In vitro temperature changes in the pulp chamber caused by laser and Quadwave LED-light curing units

被引:6
|
作者
Maucoski, Cristiane [1 ,2 ]
Price, Richard Bengt [1 ]
Arrais, Cesar Augusto Galvao [2 ]
Sullivan, Braden [1 ]
机构
[1] Dalhousie Univ, Dept Dent Clin Sci, Halifax, NS B3H 4R2, Canada
[2] Univ Estadual Ponta Grossa, Dept Restorat Dent, 4748 Gen Carlos Cavalcanti Ave, BR-84051130 Ponta Grossa, PR, Brazil
关键词
Laser; Light-emitting diode; Dental curing lights; Temperature; ANESTHETIZED HUMAN PULP; HEAT; RISE; POLYMERIZATION; EXPOSURE; MICROCIRCULATION; INCREASE; GINGIVA; DENTIN; DIODE;
D O I
10.1007/s10266-022-00780-y
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The study evaluated the pulp temperature (PT) increase in Class I and V preparations when exposed to the Monet Laser (for 1 and 3 s), the PinkWave (for 3 and 10 s), the Valo Grand (for 3 and 10 s), the PowerCure, (for 3 and 10 s) and the SmartLite Pro (for 10 s). Non-retentive Class I and Class V cavities were prepared in one molar fixed in an acrylic plate and positioned in a warm water bath. The PT baseline was kept at 32 & DEG;C to simulate physiological conditions. Two T-type thermocouples were inserted through the roots into the pulp chamber in two positions: close to the pulp horn and the buccal wall close to the Class V cavity. The water flow was adjusted to 0.026 mL/min, and real-time temperature data were collected every 0.5 s. PT measurements were made with the tip of the LCU 0 and 6 mm away from the tooth surface. The radiant exitance (mW/cm(2)) and radiant exposure (J/cm(2)) were calculated. One-way ANOVA compared the effect of the pulpal flow, and & UDelta;T values were subjected to two-way ANOVA, followed by Scheffe's post hoc tests. The Monet Laser used for 3 s and the PinkWave used for 10 s produced the greatest PT rise in the Class I cavity. The simulated pulpal flow did not influence the PT rise. Overall, cavities exposed at the 0 mm distance had higher & UDelta;T values than groups at 6 mm distance. The placement of a rubber dam for Class V restorations may prevent positioning LCUs directly over the cavity, which may affect the rise in PT.
引用
收藏
页码:668 / 679
页数:12
相关论文
共 50 条
  • [1] In vitro temperature changes in the pulp chamber caused by laser and Quadwave LED-light curing units
    Cristiane Maucoski
    Richard Bengt Price
    Cesar Augusto Galvão Arrais
    Braden Sullivan
    Odontology, 2023, 111 : 668 - 679
  • [2] Comparison of temperature changes in the pulp chamber induced by various light curing units, in vitro
    Yazici, AR
    Müftü, A
    Kugel, G
    Perry, RD
    OPERATIVE DENTISTRY, 2006, 31 (02) : 261 - 265
  • [3] Pulp Chamber Temperature Increase from Curing Light Units: An In Vitro Study
    Cantekin, Kenan
    Buyuk, Suleyman Kutalmis
    Delikan, Ebru
    Peduk, Kubra
    Demirbuga, Sezer
    JOURNAL OF DENTISTRY FOR CHILDREN, 2014, 81 (03): : 128 - 132
  • [4] Comparison of the Amount of Temperature Rise in the Pulp Chamber of Teeth Treated With QTH, Second and Third Generation LED Light Curing Units: An In Vitro Study
    Mahant, Rajesh Harivadanbhai
    Chokshi, Shraddha
    Vaidya, Rupal
    Patel, Pruthvi
    Vora, Asima
    Mahant, Priyanka
    JOURNAL OF LASERS IN MEDICAL SCIENCES, 2016, 7 (03) : 184 - 191
  • [5] Temperature changes caused by light curing units on dentine of primary teeth
    Melara Munguia, A.
    Arregui Gambus, M.
    Guinot Jimeno, F.
    Bellet Dalmau, L. J.
    EUROPEAN JOURNAL OF PAEDIATRIC DENTISTRY, 2011, 12 (01) : 7 - 12
  • [6] Controlling In Vivo, Human Pulp Temperature Rise Caused by LED Curing Light Exposure
    Zarpellon, D. C.
    Runnacles, P.
    Maucoski, C.
    Coelho, U.
    Rueggeberg, F. A.
    Arrais, C. A. G.
    OPERATIVE DENTISTRY, 2019, 44 (03) : 235 - 241
  • [7] Evaluation of temperature changes in the pulp chamber during polymerization of light-cured pulp-capping materials by using a VALO LED light curing unit at different curing distances
    Savas, Selcuk
    Botsali, Murat S.
    Kucukyilmaz, Ebru
    Sari, Tugrul
    DENTAL MATERIALS JOURNAL, 2014, 33 (06) : 764 - 769
  • [8] Laser and curing light induced in vitro pulpal temperature changes
    Powell, GL
    Anderson, JR
    Blankenau, RJ
    JOURNAL OF DENTAL RESEARCH, 1997, 76 : 526 - 526
  • [9] Laser and curing light induced in vitro pulpal temperature changes
    Powell, GL
    Anderson, JR
    Blankenau, RJ
    JOURNAL OF CLINICAL LASER MEDICINE & SURGERY, 1999, 17 (01): : 3 - 5
  • [10] Comparative evaluation of third-generation LED-light curing units and their effect on bulk-fill composites-An in vitro study
    Puri, Aakansha
    Prathap, M. S.
    Jayaprakash, K.
    Sreegowri
    Jose, Jeslee A.
    JOURNAL OF INTERNATIONAL ORAL HEALTH, 2024, 16 (02): : 166 - 173