共 50 条
In vitro temperature changes in the pulp chamber caused by laser and Quadwave LED-light curing units
被引:6
|作者:
Maucoski, Cristiane
[1
,2
]
Price, Richard Bengt
[1
]
Arrais, Cesar Augusto Galvao
[2
]
Sullivan, Braden
[1
]
机构:
[1] Dalhousie Univ, Dept Dent Clin Sci, Halifax, NS B3H 4R2, Canada
[2] Univ Estadual Ponta Grossa, Dept Restorat Dent, 4748 Gen Carlos Cavalcanti Ave, BR-84051130 Ponta Grossa, PR, Brazil
来源:
关键词:
Laser;
Light-emitting diode;
Dental curing lights;
Temperature;
ANESTHETIZED HUMAN PULP;
HEAT;
RISE;
POLYMERIZATION;
EXPOSURE;
MICROCIRCULATION;
INCREASE;
GINGIVA;
DENTIN;
DIODE;
D O I:
10.1007/s10266-022-00780-y
中图分类号:
R78 [口腔科学];
学科分类号:
1003 ;
摘要:
The study evaluated the pulp temperature (PT) increase in Class I and V preparations when exposed to the Monet Laser (for 1 and 3 s), the PinkWave (for 3 and 10 s), the Valo Grand (for 3 and 10 s), the PowerCure, (for 3 and 10 s) and the SmartLite Pro (for 10 s). Non-retentive Class I and Class V cavities were prepared in one molar fixed in an acrylic plate and positioned in a warm water bath. The PT baseline was kept at 32 & DEG;C to simulate physiological conditions. Two T-type thermocouples were inserted through the roots into the pulp chamber in two positions: close to the pulp horn and the buccal wall close to the Class V cavity. The water flow was adjusted to 0.026 mL/min, and real-time temperature data were collected every 0.5 s. PT measurements were made with the tip of the LCU 0 and 6 mm away from the tooth surface. The radiant exitance (mW/cm(2)) and radiant exposure (J/cm(2)) were calculated. One-way ANOVA compared the effect of the pulpal flow, and & UDelta;T values were subjected to two-way ANOVA, followed by Scheffe's post hoc tests. The Monet Laser used for 3 s and the PinkWave used for 10 s produced the greatest PT rise in the Class I cavity. The simulated pulpal flow did not influence the PT rise. Overall, cavities exposed at the 0 mm distance had higher & UDelta;T values than groups at 6 mm distance. The placement of a rubber dam for Class V restorations may prevent positioning LCUs directly over the cavity, which may affect the rise in PT.
引用
收藏
页码:668 / 679
页数:12
相关论文