Evaluating Screw Stability After Pedicle Screw Fixation With PEEK Rods

被引:10
|
作者
Wu, Jie [1 ,2 ]
Shi, Lei [1 ]
Liu, Daohong [2 ]
Wu, Zhigang [1 ]
Gao, Peng [1 ]
Liu, Wenwen [1 ]
Li, Xiaokang [1 ]
Guo, Zheng [1 ]
机构
[1] Fourth Mil Med Univ, Xijing Hosp, Dept Orthoped, Xian 710032, Shanxi, Peoples R China
[2] Chinese Peoples Liberat Army Gen Hosp, Med Ctr 8, Dept Orthopaed, Beijing, Peoples R China
基金
中国国家自然科学基金; 国家重点研发计划;
关键词
PEEK rod; pedicle screw; screw loosening; cervical vertebra; LUMBAR SPINE FUSION; BIOMECHANICAL EVALUATION; CEMENT LEAKAGE; AUGMENTATION; OSTEOPOROSIS; CONSTRUCT; DEBRIS;
D O I
10.1177/2192568221996692
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design: Animal experiment. Objective: To evaluate whether the use of polyetheretherketone (PEEK) rods for posterior spinal fixation can improve screw stability. Methods: Sheep models of anterior-posterior cervical fusion were used in this study. Six sheep were randomly assigned to the PEEK rod group and titanium alloy group. A total of 8 screws and 2 fixing rods were implanted in each sheep. At 24 weeks postoperatively, a computed tomography (CT) evaluation, pull-out test, micro-CT evaluation and histological evaluation were conducted to evaluate screw stability in the harvested surgical segments. Result: According to the CT evaluation, there were no signs of screw loosening in either group. The pull-out force and energy of the PEEK rod group were significantly higher than those of the titanium alloy rod group. Denser and thicker trabecular bone around the screw was observed in the PEEK rod group according to the micro-CT reconstructed images, and quantitative analysis of the micro-CT data confirmed this finding. In the histological evaluation, more abundant and denser bone trabeculae were also observed in the PEEK rod group. However, there was no significant difference in the bone-screw interface between the 2 groups. Conclusion: Posterior spinal fixation with PEEK rods can increase screw stability by promoting bone growth around the screw but cannot promote bone integration at the bone-screw interface in an animal model study. This finding presents a new idea for clinical practices to reduce screw loosening rate.
引用
收藏
页码:393 / 399
页数:7
相关论文
共 50 条
  • [41] Pedicle screw shift without loosening following instrumented posterior fusion: limitations of pedicle screw fixation
    Miyashita, Tomohiro
    Ataka, Hiromi
    Kato, Kei
    Takaoka, Hiromitsu
    Tanno, Takaaki
    NEUROSURGICAL REVIEW, 2019, 42 (03) : 691 - 698
  • [42] The rate of screw misplacement in segmental pedicle screw fixation in adolescent idiopathic scoliosis
    Abul-Kasim, Kasim
    Ohlin, Acke
    ACTA ORTHOPAEDICA, 2011, 82 (01) : 50 - 55
  • [43] Is unilateral pedicle screw fixation superior than bilateral pedicle screw fixation for lumbar degenerative diseases: a meta-analysis
    Lu, Pei
    Pan, Ting
    Dai, Teng
    Chen, Gang
    Shi, Ke-qin
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2018, 13
  • [44] Is unilateral pedicle screw fixation superior than bilateral pedicle screw fixation for lumbar degenerative diseases: a meta-analysis
    Pei Lu
    Ting Pan
    Teng Dai
    Gang Chen
    Ke-qin Shi
    Journal of Orthopaedic Surgery and Research, 13
  • [45] Comparison between posterolateral fusion with pedicle screw fixation and anterior interbody fusion with pedicle screw fixation in adult spondylolytic spondylolisthesis
    Suk, KS
    Jeon, CH
    Park, MS
    Moon, SH
    Kim, NH
    Lee, HM
    YONSEI MEDICAL JOURNAL, 2001, 42 (03) : 316 - 323
  • [46] Mechanical stability of thoracolumbar pedicle screw fixation - The effect of crosslinks - Point of view
    Esses, SI
    SPINE, 2000, 25 (06) : 36S - 36S
  • [48] Internal fixation (pedicle screw fixation) for fusions of the lumbar spi
    Vaccaro, AR
    Garfin, SR
    SPINE, 1995, 20 (24) : S157 - S165
  • [49] Impact of lumbar pedicle screw positioning on screw stability - A biomechanical investigation
    Grevenstein, David
    Scheyerer, Max J.
    Meyer, Carolin
    Borggrefe, Jan
    Hackl, Michael
    Oikonomidis, Stavros
    Eysel, Peer
    Prescher, Andreas
    Wegmann, Kilian
    CLINICAL BIOMECHANICS, 2020, 74 (74) : 66 - 72
  • [50] Prediction of Screw Loosening After Dynamic Pedicle Screw Fixation With Lumbar Polyetheretherketone Rods Using Magnetic Resonance Imaging-Based Vertebral Bone Quality Score
    Jiang, Guozheng
    Xu, Luchun
    Ma, Yukun
    Guan, Jianbin
    Yang, Yongdong
    Zhong, Wenqing
    Li, Wenhao
    Zhou, Shibo
    Song, JiaWei
    Feng, Ningning
    Qiu, Ziye
    Li, Zeyu
    Zhou, Yishu
    Meng, Letian
    Qu, Yi
    Yu, Xing
    NEUROSPINE, 2024, 21 (02) : 712 - 720