A Self-Calibration of Leaky VNA System for In-Fixture Measurements

被引:0
|
作者
Zbik, Mateusz [1 ,2 ]
Wiatr, Wojciech [2 ]
机构
[1] VIGO PHOTONICS, PL-05850 Ozarow Mazowiecki, Poland
[2] Warsaw Univ Technol, Inst Elect Syst, PL-00665 Warsaw, Poland
关键词
Calibration; Standards; Fixtures; Transmission line measurements; Mathematical models; Measurement uncertainty; Load modeling; De-embedding; fixture; leakage; reciprocal networks; self-calibration; NETWORK-ANALYZER CALIBRATION; 16-TERM ERROR MODEL;
D O I
10.1109/TMTT.2023.3305184
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We propose a new approach to calibrating four-port reciprocal test fixtures affected by leakage. It is based on the well-known 16-term error model that represents a stationary embedding of a two-port device tested. We determine all the terms from scattering matrices measured by a calibrated vector network analyzer (VNA) for just four symmetric standards (patterns), consecutively attached to the test ports. For the calibration, we employ three pairs of shorts, open circuits, and matched loads, and a reciprocal and symmetric transmitting pattern, the scattering parameters of which are unknown. We calculate the unknown parameters and error terms using a self-calibration algorithm and compact closed-form, analytical equations expressed in the scattering parameter notation. According to the abbreviated names of the employed patterns and the model, we call our procedure SOLR16 treating it as an extension of the popular short-open-load-reciprocal (SOLR) approach based on the eight-term error model for the leaky embedding. The SOLR16 method is particularly useful to calibrate test fixtures dedicated to mixed-mode device characterization as, e.g., leaky on-wafer probes. Experimental results of the calibration and in-fixture measurements of a photodetector encapsulated in a transistor outline (TO) package type confirm its efficacy and robustness.
引用
收藏
页码:4649 / 4657
页数:9
相关论文
共 50 条
  • [41] Aspects of body self-calibration
    Lackner, JR
    DiZio, PA
    TRENDS IN COGNITIVE SCIENCES, 2000, 4 (07) : 279 - 288
  • [42] An algorithm of camera self-calibration
    Dong, L
    Lu, W
    APPLICATIONS OF DIGITAL IMAGE PROCESSING XXVII, PTS 1AND 2, 2004, 5558 : 862 - 866
  • [43] Self-calibration of a space robot
    deAngulo, VR
    Torras, C
    IEEE TRANSACTIONS ON NEURAL NETWORKS, 1997, 8 (04): : 951 - 963
  • [44] Self-Calibration of Accelerometer Arrays
    Schopp, Patrick
    Graf, Hagen
    Burgard, Wolfram
    Manoli, Yiannos
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2016, 65 (08) : 1913 - 1925
  • [45] An algorithm of camera self-calibration
    Wang, Nian
    Tang, Jun
    Fan, Yi-Zheng
    Liang, Dong
    2006 8TH INTERNATIONAL CONFERENCE ON SIGNAL PROCESSING, VOLS 1-4, 2006, : 1345 - +
  • [46] A survey of camera self-calibration
    Hemayed, EE
    IEEE CONFERENCE ON ADVANCED VIDEO AND SIGNAL BASED SURVEILLANCE, PROCEEDINGS, 2003, : 351 - 357
  • [47] Self-calibration of sensor networks
    Moses, RL
    Patterson, R
    UNATTENDED GROUND SENSOR TECHNOLOGIES AND APPLICATIONS IV, 2002, 4743 : 108 - 119
  • [48] DYNAMIC STEREO WITH SELF-CALIBRATION
    TIRUMALAI, AP
    SCHUNCK, BG
    JAIN, RC
    IEEE TRANSACTIONS ON PATTERN ANALYSIS AND MACHINE INTELLIGENCE, 1992, 14 (12) : 1184 - 1189
  • [49] Statistical perspectives of self-calibration
    Raugh, MR
    Minor, JM
    METROLOGY, INSPECTION, AND PROCESS CONTROL FOR MICROLITHOGRAPHY X, 1996, 2725 : 114 - 121
  • [50] Self-Calibration for Star Sensors
    Fu, Jingneng
    Lin, Ling
    Li, Qiang
    SENSORS, 2024, 24 (11)