Usefulness of modular neck adapter in partial hip revision

被引:0
|
作者
Pautasso, Andrea [1 ]
Bardellini, Giuseppe [2 ]
Stissi, Placido [1 ]
D'Angelo, Fabio [1 ,3 ]
机构
[1] ASST Sette Laghi, Circolo Hosp & Macchi Fdn Varese, Univ Ctr, Orthopaed & Traumatol Dept, Viale Luigi Borri 57, I-21100 Varese, Italy
[2] Univ Studies Insubria, Orthopaed & Traumatol Dept, Varese, Italy
[3] Univ Studies Insubria, Dept Biotechnol & Life Sci DBSV, Varese, Italy
来源
ANNALS OF JOINT | 2023年 / 8卷
关键词
Hip revision arthroplasty; head-neck adapter; total hip arthroplasty (THA); Merete BioBall; FEMORAL OFFSET; CERAMIC HEADS; LEG LENGTH; ARTHROPLASTY; RESTORATION; MULTICENTER; FRACTURE;
D O I
10.21037/aoj-23-22
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background and Objective: Modular neck adapters allow different length and offset changes to reach a stable total hip arthroplasty (THA) and permit a quick partial hip revision procedure without removing the existing components. The literature is poor on this matter and about the long-term related outcomes. This narrative review summarizes the most recent literature about these devices as an option of surgical treatment in partial total hip arthroplasty revision (THAr) focusing on indications, clinical and radiological outcomes, and related complications. Methods: The narrative review of the current available literature was conducted in December 2022 through electronic database. The terms used were: "Head neck taper" OR "Merete BioBall" AND "revision Total Hip Arthroplasty (MeSH Terms)". The timeframe was limited between 01/01/2000 and 01/12/2022. The studies regarding the clinical use of the Merete BioBall (R) system in hip revision surgery were included, while all the papers concerning modular stem prosthesis were excluded. Key Content and Findings: The surgical procedure is safe, quick and allows the surgeon to correct a well-fixed stem version, length and offset, besides retensioning soft tissues. Clinical and radiological outcomes are good with low complications rates. Conclusions: The modular neck adapter system seems to be a good surgical procedure for recurrent dislocation of THA, especially in case of a second THAr surgery. However, the main indication of adapter use remains the isolated acetabular cup revision. The related complications are rare: the worst is the redislocation due to an insufficient stem version and length correction. Re-dislocation rates reported in literature vary from 5.2% to 15%. Corrosion or fretting of the modular system are not reported in literature.
引用
收藏
页数:7
相关论文
共 50 条
  • [41] Modular Revision System for Complex Hip Revisions: A Case Series
    Hameed, Daniel
    Oster, Brittany
    Springer, Bryan d.
    Malkani, Arthur l.
    Mont, Michael a.
    SURGICAL TECHNOLOGY INTERNATIONAL-INTERNATIONAL DEVELOPMENTS IN SURGERY AND SURGICAL RESEARCH, 2024, 44
  • [42] Factors influencing taper failure of modular revision hip stems
    Krull, A.
    Morlock, M. M.
    Bishop, N. E.
    MEDICAL ENGINEERING & PHYSICS, 2018, 54 : 65 - 73
  • [43] Modular Tantalum Augments for Acetabular Defects in Revision Hip Arthroplasty
    Siegmeth, Alexander
    Duncan, Clive P.
    Masri, Bassam A.
    Kim, Winston Y.
    Garbuz, Donald S.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2009, 467 (01) : 199 - 205
  • [44] Revision hip arthroplasty using a cementless modular tapered stem
    Christophe Pattyn
    Alexander Mulliez
    René Verdonk
    Emmanuel Audenaert
    International Orthopaedics, 2012, 36 : 35 - 41
  • [45] Catastrophic failure of a modular revision total hip polyethylene insert
    Stiehl, James Bowen
    Mahfouz, Mohammed R.
    JOURNAL OF ARTHROPLASTY, 2007, 22 (01): : 143 - 147
  • [46] Revision Total Hip Arthroplasty With a Modular Cementless Femoral Stem
    Pelt, Christoper E.
    Madsen, Wes
    Erickson, Jill A.
    Gililland, Jeremy M.
    Anderson, Mike B.
    Peters, Christopher L.
    JOURNAL OF ARTHROPLASTY, 2014, 29 (09): : 1803 - 1807
  • [47] Implant-Related Fractures of the Femoral Neck Cone Adapter of a Modular Short-Stem Hip Prosthesis - Patient Management and Operative Technique
    Kohler, S.
    Ratayski, H.
    Zacher, J.
    ZEITSCHRIFT FUR ORTHOPADIE UND UNFALLCHIRURGIE, 2011, 149 (02): : 185 - 190
  • [48] Revision of Recalled Modular Neck Rejuvenate and ABG Femoral Implants
    Walsh, Christopher P.
    Hubbard, James C.
    Nessler, Joseph P.
    Markel, David C.
    JOURNAL OF ARTHROPLASTY, 2015, 30 (05): : 822 - 826
  • [49] Metallosis During Partial Component Hip Revision Arthroplasty
    Vermesan, Dinu
    Prejbeanu, Radu
    Haragus, Horia
    Ahmadi, Mirela
    Damian, Gratian
    REVISTA DE CHIMIE, 2012, 63 (09): : 953 - 955
  • [50] Acetabular defect reconstruction in revision hip arthroplasty with a modular revision system and biological defect augmentation
    Jan Schmolders
    Max J. Friedrich
    Robert D. Michel
    Thomas M. Randau
    Matthias D. Wimmer
    Andreas C. Strauss
    Hendrik Kohlhof
    Dieter C. Wirtz
    Sascha Gravius
    International Orthopaedics, 2015, 39 : 623 - 630