X-ray fluorescence analysis using ultrasonic levitation technique

被引:0
|
作者
Okuda, Masaki [1 ,2 ]
Matsuyama, Tsugufumi [1 ]
Tsuji, Kouichi [1 ,2 ]
机构
[1] Osaka Metropolitan Univ OMU, Grad Sch Engn, Dept Chem & Bioengn, Div Sci & Engn Mat, Osaka, Japan
[2] Osaka Metropolitan Univ OMU, Grad Sch Engn, Dept Chem & Bioengn, Div Sci & Engn Mat, 3-3-138 Sugimoto, Sumiyoshi, Osaka 5588585, Japan
关键词
trace elements analysis without sample holder; ultrasonic levitation; X-ray fluorescence analysis; SAMPLES; LIQUID;
D O I
10.1002/xrs.3337
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
A thin film is used as the sample holder for analyzing solutions and solid samples using x-ray fluorescence (XRF). The incident x-rays are scattered on the sample and thin film, and the scattered x-rays serve as background signals in the XRF spectra. To reduce the intensity of the background signal, an ultrasonic levitation method was developed herein for XRF measurements. The ultrasonic levitation device is equipped with a transducer (horn) and reflector, and a small-size sample is levitated around a node of a generated standing wave; thus, a sample holder is not necessary. Fe solution (2 mu L) was analyzed to evaluate the developed instrument. The results obtained using levitation times of 0 and 60 min were compared. The solvent droplet almost evaporated during ultrasonic levitation; therefore, the absorption effects of XRFs and incident x-rays and the scattering effect of incident x-rays were reduced. At a levitation time of 60 min, the net intensity of the Fe K alpha peak was increased, and the background intensity decreased. The relationship between the levitation time and the net or background intensities of the Fe K alpha peak was defined; thus, condensation of the sample droplet was successfully observed. Quantitative analysis was performed using an internal standard method. The relative intensity (ratio of the net intensities of the target and internal standard elements) was proportional to the concentration ratio. Therefore, this method is useful for measuring trace elements at low background intensity and elucidating the process of condensation of a sample droplet.
引用
收藏
页码:364 / 370
页数:7
相关论文
共 50 条
  • [21] A robust X-ray fluorescence technique for multielemental analysis of solid samples
    Kallithrakas-Kontos, Nikolaos
    Foteinis, Spyros
    Paigniotaki, Katherine
    Papadogiannakis, Minos
    ENVIRONMENTAL MONITORING AND ASSESSMENT, 2016, 188 (02) : 1 - 10
  • [22] A robust X-ray fluorescence technique for multielemental analysis of solid samples
    Nikolaos Kallithrakas-Kontos
    Spyros Foteinis
    Katherine Paigniotaki
    Minos Papadogiannakis
    Environmental Monitoring and Assessment, 2016, 188
  • [24] Analysis of Chemical Features of a Soil Used as Landfill: Using the X-Ray Fluorescence (XRF) Technique
    de Borba, Willian Fernando
    Silverio da Silva, Jose Luiz
    Cunha Kemerich, Pedro Daniel
    Boito de Souza, Ericklis Edson
    Fernandes, Gabriel D'avila
    Carvalho, Ivan Ricardo
    WATER AIR AND SOIL POLLUTION, 2020, 231 (06):
  • [25] Analysis of Chemical Features of a Soil Used as Landfill: Using the X-Ray Fluorescence (XRF) Technique
    Willian Fernando de Borba
    José Luiz Silvério da Silva
    Pedro Daniel da Cunha Kemerich
    Éricklis Edson Boito de Souza
    Gabriel D’ávila Fernandes
    Ivan Ricardo Carvalho
    Water, Air, & Soil Pollution, 2020, 231
  • [26] Energy dispersion X-ray fluorescence analysis of macroelements in soils - Using the reference element technique
    Savichev, AT
    Sorokin, SE
    EURASIAN SOIL SCIENCE, 2002, 35 (12) : 1282 - 1286
  • [27] Sample preparation for total reflection X-ray fluorescence analysis using resist pattern technique
    Tsuji, K.
    Yomogita, N.
    Konyuba, Y.
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2018, 144 : 68 - 74
  • [28] Using Organic Slab to Obtain X-ray Tube Spectra for Quantitative Analysis of X-ray Fluorescence Analysis
    Liu, Song
    Li, Qinghui
    ADVANCES IN X-RAY/EUV OPTICS AND COMPONENTS VIII, 2013, 8848
  • [29] Polarization X-Ray Fluorescence Analysis by Milliwatt X-ray Tube
    Kawai, Jun
    Kato, Toshihide
    Imashuku, Susumu
    TETSU TO HAGANE-JOURNAL OF THE IRON AND STEEL INSTITUTE OF JAPAN, 2024, 110 (12): : 981 - 987
  • [30] X-ray Peltier cooled detectors for X-ray fluorescence analysis
    Loupilov, A
    Sokolov, A
    Gostilo, V
    RADIATION PHYSICS AND CHEMISTRY, 2001, 61 (3-6) : 463 - 464