Comparative Evaluation of Cone-Beam Consistency Conditions for Mono-Material Beam Hardening Correction in Flat Detector Computed Tomography

被引:0
|
作者
Abdurahman, Shiras [1 ]
Frysch, Robert [1 ]
Pfeiffer, Tim [1 ]
Beuing, Oliver [2 ]
Rose, Georg [1 ]
机构
[1] Otto von Guericke Univ, Inst Med Engn & Res Campus STIMULATE, D-39106 Magdeburg, Germany
[2] AMEOS Klinikum Bernburg, Dept Intervent Neuroradiol, D-06406 Bernburg, Germany
关键词
Detectors; Three-dimensional displays; Computed tomography; Image reconstruction; Geometry; Transforms; Trajectory; Beam hardening correction; cone-beam computed tomography (CT); consistency conditions; RADON-TRANSFORM; RECONSTRUCTION; PROJECTIONS; CT;
D O I
10.1109/TRPMS.2022.3228694
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Consistency conditions have been successfully utilized for data-driven artifact reductions in cone-beam computed tomography systems equipped with a large-area flat-panel detector. Recently, many formulations and applications of pairwise cone-beam consistency conditions have been published, including the Grangeat consistency condition (GCC), Smith consistency condition (SCC), and fan-beam consistency condition (FBCC). Previous works demonstrated that the polynomial coefficients for beam hardening correction could be directly computed from cone-beam raw data by enforcing consistency conditions on projection pairs. This article compares the effectiveness of pairwise consistency conditions for mono-material beam hardening correction using a second-degree polynomial. The results from our studies show that similar corrections could be achieved for ideal polychromatic projections. We also investigated the effectiveness of corrections after perturbing the projections with an increasing degree of errors other than those caused by beam hardening. The studies indicate the superior robustness of FBCCs toward Poisson noise, axial truncation, detector shift, and scatter, while GCCs were less vulnerable to projection intensity errors. The optimal choice of consistency conditions depends on the CBCT system geometry, physical phenomena other than beam hardening, and the availability and accuracy of preprocessing and artifact corrections algorithms before beam hardening correction.
引用
收藏
页码:354 / 371
页数:18
相关论文
共 50 条
  • [41] A moving blocker system for cone-beam computed tomography scatter correction
    Ouyang, Luo
    Song, Kwang
    Solberg, Timothy
    Wang, Jing
    MEDICAL IMAGING 2013: PHYSICS OF MEDICAL IMAGING, 2013, 8668
  • [42] Simultaneous misalignment correction for approximate circular cone-beam computed tomography
    Kyriakou, Y.
    Lapp, R. M.
    Hillebrand, L.
    Ertel, D.
    Kalender, W. A.
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (22): : 6267 - 6289
  • [44] Scatter Correction Based on Beam Stop Array for Cone-Beam Micro-Computed Tomography
    Tang Tianxu
    Duan Xiaojiao
    Zhou Zhizheng
    Wu Qi
    ACTA OPTICA SINICA, 2019, 39 (08)
  • [45] Evaluating the mandible with cone-beam computed tomography
    Stratemann, Scott A.
    Huang, John C.
    Maki, Koutaro
    Hatcher, David C.
    Miller, Arthur J.
    AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2010, 137 (04) : S58 - S70
  • [46] Simple ROI Cone-Beam Computed Tomography
    Maass, Clemens
    Knaup, Michael
    Sawall, Stefan
    Kachelriess, Marc
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 3161 - 3168
  • [47] Operational principles for cone-beam computed tomography
    Hatcher, David C.
    JOURNAL OF THE AMERICAN DENTAL ASSOCIATION, 2010, 141 (10): : 3S - 6S
  • [48] Cone-Beam Computed Tomography: Beyond Diagnostics
    Perel, Morton L.
    IMPLANT DENTISTRY, 2015, 24 (04) : 367 - 367
  • [49] Investigation of the Lag Effect in X-Ray Flat-Panel Detector for Cone-Beam Computed Tomography
    Avakyan, A. K.
    Dergacheva, I. L.
    Elanchik, A. A.
    Krylova, T. A.
    Lobzhanidze, T. K.
    Polikhov, S. A.
    Smirnov, V. P.
    BIOMEDICAL ENGINEERING-MEDITSINSKAYA TEKNIKA, 2020, 54 (03): : 179 - 183
  • [50] Investigation of the Lag Effect in X-Ray Flat-Panel Detector for Cone-Beam Computed Tomography
    A. K. Avakyan
    I. L. Dergacheva
    A. A. Elanchik
    T. A. Krylova
    T. K. Lobzhanidze
    S. A. Polikhov
    V. P. Smirnov
    Biomedical Engineering, 2020, 54 : 179 - 183