Design of a piezoelectrically actuated hydrocephalus shunt valve

被引:0
|
作者
Salih, O. [1 ]
Messina, M. [2 ]
Al-Jumeily, D. [3 ]
机构
[1] Liverpool John Moores Univ, Engn & Technol Res Inst, Liverpool, Merseyside, England
[2] Liverpool John Moores Univ, Mech Engn Dept, Liverpool, Merseyside, England
[3] Liverpool John Moores Univ, Sch Comp Sci & Math, Liverpool, Merseyside, England
关键词
Actuation; Hydrocephalus; Mechatronic valve; Piezoelectric; Shunt; Ultra-sonic motor; CEREBROSPINAL-FLUID; PRESSURE;
D O I
10.1007/s11517-023-02822-1
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Hydrocephalus is currently managed by using traditional mechanical passive shunts. Due to their nature, these shunts have fundamental shortcomings including an increase in patient shunt dependency, absence of fault detection, and over drainage due to lack of shunt proactivity. There is a scientific consensus that the way forward to address these issues is through what is termed a smart shunt. The core component of this system is the mechatronic controllable valve. A design of a valve which utilises both the passive nature of the classical valves and the controllability feature of the fully automated valves is presented in this paper. The valve consists of a fluid compartment, a linear spring, and an ultrasonic piezoelectric element. The valve is designed to operate on a 5 V supply, to drain up to 300 mL/h, and it has an operational range between 10 and 20 mmHg. The design produced is considered feasible as it takes into consideration the multiple operation conditions associated with such an implanted system.
引用
收藏
页码:2281 / 2290
页数:10
相关论文
共 50 条
  • [1] Design of a piezoelectrically actuated hydrocephalus shunt valve
    O. Salih
    M. Messina
    D. Al-Jumeily
    Medical & Biological Engineering & Computing, 2023, 61 : 2281 - 2290
  • [2] Piezoelectrically actuated hydraulic valve design for high bandwidth and flow performance
    Branson, D. T.
    Wang, F. C.
    Johnston, D. N.
    Tilley, D. G.
    Bowen, C. R.
    Keogh, P. S.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART I-JOURNAL OF SYSTEMS AND CONTROL ENGINEERING, 2011, 225 (I3) : 345 - 359
  • [3] Simulation and verification of a piezoelectrically actuated diaphragm for check valve micropump design
    Kang, Jianke
    Auner, Gregory W.
    SENSORS AND ACTUATORS A-PHYSICAL, 2011, 167 (02) : 512 - 516
  • [4] DESIGN AND MODELING OF A PIEZOELECTRICALLY ACTUATED MICROVALVE
    Rakotondrabe, M.
    Ivan, I. A.
    Stihi, V.
    Noveanu, Simona
    Minca, Eugenia
    ROMANIAN JOURNAL OF PHYSICS, 2011, 56 (1-2): : 141 - 149
  • [5] Design of a Piezoelectrically Actuated Jumping Robot
    Thanhtam Ho
    Lee, Sangyoon
    ADVANCED MATERIALS AND PROCESSES, PTS 1-3, 2011, 311-313 : 2211 - 2214
  • [6] Characteristic studies of the piezoelectrically actuated micropump with check valve
    Cheng, Chiang-Ho
    Tseng, Yi-Pin
    MICROSYSTEM TECHNOLOGIES-MICRO-AND NANOSYSTEMS-INFORMATION STORAGE AND PROCESSING SYSTEMS, 2013, 19 (11): : 1707 - 1715
  • [7] Characteristic studies of the piezoelectrically actuated micropump with check valve
    Chiang-Ho Cheng
    Yi-Pin Tseng
    Microsystem Technologies, 2013, 19 : 1707 - 1715
  • [8] TESTING THE HYDROCEPHALUS SHUNT VALVE
    WATTS, C
    KEITH, HD
    CHILDS BRAIN, 1983, 10 (04): : 217 - 228
  • [9] The design and analysis of piezoelectrically actuated microfluidic systems
    Chiarot, P
    Sullivan, P
    Ben Mrad, R
    MICRO-ELECTRO-MECHANICAL SYSTEMS (MEMS) - 2003, 2003, : 401 - 409
  • [10] Randomized trial of cerebrospinal fluid shunt valve design in pediatric hydrocephalus
    Drake, JM
    Kestle, JRW
    Milner, R
    Cinalli, G
    Boop, F
    Piatt, J
    Haines, S
    Schiff, SJ
    Cochrane, DD
    Steinbok, P
    MacNeil, N
    Haines, S
    Sainte-Rose, C
    Poskitt, K
    Goumnerova, L
    Albright, AL
    Cinalli, G
    Pierre-Kahn, A
    Renier, D
    Zerah, M
    Rutka, J
    Humphreys, R
    Hoffman, H
    Lamberti-Pasculi, M
    Teo, C
    Cherny, B
    Aureli, S
    Vandertop, P
    Broomstra, S
    Chadduck, W
    Donahue, D
    Hall, W
    Parent, A
    Turmel, A
    Myles, T
    Hamilton, M
    Oakes, J
    Mapstone, T
    NEUROSURGERY, 1998, 43 (02) : 294 - 303