Influence of reconstruction techniques on PET/CT image quality and quantitative accuracy: a phantom study

被引:1
|
作者
Su, Xuesong [1 ]
Geng, Jianhua [1 ]
Liu, Jianing [1 ]
Liu, Fengshuo [1 ]
Wu, Yichen [1 ]
Zheng, Rong [1 ]
Wang, Xuejuan [1 ]
机构
[1] Chinese Acad Med Sci & Peking Union Med Coll, Canc Hosp, Dept Nucl Med, Natl Canc Ctr,Natl Clin Res Ctr Canc, Beijing 100021, Peoples R China
关键词
Positron emission tomography (PET); Image quality; Quantification accuracy; Reconstruction techniques; Phantom; TIME-OF-FLIGHT; POINT-SPREAD-FUNCTION; PHYSICAL FIGURES; PERFORMANCE; RESOLUTION; IMPACT; ALGORITHMS; NOISE; MERIT; OSEM;
D O I
10.1007/s41605-023-00441-w
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
PurposeTo investigate the influence of different reconstruction techniques on the quantitative accuracy and image quality of PET/CT.MethodsThe NEMA NU2-2018 image quality phantom was scanned on a GE Discovery Elite PET/CT scanner and the spatial resolution was measured based on NEMA NU2 standard. The same raw data were reconstructed using five different algorithms: ordered subset expectation maximization (OSEM), OSEM with point spread function (PSF) modeling, OSEM with time-of-flight (TOF), OSEM with TOF and PSF, and filtered back-projection (FBP). The recovery coefficient (RC), contrast recovery coefficient (CRC), and contrast-to-noise ratio (CNR) were calculated for the six hot spheres, the percent background variability (PBV) and coefficient of variation (COV) were calculated for the background, and the residual error (RE) was calculated for lung insert in different image slices.ResultsThe incorporation of PSF modeling showed the smallest transverse FWHM and FWTM at both 1 and 10 cm radical offsets. The combination of PSF modeling and TOF improved RCmean and CRC for all spheres and resulted in the highest ratings for the detectability of 10 mm spheres in human observer assessment. PSF modeling played a role in reducing the COV within the background region of interest and increasing the CNR of the spheres, and decreased background noise ratings in human observer assessment. Besides, TOF significantly reduced the RE in lung insert. Neither PSF modeling nor TOF had a significant effect on PBV. Compared to FBP, the OSEM algorithm showed significant advantages in PBV, COV, CNR and RE and human observer ratings of image quality, but worse results for RCmax, RCmean and CRC.ConclusionsThe integration of TOF and PSF modeling into the OSEM algorithm achieves improvements in both quantitative accuracy and image quality, providing distinct advantages. PSF modeling improves the spatial resolution and decreases the visual appearance of background noise. The OSEM algorithm shows significantly better image quality than the FBP algorithm but no distinct advantages concerning quantitative accuracy.
引用
收藏
页码:1171 / 1186
页数:16
相关论文
共 50 条
  • [1] Image quality with iterative reconstruction techniques in CT of the lungs-A phantom study
    Andersen, Hilde Kjernlie
    Volgyes, David
    Martinsen, Anne Latrine Trgde
    EUROPEAN JOURNAL OF RADIOLOGY OPEN, 2018, 5 : 35 - 40
  • [2] Impact of BSREM reconstruction algorithm on image quality and quantification accuracy of PET-a phantom study
    Chen, Wei
    Liang, Ying
    JOURNAL OF NUCLEAR MEDICINE, 2021, 62
  • [3] 90Y PET/CT quantitative accuracy and image quality
    Siman, Wendy
    Mawlawi, Osama
    Kappadath, Srinivas
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [4] 90Y PET/CT quantitative accuracy and image quality
    Siman, Wendy
    Mawlawi, Osama
    Kappadath, Srinivas
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [5] Effect of iterative reconstruction techniques on image quality in low radiation dose chest CT: a phantom study
    Xu, Yan
    Zhang, Ting-ting
    Hu, Zhi-hai
    Li, Juan
    Hou, Hong-jun
    Xu, Zu-shan
    He, Wen
    DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY, 2019, 25 (06) : 442 - 450
  • [6] A Phantom Study of Regularized Image Reconstruction in PET
    Wilson, Joshua M.
    Ross, Steven G.
    Deller, Timothy
    Asma, Evren
    Manjeshwar, Ravindra
    Turkington, Timothy G.
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 3661 - 3665
  • [7] Usefulness of CT iterative reconstruction method for low-dose CT attenuation correction on the accuracy of PET image in obesity phantom study
    Matsutomo, Norikazu
    Onishi, Hideo
    Nagaki, Akio
    Matsumoto, Keiichi
    Sasaki, Masayuki
    JOURNAL OF NUCLEAR MEDICINE, 2013, 54
  • [8] Image quality and detectability in Siemens Biograph PET/MRI and PET/CT systems—a phantom study
    Silje Kjærnes Øen
    Lars Birger Aasheim
    Live Eikenes
    Anna Maria Karlberg
    EJNMMI Physics, 6
  • [9] Quantitative Comparison of Image Quality and Standardized Uptake Values for PET/CT and PET/MR - a Phantom Study with 18F-FDG
    Karlberg, A. M.
    Saether, O.
    Lohmann, K.
    Eikenes, L.
    Goa, P. E.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2014, 41 : S254 - S254
  • [10] PET NEMA IQ Phantom dataset: image reconstruction settings for quantitative PET imaging
    Vosoughi, Habibeh
    Hajizadeh, Mohsen
    Emami, Farshad
    Momennezhad, Mehdi
    Geramifar, Parham
    DATA IN BRIEF, 2021, 37