Performance Assessment of Sysmex DI-60: Is Digital Morphology Analyzer Reliable for White Blood Cell Differentials in Body Fluids?

被引:1
|
作者
Shin, Eunju [1 ]
Hur, Mina [1 ]
Kim, Hanah [1 ]
Lee, Gun-Hyuk [1 ]
Hong, Mi-Hyun [1 ]
Nam, Minjeong [2 ]
Lee, Seungho [3 ]
机构
[1] Konkuk Univ, Dept Lab Med, Sch Med, Seoul 05030, South Korea
[2] Korea Univ, Anam Hosp, Dept Lab Med, Seoul 02841, South Korea
[3] Dong A Univ, Coll Med, Dept Prevent Med, Busan 49315, South Korea
关键词
digital morphology analyzer; DI-60; body fluid; white blood cell differential; performance; turnaround time; IMAGE-ANALYSIS; MODE; RISK;
D O I
10.3390/diagnostics14060592
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Few studies have evaluated digital morphology (DM) analyzers on body fluids (BF). We evaluated the performance of a DM analyzer, Sysmex DI-60 (Sysmex, Kobe, Japan) for white blood cell (WBC) differentials in BF samples. Methods: In five BF samples (two pleural fluids and three ascites) containing a single, dominant cell type (>80%, neutrophils, lymphocytes, macrophages, abnormal lymphocytes, and malignant cells in each sample), we evaluated the precision of the DI-60 and compared the WBC differentials and turnaround times (TAT) between DI-60 and manual counting. Results: The precision of the DI-60 pre-classification and verification was excellent (%CV, 0.01-3.16%). After verification, the DI-60 showed high sensitivity, specificity, and efficiency (ranges: 90.8-98.1%, 96.8-97.9%, and 92.5-98.0%, respectively) for the dominant cell types in neutrophil- and lymphocyte-dominant samples. For all samples, the DI-60 and manual counting showed high correlations for major cell types (neutrophils, lymphocytes, macrophages, and others, r = 0.72 to 0.94) after verification. The agreement between the pre-classification and verification of the DI-60 was strong in the neutrophil-dominant sample (kappa = 0.81). The DI-60 showed a significantly longer TAT (min: s) than manual counting for all samples (median TAT/slide: 6:28 vs. 1:53, p < 0.0001), with remarkable differences in abnormal lymphocyte- and malignant cell-dominant samples (21:05 vs. 2:06; 12:34 vs. 2:25). Conclusions: The DI-60 may provide reliable data in neutrophil- and lymphocyte-dominant BF samples. However, it may require longer times and higher workloads for WBC differentials especially in BF samples containing atypical cells. Further improvement would be needed before applying DM analyzers for routine clinical practice in BF analysis.
引用
收藏
页数:15
相关论文
共 36 条
  • [1] PERFORMANCE OF AUTOMATED DIGITAL CELL MORPHOLOGY ANALYZER SYSMEX DI-60
    Kim, Hanah
    Kim, Hyeong Nyeon
    Hur, Mina
    Kim, Seung Wan
    Moon, Hee-Won
    Yun, Yeo-Min
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2016, 38 : 54 - 54
  • [2] Performance evaluation of the digital morphology analyser Sysmex DI-60 for white blood cell differentials in abnormal samples
    Zhao, Yan
    Diao, Yingying
    Zheng, Jun
    Li, Xinyao
    Luan, Hong
    SCIENTIFIC REPORTS, 2024, 14 (01):
  • [3] Analytical validation of white blood cell differential and platelet assessment on the Sysmex DI-60 digital morphology analyzer
    Lapic, Ivana
    Milos, Marija
    Dorotic, Marija
    Drenski, Valentina
    Herak, Desiree Coen
    Rogic, Dunja
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2023, 45 (05) : 668 - 677
  • [4] Analytical performance of the digital morphology analyzer Sysmex DI-60 for body fluid cell differential counts
    Yoon, Sumi
    Kim, Hye Ryoun
    PLOS ONE, 2023, 18 (07):
  • [5] Digital Morphology Analyzer Sysmex DI-60 vs . Manual Counting for White Blood Cell Differentials in Leukopenic Samples: A Comparative Assessment of Risk and Turnaround Time
    Nam, Minjeong
    Yoon, Sumi
    Hur, Mina
    Lee, Gun Hyuk
    Kim, Hanah
    Park, Mikyoung
    Kim, Hyeong Nyeon
    ANNALS OF LABORATORY MEDICINE, 2022, 42 (04) : 398 - +
  • [6] Performance of automated digital cell imaging analyzer Sysmex DI-60
    Kim, Hyeong Nyeon
    Hur, Mina
    Kim, Hanah
    Kim, Seung Wan
    Moon, Hee-Won
    Yun, Yeo-Min
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2018, 56 (01) : 94 - 102
  • [7] Performance of the automated digital cell image analyzer UIMD PBIA in white blood cell classification: a comparative study with sysmex DI-60
    Kim, Hongkyung
    Kweon, Oh Joo
    Yoon, Sumi
    Lim, Yong Kwan
    Kim, Bohyun
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2025,
  • [8] Detection of Plasmodium falciparum using automated digital cell morphology analyzer Sysmex DI-60
    Park, Mikyoung
    Hur, Mina
    Kim, Hanah
    Kim, Hyeong Nyeon
    Kim, Seung Wan
    Moon, Hee-Won
    Yun, Yeo-Min
    Cheong, Hae Suk
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2018, 56 (12) : E284 - E287
  • [9] Red and white blood cell morphology characterization and hands-on time analysis by the digital cell imaging analyzer DI-60
    Kweon, Oh Joo
    Lim, Yong Kwan
    Lee, Mi-Kyung
    Kim, Hye Ryoun
    PLOS ONE, 2022, 17 (04):
  • [10] Performance evaluation of the digital cell imaging analyzer DI-60 integrated into the fully automated Sysmex XN hematology analyzer system
    Tabe, Yoko
    Yamamoto, Takemasa
    Maenou, Imiko
    Nakai, Rie
    Idei, Mayumi
    Horii, Takashi
    Miida, Takashi
    Ohsaka, Akimichi
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2015, 53 (02) : 281 - 289