas pressure on the performance of the instrument, we found that the instrument shows some unique advantages when the first ionization chamber is under sub-atmospheric pressure (SAP) conditions, such as reducing the load of the vacuum pump by 40%, achieving pump-free injection for gas and liquid samples, and improving the resolution by a factor of 2 without loss of detection sensitivity. Therefore, we propose a new integrated interface called continuous sub-atmospheric pressure interface (CSAPI) for building a miniature ion trap mass spectrometer. The CSAPI specially integrates sample introduction, gas/ions interface, and ionizations, including electrospray ionization (ESI) and secondary electrospray ionization (SESI), making this system more convenient for non-professional handlers to rapidly identify or monitor target analytes in gaseous- and solution-phase samples. We also use this system to study gas composition to further improve performance, being able to achieve a 5-fold sensitivity and 2-fold resolution improvement. At last, some custom applications of the current CSAPI-MS platform are explored and demonstrated, including real-time monitoring of chemical reactions in solution and long-distance sampling and analysis of dried Chinese herbs. In conclusion, this study provides a new approach to constructing a complete, versatile and practical miniature MS instrument.
引用
收藏
页数:7
相关论文
共 50 条
  • [41] Impurity measurement in specialty gases using an atmospheric pressure ionization mass spectrometer with a two-compartment ion source
    Kitano, M
    Shirai, Y
    Ohki, A
    Babasaki, S
    Ohmi, T
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 2001, 40 (4B): : 2688 - 2693
  • [42] Continuous boilover behaviors of large-scale kerosene pool fires under sub-atmospheric pressure
    Zhao, Jinlong
    Zhang, Qingyuan
    Wang, Zhenhua
    Yang, Rui
    Zhang, Jianping
    PROCESS SAFETY AND ENVIRONMENTAL PROTECTION, 2023, 177 : 1431 - 1439
  • [43] Development of an integrated, single electrode liquid sampling - atmospheric pressure glow discharge microplasma ionization source
    Williams, Tyler J.
    Marcus, R. Kenneth
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2021, 179
  • [44] Optimization of a supercritical fluid chromatograph atmospheric pressure chemical ionization mass spectrometer interface using an ion trap and two quadrupole mass spectrometers
    Morgan, DG
    Harbol, KL
    Kitrinos, NP
    JOURNAL OF CHROMATOGRAPHY A, 1998, 800 (01) : 39 - 49
  • [45] LOW-PRESSURE CHEMICAL IONIZATION SOURCE - APPLICATION OF A NOVEL TYPE OF ION STORAGE MASS-SPECTROMETER
    BONNER, RF
    TODD, JFJ
    LAWSON, G
    JOURNAL OF THE CHEMICAL SOCIETY-CHEMICAL COMMUNICATIONS, 1972, (21) : 1179 - &
  • [46] Following the Ions through a Mass Spectrometer with Atmospheric Pressure Interface: Simulation of Complete Ion Trajectories from Ion Source to Mass Analyzer
    Zhou, Xiaoyu
    Ouyang, Zheng
    ANALYTICAL CHEMISTRY, 2016, 88 (14) : 7033 - 7040
  • [47] Radial Electric Field Driven Collision-Induced Dissociation in a Miniature Continuous Atmospheric Pressure Interfaced Ion Trap Mass Spectrometer
    Ruan, Huiwen
    Xu, Chuting
    Fan, Zhigang
    Wang, Weiguo
    Li, Haiyang
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2023, 34 (12) : 2822 - 2830
  • [48] Studies of the mechanism of the cluster formation in a thermally sampling atmospheric pressure ionization mass spectrometer
    Albrecht, Sascha
    Klopotowski, Sebastian
    Derpmann, Valerie
    Klee, Sonja
    Brockmann, Klaus J.
    Stroh, Fred
    Benter, Thorsten
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2014, 85 (01):
  • [49] APPLICATION OF LIQUID-CHROMATOGRAPHY ATMOSPHERIC-PRESSURE CHEMICAL-IONIZATION MASS-SPECTROMETRY TO A SECTOR MASS-SPECTROMETER
    NOJIMA, K
    FUJIMAKI, S
    HERTSENS, RC
    MORITA, T
    JOURNAL OF CHROMATOGRAPHY A, 1995, 712 (01) : 17 - 19
  • [50] Simulated and developed an electron impact ionization source for space miniature time-of-flight mass spectrometer
    Ren, Zhengyi
    Guo, Meiru
    Cheng, Yongjun
    Sun, Wenjun
    Dong, Meng
    Li, Gang
    Pei, Xiaoqiang
    Wu, Chengyao
    VACUUM, 2020, 174