Comprehensive commissioning and quality assurance validation of EthosTM therapy

被引:0
|
作者
Morel, A. [1 ]
Prunaretty, J. [1 ]
Trauchessec, D. [1 ]
Ailleres, N. [1 ]
Fenoglietto, P. [1 ]
Azria, D. [1 ]
机构
[1] Inst Canc Montpellier ICM, Montpellier, France
来源
CANCER RADIOTHERAPIE | 2023年 / 27卷 / 05期
关键词
EthosTM; Commissioning; QA validation; ADAPTIVE RADIATION-THERAPY; CLINICAL IMPLEMENTATION; RADIOTHERAPY; CANCER;
D O I
10.1016/j.canrad.2022.10.001
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose. - Adaptive radiotherapy with the Ethos((R)) therapy Varian system has been recently implementedat the Montpellier Cancer Institute, France. This article details the commissioning performed before the implementation of this new treatment planning system (TPS). Material and methods. - To validate the golden beam data of the machine (Halcyon linear accelerator), percentage depth doses (PDD) and profiles were measured for several field sizes and at different depths with a microdiamond chamber. The final doses calculated for different plan types with the Ethos Acuros XB algorithm and the Halcyon Eclipse Analytic Anisotropic Algorithm were compared using the gamma index method. Lastly, for the patient quality assurance (QA) process, the patient treatment plan results obtained with the Mobius3D QA platform (Varian) were compared with the portal dosimetry results obtained with Epiqa (Epidos). Results. - Minor differences were observed for the PDD and profile curves (mean difference of 0.2% and 2%, respectively). The chi index pass rate was above 98% for all measures using the 1%/1 mm and 2%/2 mm criteria for PDD and profile evaluations. The Ethos AXB algorithm was validated for every configuration( fixed fields, standard IMRT and VMAT fields, and clinical plans) with 2D/3D gamma index values > 99%. Seventy-three 3-arcs-VMAT QA plans and 27 9-fields-IMRT QA plans were evaluated. Both showed excellent agreement with the TPS calculations (mean gamma pass rate higher than 99%). No difference was observed between IMRT and VMAT. Conclusion. - The beam delivery, the Ethos AXB algorithm, and the patient QA were comprehensively validated using independent tools. (c) 2023 Societe francaise de radiotherapie oncologique (SFRO). Published by Elsevier Masson SAS. Allrights reserved.
引用
收藏
页码:355 / 361
页数:7
相关论文
共 50 条
  • [21] Commissioning and Quality Assurance of VMAT with An Ion Chamber Profiler
    Yan, G.
    Simon, T.
    Peng, L.
    Liu, C.
    Li, J.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [22] Commissioning and routine quality assurance of the Radixact Synchrony system
    Goddard, Lee
    Jeong, Kyoungkeun
    Tome, Wolfgang A.
    MEDICAL PHYSICS, 2022, 49 (02) : 1181 - 1195
  • [23] On the acceptance, commissioning, and quality assurance of electron FLASH units
    Palmiero, Allison
    Liu, Kevin
    Colnot, Julie
    Chopra, Nitish
    Neill, Denae
    Connell, Luke
    Velasquez, Brett
    Koong, Albert C.
    Lin, Steven H.
    Balter, Peter
    Tailor, Ramesh
    Robert, Charlotte
    Germond, Jean-Francois
    Jorge, Patrik Goncalves
    Geyer, Reiner
    Beddar, Sam
    Moeckli, Raphael
    Schuler, Emil
    MEDICAL PHYSICS, 2025, 52 (02) : 1207 - 1223
  • [24] Commissioning and quality assurance of Halcyon™ 2.0 linear accelerator
    Pathak, Pushpraj K.
    Vashisht, S. K.
    Baby, S.
    Jithin, P. K.
    Jain, Y.
    Mahawar, R.
    Sharan, V. G. G. K.
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2021, 26 (03) : 433 - 444
  • [25] COMMISSIONING AND QUALITY ASSURANCE OF TREATMENT PLANNING COMPUTERS - COMMENTS
    MCSHAN, DL
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 26 (02): : 371 - 372
  • [26] Commissioning and quality assurance of commercial in vivo dosimetry systems
    Esposito, M.
    RADIOTHERAPY AND ONCOLOGY, 2022, 170 : S1 - S2
  • [27] Comprehensive Quality Assurance Services
    Waste Disposal and Water Management in Australia, 1995, 22 (01):
  • [28] Commissioning and quality assurance of RapidArc radiotherapy delivery system
    Ling, C. Clifton
    Zhang, Pengpeng
    Archambault, Yves
    Bocanek, Jim
    Tang, Grace
    LoSasso, Thomas
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (02): : 575 - 581
  • [29] Comprehensive Quality Assurance Procedures for Uniform Scanning Proton Therapy Machines
    Zheng, Y.
    Ding, X.
    Mascia, A.
    Hsi, W.
    Kang, Y.
    Ramirez, E.
    Zeidan, O.
    Schreuder, N.
    MEDICAL PHYSICS, 2012, 39 (06) : 3749 - 3749
  • [30] Quality assurance and validation with immunoassays
    Schneider, Rudolf J.
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2006, 232 : 469 - 469