Vancomycin selective electrode based on molecularly imprinted polymer

被引:1
|
作者
Diaz-Villavicencio, Linet [1 ]
Baeza-Fonte, Alen N. [2 ]
Luaces-Alberto, Markel D. [3 ]
Valdes-Gonzalez, Aristides C. [4 ]
Milian-Rodriguez, Yanio E. [2 ,5 ]
Lazo-Fraga, Ana R. [2 ]
机构
[1] Beekeeping Res Ctr, Havana, Cuba
[2] Univ Havana, Inst Sci & Technol Mat, Havana, Cuba
[3] Univ Havana, Fac Chem, Havana, Cuba
[4] Inst Sanitat & Epidemiol, Havana, Cuba
[5] North Catholic Univ, Fac Engn & Geol Sci, Dept Chem Engn & Environm, Antofagasta 1270709, Chile
关键词
Vancomycin; All-solid-state electrode; PVC membranes; Molecularly-imprinted-polymer; POTENTIOMETRIC SENSOR; ASCORBIC-ACID;
D O I
10.1007/s11696-023-03051-4
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Vancomycin (VCM) is an antibiotic used in the treatment of serious gram-positive infections. However, antibiotic contamination has induced the emergence of resistant microorganism, especially for VCM. Accordingly, detection and control of VCM in aqueous samples after the cleaning process of pharmaceutical equipment employed to prepare injections is very important. For this, routine analyses require simple, sensitive and low-cost analytical methods. In this work, a selective electrode type all-solid-state for the determination of VCM antibiotic was developed. A molecularly imprinted polymer (MIP) based on silicon alkoxysilanes was used as recognition element in a polyvinyl chloride membrane, with o-nitrophenyl octyl ether as a mediating solvent. The influence of the MIP particle size on sensor response was studied. The best results were obtained for the electrode containing particles smaller than 15 & mu;m. This sensor exhibited sensitivity of 55.5 mV dec-1, limit of detection of 6.6 & mu;M and a linear response starting on 10 & mu;M. An electrode response time of 10 s and a lifetime of 21 days were achieved. Selectivity study showed that antibiotics like cefotaxime, ciprofloxacin and amoxicillin were interfering compounds, while chloramphenicol, metronidazole and inorganic cations Ca2+, Mg2+, Na+ did not interfere.
引用
收藏
页码:165 / 172
页数:8
相关论文
共 50 条
  • [41] A Fluorimetric Method Based On Magnetic Molecularly Imprinted Polymer For Selective Detection Of Chlorferon
    Ilktac, Raif
    JOURNAL OF FLUORESCENCE, 2022, 32 (01) : 369 - 379
  • [42] A Fluorimetric Method Based On Magnetic Molecularly Imprinted Polymer For Selective Detection Of Chlorferon
    Raif İLKTAÇ
    Journal of Fluorescence, 2022, 32 : 369 - 379
  • [43] Reagentless Sensing of Vancomycin Using an Indium Tin Oxide Electrode Grafted with Molecularly Imprinted Polymer including Ferrocenyl Group
    Eguchi, Haruto
    Hatano, Akihiko
    Yoshimi, Yasuo
    SENSORS, 2021, 21 (24)
  • [44] An electrochemical sensor for the determination of phoxim based on a graphene modified electrode and molecularly imprinted polymer
    Tan, Xuecai
    Wu, Jiawen
    Hu, Qi
    Li, Xiaoyu
    Li, Pengfei
    Yu, Huicheng
    Li, Xiaoyan
    Lei, Fuhou
    ANALYTICAL METHODS, 2015, 7 (11) : 4786 - 4792
  • [45] Molecularly Imprinted Polymer-Based Electrode for Tannic Acid Detection in Black Tea
    Moulick, Madhurima
    Das, Debangana
    Nag, Shreya
    Tudu, Bipan
    Bandyopadhyay, Rajib
    Roy, Runu Banerjee
    IEEE SENSORS JOURNAL, 2023, 23 (06) : 5535 - 5542
  • [46] Electrochemical Sensor for Determination of Chlorpyrifos Based on Graphene Modified Electrode and Molecularly Imprinted Polymer
    Tan Xue-Cai
    Wu Jia-Wen
    Hu Qi
    Li Xiao-Yu
    Li Peng-Fei
    Yu Hui-Cheng
    Li Xiao-Yan
    Lei Fu-Hou
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2015, 43 (03) : 387 - 393
  • [47] Electrochemical sensor based on molecularly imprinted polymer embedded graphite electrode for detecting curcumin
    Ahmed A.H.M.T.
    Naskar H.
    Banerjee S.
    Ghatak B.
    Das N.
    Tudu B.
    Bandyopadhyay R.
    Sensors and Actuators A: Physical, 2022, 344
  • [48] Selective and sensitive voltammetric sensor for methocarbamol determination by molecularly imprinted polymer modified carbon paste electrode
    Payehghadr, Mahmood
    Taherkhani, Yeganeh
    Maleki, Abbas
    Nourifard, Farzaneh
    EURASIAN CHEMICAL COMMUNICATIONS, 2020, 2 (09): : 982 - 990
  • [49] Selective detection of salivary cortisol using screen-printed electrode coated with molecularly imprinted polymer
    Karthika, Palanisamy
    Shanmuganathan, Saravanakumar
    Subramanian, Viswanathan
    Delerue-Matos, Cristina
    TALANTA, 2024, 272
  • [50] Zwitterionic Molecularly Imprinted Hairy Cellulose Nanocrystals Enable Selective Vancomycin Removal
    Koshani, Roya
    Yeh, Shang-Lin
    Kheirabadi, Sina
    He, Zeming
    Park, Min Ju
    Leconey, Sean
    Read, Andrew F.
    Sheikhi, Amir
    LANGMUIR, 2025, 41 (13) : 8554 - 8564