Does radiolucency really predict loose components in revision shoulder arthroplasty?

被引:2
|
作者
Streck, Laura E. [1 ]
Gaal, Chiara [1 ]
Gohlke, Frank [2 ]
Rudert, Maximilian [1 ]
List, Kilian [1 ]
机构
[1] Univ Wurzburg, Dept Orthoped Surg, Koenig Ludwig Haus,Brettreichstr 11, D-97074 Wurzburg, Germany
[2] Rhoen Clin, Dept Shoulder Surg, Bad Neustadt an der Saale, Germany
关键词
Radiolucent lines; Shoulder replacement; Arthroplasty revision; Radiograph; GLENOID COMPONENTS; FOLLOW-UP; REPLACEMENT; TOMOGRAPHY; COMPLICATIONS; PROSTHESIS; LINES; NEER;
D O I
10.1007/s00256-023-04330-7
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
ObjectiveThe number of shoulder arthroplasties is increasing along with the need for revision surgeries. Determining the stability of the implant is crucial in preoperative planning. This study aims to investigate whether radiolucent lines (RLL) in preoperative radiographs predict component loosening.Materials and methodsPreoperative radiographs of 93 cases in 88 patients who underwent shoulder arthroplasty revision were evaluated regarding the presence of RLL. Correlation analyses were performed for radiographic findings and demographic factors (age, gender, BMI, prior surgeries) compared to intraoperative findings.ResultsThe presence of RLL around the humeral component correlated with loosening (p < 0.001, Phi 0.511), and the distal zones 3 and 5 showed the strongest correlation (Phi 0.536). While RLL in only one zone did not predict loosening (p = 0.337), RLL present in two or more zones showed correlation with loosening (p < 0.001). Risk factors associated with loosening were a higher age at the time of revision surgery (p = 0.030) and the number of zones with RLL (p < 0.001). The glenoid component was loose in 39.0% of the cases; 5.5% of the glenoid components with RLL were stable. Nevertheless, the presence of RLL was highly associated with loosening (p < 0.001, Phi 0.603). A longer time between implantation and revision correlated with loosening of the glenoid component (p = 0.046).ConclusionWhile RLL do not predict loosening of the implant in general, occurrence in more than one zone correlates with loosening. If located in distal zones and with increasing number of zones with RLL, the correlation becomes even stronger and loosening is more likely.
引用
收藏
页码:1759 / 1765
页数:7
相关论文
共 50 条
  • [21] Revision surgery of reverse shoulder arthroplasty
    Boileau, Pascal
    Melis, Barbara
    Duperron, David
    Moineau, Gregory
    Rumian, Adam P.
    Han, Yung
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2013, 22 (10) : 1359 - 1370
  • [22] Humeral windows in revision shoulder arthroplasty
    Sperling, JW
    Cofield, RH
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2005, 14 (03) : 258 - 263
  • [23] Glenoid reconstruction in revision shoulder arthroplasty
    Gohlke, F.
    Werner, B.
    Wiese, I.
    OPERATIVE ORTHOPADIE UND TRAUMATOLOGIE, 2019, 31 (02): : 98 - 114
  • [24] Revision of failed total shoulder arthroplasty
    Petersen, SA
    Hawkins, RJ
    ORTHOPEDIC CLINICS OF NORTH AMERICA, 1998, 29 (03) : 519 - +
  • [25] Risk of Revision Shoulder Arthroplasty After Anatomic and Reverse Total Shoulder Arthroplasty
    Cutler, Holt S.
    DeClercq, Josh
    Ayers, Gregory D.
    Serbin, Philip
    Jain, Nitin
    Khazzam, Michael
    JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2023, 31 (01) : 17 - 25
  • [26] Prevalence of Shoulder Arthroplasty in the United States and the Increasing Burden of Revision Shoulder Arthroplasty
    Farley, Kevin X.
    Wilson, Jacob M.
    Kumar, Anjali
    Gottschalk, Michael B.
    Daly, Charles
    Sanchez-Sotelo, Joaquin
    Wagner, Eric R.
    JBJS OPEN ACCESS, 2021, 6 (03)
  • [27] Revision reverse shoulder arthroplasty for the management of baseplate failure: an analysis of 676 revision reverse shoulder arthroplasty procedures
    Ii, Mitchell P. John
    Mesa, Lazaro
    Simon, Peter
    Frankle, Mark A.
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2024, 33 (03) : 707 - 714
  • [28] Extent of vertical cementing as a predictive factor for radiolucency in revision total knee arthroplasty
    Du Hyun Ro
    Yool Cho
    Sahnghoon Lee
    Kee Yun Chung
    Seong Hwan Kim
    Young Min Lee
    Joon Kyu Lee
    Myung Chul Lee
    Knee Surgery, Sports Traumatology, Arthroscopy, 2016, 24 : 2710 - 2717
  • [29] Extent of vertical cementing as a predictive factor for radiolucency in revision total knee arthroplasty
    Ro, Du Hyun
    Cho, Yool
    Lee, Sahnghoon
    Chung, Kee Yun
    Kim, Seong Hwan
    Lee, Young Min
    Lee, Joon Kyu
    Lee, Myung Chul
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2016, 24 (08) : 2710 - 2717
  • [30] Preoperative prophylactic antibiotics administration does not influence culture yield in revision shoulder arthroplasty
    Stoll, Kurt
    Alfonsi III, Samuel
    Khan, Adam Z.
    Vaughan, Alayna
    Namdari, Surena
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2023, 32 (11) : 2366 - 2370