Three-Dimensional Bioelectrical Impedance Spectroscopic Tomography for Visualization of Breast Tumor

被引:5
|
作者
Deng, Qi [1 ]
He, Jie [1 ]
Zhu, Chengjun [2 ]
Wang, Hao [3 ]
Liu, Kai [1 ]
Sun, Bo [4 ]
Yao, Jiafeng [1 ]
机构
[1] Nanjing Univ Aeronaut & Astronaut, Coll Mech & Elect Engn, Nanjing 210016, Peoples R China
[2] Jiangsu Prov Hosp, Dept Oncol, Nanjing 210008, Peoples R China
[3] Jinan Univ, Affiliated Hosp 1, Dept Anesthesiol, Guangzhou 215004, Peoples R China
[4] Xian Univ Technol, Sch Mech & Precis Instrument Engn, Xian 710048, Peoples R China
基金
中国国家自然科学基金;
关键词
3-D bioelectrical impedance spectroscopic tomography (3D-BIST); 3-D visualization; breast tumor; conical sensor; identification of benign and malignant; RECONSTRUCTION; ALGORITHM; EIT;
D O I
10.1109/JSEN.2023.3294185
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Visualization of breast tumors with 3-D bioelectrical impedance spectroscopic tomography (3D-BIST) is proposed to identify the position, size, benignity, and malignancy of breast tumors. First, to visualize in 3-D, a conical sensor is developed based on the actual shape of the breast. Second, to confirm the feasibility of the proposed method, finite element simulation is employed. The influence of different impedance spectroscopy data acquisition methods (0-, 1-, and 3-skip) and different classifiers [Linear-SVM, radial basis function-support vector machines (RBF-SVM), and k-nearest neighbors (KNN)] on the classification results are explored in the simulation. The simulation results show that the 3D-BIST method allows visualization of breast tumor types; the image correlation coefficient (ICC) of imaging results at each position is larger than 0.7 when the breast tumor diameter d >= 6 mm; the best acquisition method for impedance spectroscopy data is 3-skip acquisition; and the best classifier is Linear-SVM classifier. The accuracy of the verification set calculated by fivefold cross-validation achieved 93.5%. The accuracy of identifying benign and malignant breast tumors with d >= 8 mm exceeded 89.0% on the test set. Finally, the proposed method was experimentally validated by using three kinds of biological tissues in place of normal breast tissue, benign breast tumor, and malignant breast tumor. In conclusion, the 3D-BIST method enables 3-D visualization of breast tumors and identification of breast tumor types. The proposed method has the advantages of noninvasive and non-radiation, which is expected to be applied to the detection of breast tumors.
引用
收藏
页码:19598 / 19605
页数:8
相关论文
共 50 条
  • [1] A Sensor for Detection of Breast Tumor with Three-dimensional Electrical Impedance Tomography
    Kai, Liu
    An-Qi, Li
    Li, Fang
    Cheng-Jun, Zhu
    Hang, Tian
    Jia-Feng, Yao
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2024, 52 (02) : 248 - 255
  • [2] Three-dimensional electrical impedance tomography
    Metherall, P
    Barber, DC
    Smallwood, RH
    Brown, BH
    NATURE, 1996, 380 (6574) : 509 - 512
  • [3] Three-dimensional visualization of breast cancer
    Maidment, ADA
    Albert, M
    Conant, EF
    Feig, SA
    DIGITAL MAMMOGRAPHY, 1998, 13 : 57 - 60
  • [4] Three-dimensional reconstruction in electrical impedance tomography
    Liu, W.P.
    Hua, P.
    Webster, J.G.
    Clinical Physics and Physiological Measurement, 1988, 9 (SUPPL. A): : 131 - 135
  • [5] Static three-dimensional electrical impedance tomography
    Vauhkonen, PJ
    Vauhkonen, H
    Savolainen, T
    Kaipio, JP
    ELECTRICAL BIOIMPEDANCE METHODS: APPLICATIONS TO MEDICINE AND BIOTECHNOLOGY, 1999, 873 : 472 - 481
  • [6] Three-Dimensional Visualization of Whole Synapses by Stem Tomography
    Sousa, Alioscka A.
    Zhang, Jun
    Chen, Xiaobing
    Reese, Thomas S.
    Diamond, Jeffrey S.
    Leapman, Richard D.
    BIOPHYSICAL JOURNAL, 2013, 104 (02) : 353A - 353A
  • [7] Excitation patterns in three-dimensional electrical impedance tomography
    Dehghani, H
    Soni, N
    Halter, R
    Hartov, A
    Paulsen, KD
    PHYSIOLOGICAL MEASUREMENT, 2005, 26 (02) : S185 - S197
  • [8] Three-Dimensional Electrical Impedance Tomography With Multiplicative Regularization
    Zhang, Ke
    Li, Maokun
    Yang, Fan
    Xu, Shenheng
    Abubakar, Aria
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2019, 66 (09) : 2470 - 2480
  • [9] Three-dimensional image reconstruction for electrical impedance tomography
    Kleinermann, F
    Avis, NJ
    Judah, SK
    Barber, DC
    PHYSIOLOGICAL MEASUREMENT, 1996, 17 : A77 - A83
  • [10] Visualization of three-dimensional nephron structure with microcomputed tomography
    Bentley, Michael D.
    Jorgensen, Steven M.
    Lerman, Lilach O.
    Ritman, Erik L.
    Romero, J. Carlos
    ANATOMICAL RECORD-ADVANCES IN INTEGRATIVE ANATOMY AND EVOLUTIONARY BIOLOGY, 2007, 290 (03): : 277 - 283