Comparison of the Sensitive Volume Response of the Three Different Cylindrical Ionization Chambers with GEANT4 Simulations Using the Concept of Output Factor

被引:0
|
作者
Ayrancioglu, Oguzhan [1 ,2 ,5 ]
Sisman, Gizem [2 ]
Kandemir, Recep [2 ]
Ayrancioglu, Ceren [2 ,3 ]
Akgungor, Kadir [4 ]
Alicikus, Lutfiye Zumre A. R. I. C. A. N. [1 ]
机构
[1] Izmir Tınaztepe Univ, Galen Hosp, Dept Radiat Oncol, Izmir, Turkiye
[2] Dokuz Eylul Univ, Inst Hlth Sci, Dept Med Phys, Izmir, Turkiye
[3] Izmir Tinaztepe Univ, Vocat Sch Hlth Serv, Izmir, Turkiye
[4] Dokuz Eylul Univ, Fac Sci, Dept Atom & Mol Phys, Izmir, Turkiye
[5] Izmir Tınaztepe Univ, Galen Hastanesi, Radyasyon Onkolojisi Klin, Izmir, Turkiye
来源
关键词
Detector sensitive volume; GEANT4; Monte Carlo; output factor; MONTE-CARLO SIMULATIONS; BEAMS;
D O I
10.5505/tjo.2023.4183
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
This study aims to contribute to the literature the broadened demonstration of energy-related and field size-related responses of three different cylindrical ionization chambers, commonly used for clinical applications of the therapeutic photon beams, using GEANT4 code. METHODS GEANT4 simulation of Varian TrueBeam STx was compared with the real-time measurements. The variations in the output factor (OF) measurements depending on field sizes ranging from 0.6x0.6 cm2 to 30x30 cm2 and photon energies of 6MV, 6MV-FFF, 10MV, 10MV-FFF, and 15MV were recorded for each chamber and compared with the GEANT4 simulations. Response of the chambers was analyzed depending on those OFs. RESULTS The GEANT4 simulation model was validated. All the chambers were in a good agreement within 1.5% for field sizes between 4x4 cm2 and 30x30 cm2. The farmer chamber presented a major deviation up to 22% for the field sizes smaller than 4x4 cm2. Semiflex and pinpoint chambers have an agreement within 2% and 5.5% for 3x3cm2 and 2x2 cm2, respectively. CONCLUSION The chambers evaluated in this study can be used reliably for all photon energies and field sizes between 4x4 cm2 and 30x30 cm2. To obtain accurate measurements for field sizes smaller than 3x3 cm2, it is recommended to use detectors with a smaller sensitive volume instead of the detectors used in this study.
引用
收藏
页码:27 / 38
页数:12
相关论文
共 18 条
  • [1] The influence of different versions of FLUKA and GEANT4 on the calculation of response functions of ionization chambers in clinical proton beams
    Baumann, Kilian-Simon
    Derksen, Larissa
    Witt, Matthias
    Adeberg, Sebastian
    Zink, Klemens
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (24):
  • [2] Assessment of a new multileaf collimator concept using GEANT4 Monte Carlo simulations
    Tacke, MB
    Szymanowski, H
    Oelfke, U
    Schulze, C
    Nuss, S
    Wehrwein, E
    Leidenberger, S
    MEDICAL PHYSICS, 2006, 33 (04) : 1125 - 1132
  • [3] Peak Efficiency of NaI Detector and Coincidence Summing Factor for Different Cylindrical Sources Using Geant4 Simulation
    Alzahrani, Jamila S.
    Almuqrin, Aljawhara H.
    HEALTH PHYSICS, 2021, 121 (03): : 202 - 208
  • [4] GEANT4 SIMULATIONS RESPONSE OF CYLINDRICAL SILICON-ON-INSULATOR MICRODOSIMETER ARRAY DETECTOR FOR NEUTRONS DETECTION
    Lei Ming
    Liu Shu-Huan
    He Chao-Hui
    Xiong Cen
    PROCEEDINGS OF THE 21ST INTERNATIONAL CONFERENCE ON NUCLEAR ENGINEERING - 2013, VOL 1, 2014,
  • [5] Comparison of scattering experiments using synchrotron radiation with Monte Carlo simulations using Geant4
    Gerlach, M.
    Krumrey, M.
    Cibik, L.
    Mueller, P.
    Ulm, G.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2009, 608 (02): : 339 - 343
  • [6] COMPARISON OF BNCT DOSIMETRY CALCULATIONS USING DIFFERENT GEANT4 PHYSICS LISTS
    Chen, Zhao
    Yang, Peng
    Lei, Qin
    Wen, Yumei
    He, Donglin
    Wu, Zhangwen
    Gou, Chengjun
    RADIATION PROTECTION DOSIMETRY, 2019, 187 (01) : 88 - 97
  • [7] Monte Carlo Calculation of Perturbation Correction Factors for AirFilled Ionization Chambers in Clinical Proton Beams Using TOPAS/GEANT4
    Baumann, K.
    Kaupa, S.
    Engenhart-Cabillic, R.
    Zink, K.
    MEDICAL PHYSICS, 2020, 47 (06) : E281 - E282
  • [8] BETA-EFFICIENCY OF A TYPICAL GAS-FLOW IONIZATION CHAMBER USING GEANT4 MONTE CARLO SIMULATIONS
    Hussain, Abid
    Mirza, Sikander M.
    Mirza, Nasir M.
    Siddique, Muhammad T.
    NUCLEAR TECHNOLOGY & RADIATION PROTECTION, 2011, 26 (03): : 193 - 200
  • [9] The energy response of a TPA Mk-II ionization chamber using GEANT4 Monte Carlo simulation
    Seneviratne, M. D. S.
    Reinhard, M. I.
    Baldock, C.
    PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (13): : 3837 - 3846
  • [10] Dosimetric comparison between different radiotherapy protocols for prostate cancer using Geant4 Monte Carlo simulation
    Syahir, M. K.
    Fahmi, M. R.
    Hashikin, N. A. A.
    11TH INTERNATIONAL SEMINAR ON MEDICAL PHYSICS (ISMP) 2019, 2020, 1497