Influence of reconstruction techniques on PET/CT image quality and quantitative accuracy: a phantom study

被引:1
|
作者
Su, Xuesong [1 ]
Geng, Jianhua [1 ]
Liu, Jianing [1 ]
Liu, Fengshuo [1 ]
Wu, Yichen [1 ]
Zheng, Rong [1 ]
Wang, Xuejuan [1 ]
机构
[1] Chinese Acad Med Sci & Peking Union Med Coll, Canc Hosp, Dept Nucl Med, Natl Canc Ctr,Natl Clin Res Ctr Canc, Beijing 100021, Peoples R China
关键词
Positron emission tomography (PET); Image quality; Quantification accuracy; Reconstruction techniques; Phantom; TIME-OF-FLIGHT; POINT-SPREAD-FUNCTION; PHYSICAL FIGURES; PERFORMANCE; RESOLUTION; IMPACT; ALGORITHMS; NOISE; MERIT; OSEM;
D O I
10.1007/s41605-023-00441-w
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
PurposeTo investigate the influence of different reconstruction techniques on the quantitative accuracy and image quality of PET/CT.MethodsThe NEMA NU2-2018 image quality phantom was scanned on a GE Discovery Elite PET/CT scanner and the spatial resolution was measured based on NEMA NU2 standard. The same raw data were reconstructed using five different algorithms: ordered subset expectation maximization (OSEM), OSEM with point spread function (PSF) modeling, OSEM with time-of-flight (TOF), OSEM with TOF and PSF, and filtered back-projection (FBP). The recovery coefficient (RC), contrast recovery coefficient (CRC), and contrast-to-noise ratio (CNR) were calculated for the six hot spheres, the percent background variability (PBV) and coefficient of variation (COV) were calculated for the background, and the residual error (RE) was calculated for lung insert in different image slices.ResultsThe incorporation of PSF modeling showed the smallest transverse FWHM and FWTM at both 1 and 10 cm radical offsets. The combination of PSF modeling and TOF improved RCmean and CRC for all spheres and resulted in the highest ratings for the detectability of 10 mm spheres in human observer assessment. PSF modeling played a role in reducing the COV within the background region of interest and increasing the CNR of the spheres, and decreased background noise ratings in human observer assessment. Besides, TOF significantly reduced the RE in lung insert. Neither PSF modeling nor TOF had a significant effect on PBV. Compared to FBP, the OSEM algorithm showed significant advantages in PBV, COV, CNR and RE and human observer ratings of image quality, but worse results for RCmax, RCmean and CRC.ConclusionsThe integration of TOF and PSF modeling into the OSEM algorithm achieves improvements in both quantitative accuracy and image quality, providing distinct advantages. PSF modeling improves the spatial resolution and decreases the visual appearance of background noise. The OSEM algorithm shows significantly better image quality than the FBP algorithm but no distinct advantages concerning quantitative accuracy.
引用
收藏
页码:1171 / 1186
页数:16
相关论文
共 50 条
  • [31] Deep Learning Reconstruction at CT: Phantom Study of the Image Characteristics
    Higaki, Toru
    Nakamura, Yuko
    Zhou, Jian
    Yu, Zhou
    Nemoto, Takuya
    Tatsugami, Fuminari
    Awai, Kazuo
    ACADEMIC RADIOLOGY, 2020, 27 (01) : 82 - 87
  • [32] Quantitative Bias in PET/MR from Attenuation Correction and Reconstruction: a Comparison with PET and PET/CT with an Anatomical Brain Phantom and Hoffman Brain Phantom
    Teuho, Jarmo
    Johansson, Jarkko
    Linden, Jani
    Saunavaara, Virva
    Tolvanen, Tuula
    Teras, Mika
    2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,
  • [33] Quantitative accuracy of PET/CT for image-based kinetic analysis
    Seo, Youngho
    Teo, Boon-Keng
    Hadi, Mohiuddin
    Schreck, Carole
    Bacharach, Stephen L.
    Hasegawa, Bruce H.
    MEDICAL PHYSICS, 2008, 35 (07) : 3086 - 3089
  • [34] A NIM PET/CT phantom for evaluating the PET image quality of micro-lesions and the performance parameters of CT
    Shujie Lu
    Peng Zhang
    Chengwei Li
    Jie Sun
    Wenli Liu
    Pu Zhang
    BMC Medical Imaging, 21
  • [35] A NIM PET/CT phantom for evaluating the PET image quality of micro-lesions and the performance parameters of CT
    Lu, Shujie
    Zhang, Peng
    Li, Chengwei
    Sun, Jie
    Liu, Wenli
    Zhang, Pu
    BMC MEDICAL IMAGING, 2021, 21 (01)
  • [36] CT coronary angiography: Influence of different cardiac reconstruction intervals on image quality and diagnostic accuracy
    Dewey, Marc
    Teige, Florian
    Rutsch, Wolfgang
    Schink, Tania
    Hamm, Bernd
    EUROPEAN JOURNAL OF RADIOLOGY, 2008, 67 (01) : 92 - 99
  • [37] Optimization of high definition PET/CT image reconstruction: A phantom validation for improved quantification
    Binzel, Katherine
    Zhang, Jun
    Knopp, Michael
    JOURNAL OF NUCLEAR MEDICINE, 2017, 58
  • [38] Evaluation of radiation dose and image quality of CT scan for whole-body pediatric PET/CT: A phantom study
    Yang, Ching-Ching
    Liu, Shu-Hsin
    Mok, Greta S. P.
    Wu, Tung-Hsin
    MEDICAL PHYSICS, 2014, 41 (09)
  • [39] Iterative Reconstruction Algorithm for Abdominal Multidetector CT at Different Tube Voltages: Assessment of Diagnostic Accuracy, Image Quality, and Radiation Dose in a Phantom Study
    Schindera, Sebastian T.
    Diedrichsen, Lars
    Mueller, Hubert C.
    Rusch, Oliver
    Marin, Daniele
    Schmidt, Bernhard
    Raupach, Rainer
    Vock, Peter
    Szucs-Farkas, Zsolt
    RADIOLOGY, 2011, 260 (02) : 454 - 462
  • [40] A New Iodinated Liver Phantom for the Quantitative Evaluation of Advanced CT Acquisition and Reconstruction Techniques
    Chen, Baiyu
    Marin, Daniele
    Samei, Ehsan
    MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING, 2011, 7961