Performance Assessment of Sysmex DI-60: Is Digital Morphology Analyzer Reliable for White Blood Cell Differentials in Body Fluids?

被引:1
|
作者
Shin, Eunju [1 ]
Hur, Mina [1 ]
Kim, Hanah [1 ]
Lee, Gun-Hyuk [1 ]
Hong, Mi-Hyun [1 ]
Nam, Minjeong [2 ]
Lee, Seungho [3 ]
机构
[1] Konkuk Univ, Dept Lab Med, Sch Med, Seoul 05030, South Korea
[2] Korea Univ, Anam Hosp, Dept Lab Med, Seoul 02841, South Korea
[3] Dong A Univ, Coll Med, Dept Prevent Med, Busan 49315, South Korea
关键词
digital morphology analyzer; DI-60; body fluid; white blood cell differential; performance; turnaround time; IMAGE-ANALYSIS; MODE; RISK;
D O I
10.3390/diagnostics14060592
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Few studies have evaluated digital morphology (DM) analyzers on body fluids (BF). We evaluated the performance of a DM analyzer, Sysmex DI-60 (Sysmex, Kobe, Japan) for white blood cell (WBC) differentials in BF samples. Methods: In five BF samples (two pleural fluids and three ascites) containing a single, dominant cell type (>80%, neutrophils, lymphocytes, macrophages, abnormal lymphocytes, and malignant cells in each sample), we evaluated the precision of the DI-60 and compared the WBC differentials and turnaround times (TAT) between DI-60 and manual counting. Results: The precision of the DI-60 pre-classification and verification was excellent (%CV, 0.01-3.16%). After verification, the DI-60 showed high sensitivity, specificity, and efficiency (ranges: 90.8-98.1%, 96.8-97.9%, and 92.5-98.0%, respectively) for the dominant cell types in neutrophil- and lymphocyte-dominant samples. For all samples, the DI-60 and manual counting showed high correlations for major cell types (neutrophils, lymphocytes, macrophages, and others, r = 0.72 to 0.94) after verification. The agreement between the pre-classification and verification of the DI-60 was strong in the neutrophil-dominant sample (kappa = 0.81). The DI-60 showed a significantly longer TAT (min: s) than manual counting for all samples (median TAT/slide: 6:28 vs. 1:53, p < 0.0001), with remarkable differences in abnormal lymphocyte- and malignant cell-dominant samples (21:05 vs. 2:06; 12:34 vs. 2:25). Conclusions: The DI-60 may provide reliable data in neutrophil- and lymphocyte-dominant BF samples. However, it may require longer times and higher workloads for WBC differentials especially in BF samples containing atypical cells. Further improvement would be needed before applying DM analyzers for routine clinical practice in BF analysis.
引用
收藏
页数:15
相关论文
共 36 条
  • [21] Performance Evaluation of the Automated Cell Image Analyzer DI-60 for Leukocyte Differential Count
    Chang, Soon Hee
    CLINICAL LABORATORY, 2025, 71 (03) : 538 - 544
  • [22] Performance evaluation of the automated cell image analyzer DI-60 for leukocyte differential count
    Chang, S.
    Kim, E.
    Kim, Y.
    Won, D.
    CLINICAL CHEMISTRY, 2024, 70 : I143 - I143
  • [23] Evaluation of cell count and classification capabilities in body fluids using a fully automated Sysmex XN equipped with high-sensitive Analysis (hsA) mode and DI-60 hematology analyzer system
    Takemura, Hiroyuki
    Ai, Tomohiko
    Kimura, Konobu
    Nagasaka, Kaori
    Takahashi, Toshihiro
    Tsuchiya, Koji
    Yang, Haeun
    Konishi, Aya
    Uchihashi, Kinya
    Horii, Takashi
    Tabe, Yoko
    Ohsaka, Akimichi
    PLOS ONE, 2018, 13 (04):
  • [24] Parasitised red blood cells misclassified as giant platelets by an automated digital morphology analyser (Sysmex DI-60/CellaVision): a case report and a retrospective EQA analysis
    Rosetti, Marco
    Farneti, Giorgia
    Monti, Marta
    Torri, Arianna
    Poletti, Giovanni
    Massari, Evita
    Polli, Valentina
    Clementoni, Alice
    Dorizzi, Romolo M.
    BRITISH JOURNAL OF HAEMATOLOGY, 2021, 192 (03) : E90 - E92
  • [25] Performance comparison of two automated digital morphology analyzers for leukocyte differential in patients with malignant hematological diseases: Mindray MC-80 and Sysmex DI-60
    Ye, Xianfei
    Fang, Lijuan
    Chen, Yunying
    Tong, Jixiang
    Ning, Xiaoni
    Feng, Lanjun
    Xu, Yuting
    Yang, Dagan
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2024, 46 (03) : 457 - 465
  • [26] Performance evaluation of the Sysmex XN-1000 hematology analyzer in assessment of the white blood cell count differential in pediatric specimens
    Becker, P. -H.
    Fenneteau, O.
    Da Costa, L.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2016, 38 (01) : 54 - 63
  • [27] When dysplasia deceives instrumental results: Pseudo monocytosis on the sysmex XN haematological analyser and DI-60 digital morphology system: Cases report and literature review
    La Gioia, Antonio
    Basile, Manuela
    Fiorini, Fabiana
    Fiorini, Marcello
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2023, 45 (06) : 987 - 989
  • [28] PERFORMANCE EVALUATION OF THE HEMATOLOGY ANALYZER SYSMEX XE-5000 FOR WHITE BLOOD CELL ANALYSIS IN CEREBROSPINAL FLUID
    Perne, Andrea
    Hainfellner, Johannes A.
    Szekeres, Thomas
    Schwarzinger, Ilse
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2011, 49 : S770 - S770
  • [29] Performance Evaluation of the Sysmex XE-5000 Hematology Analyzer for White Blood Cell Analysis in Cerebrospinal Fluid
    Perne, Andrea
    Hainfellner, Johannes A.
    Womastek, Irene
    Haushofer, Alexander
    Szekeres, Thomas
    Schwarzinger, Ilse
    ARCHIVES OF PATHOLOGY & LABORATORY MEDICINE, 2012, 136 (02) : 194 - 198
  • [30] VALIDATION AND IMPLEMENTATION OF THE SYSMEX XN 2000 BODY FLUID MODULE FOR WHITE BLOOD CELL DIFFERENTIAL COUNT IN SEROUS FLUIDS
    Labaere, Delphine
    Boeckx, Nancy
    Geerts, Inge
    Moens, Marc
    Van den Driessche, Marleen
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2014, 36 : 85 - 85