The effect of different posterior inclinations of tibial component on tibiofemoral contact pressures after unicompartmental knee arthroplasty

被引:3
|
作者
Yuan, Bo [1 ]
Mo, Zhongjun [2 ]
Zhang, Kuan [3 ,4 ]
Zhu, Xu [1 ]
Yan, Songhua [3 ,4 ]
Zeng, Jizhou [1 ,3 ]
机构
[1] Capital Med Univ, Beijing Luhe Hosp, Dept Bone & Joint Surg, 82 Xinhua South Rd, Beijing 101149, Peoples R China
[2] Minist Civil Affairs, Natl Res Ctr Rehabil Tech Aids, Beijing Key Lab Rehabil Tech Aids Old Age Disabil, Key Lab Human Mot Anal & Rehabil Technol, Beijing 100176, Peoples R China
[3] Beijing Key Lab Fundamental Res Biomech Clin Appli, 10 Xitoutiao, Beijing 100069, Peoples R China
[4] Capital Med Univ, Sch Biomed Engn, Beijing 100069, Peoples R China
关键词
Unicompartmental knee arthroplasty; Mobile-bearing; Posterior inclination of the tibial component; Finite element analysis; The contact pressures of tibiofemoral joint; SLOPE; VALIDATION; BEHAVIOR; LIGAMENT; MODELS; JOINT;
D O I
10.1186/s13018-023-04222-5
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background Different posterior inclinations of tibial component after unicompartmental knee arthroplasty (UKA) may lead to different biomechanical characteristics of the knee joint. This finite element study was designed to investigate the tibiofemoral contact pressures after UKA with different posterior inclinations of tibial component.Methods Finite element model of a healthy knee joint was constructed, and mobile-bearing (MB) UKA models with 5 different posterior inclinations (3 degrees, 5 degrees, 7 degrees, 9 degrees and 11 degrees) of tibial components were simulated. The maximum contact pressures of tibial plateau cartilage in the lateral compartment and polyethylene insert in the medial compartment were calculated based on the ground reaction force and the angle of the knee flexion obtained by 3D motion capture system.Results The loading ratio of medial and lateral compartments during standing stance (medial 54.49%, lateral 45.51%) and tibial anterior displacement (134 N, 3.89 mm) of healthy knee was basically consistent with previous experimental data. The maximum contact pressures of the medial meniscus and lateral tibial plateau cartilage of the healthy knee during standing stance were 2.14 MPa and 1.57 MPa, respectively. At the static standing phase, the maximum contact pressures of the polyethylene insert decreased from 17.90 to 17.29 Mpa, and the maximum contact pressures of the tibial plateau cartilage in the lateral compartment increased from 0.81 to 0.92 Mpa following an increase in the posterior inclination of the tibial component. At the first peak of ground reaction force, the maximum contact pressures of polyethylene insert increased from 22.37 to 25.16 MPa, and the maximum contact pressures of tibial plateau cartilage in the lateral compartment increased from 3.03 to 3.33 MPa, with the increase in the posterior inclination of the tibial component. At the second peak of ground reaction force, the maximum contact pressures of polyethylene insert decreased from 2.34 to 2.22 MPa with the increase in posterior inclination of tibial component.Conclusion The preoperative and postoperative finite element models of MB UKA were well established. The results showed that the maximum contact pressures of the polyethylene insert did not change significantly with the increase in the posterior inclination of the tibial prosthesis, while the maximum contact pressures of the tibial plateau cartilage of the lateral compartment increased when the posterior inclination of the tibial prosthesis was > 7 degrees. Our results also show that the maximum contact pressures were greater with an excessive inclination angle (11 degrees) of the tibial component, and the pressures of the tibial plateau cartilage in the lateral compartment were more concentrated on the posterior area. This study, therefore, proposes that excessive osteotomy should be avoided.
引用
收藏
页数:8
相关论文
共 50 条
  • [31] The Valgus Inclination of the Tibial Component Increases the Risk of Medial Tibial Condylar Fractures in Unicompartmental Knee Arthroplasty
    Inoue, Shinji
    Akagi, Masao
    Asada, Shigeki
    Mori, Shigeshi
    Zaima, Hironori
    Hashida, Masahiko
    JOURNAL OF ARTHROPLASTY, 2016, 31 (09): : 2025 - 2030
  • [32] The effect of geometry of the tibial polyethylene insert on the tibiofemoral contact kinematics in Advance Medial Pivot total knee arthroplasty
    Omori, Go
    Onda, Naoaki
    Shimura, Masashi
    Hayashi, Toyohiko
    Sato, Takashi
    Koga, Yoshio
    JOURNAL OF ORTHOPAEDIC SCIENCE, 2009, 14 (06) : 754 - 760
  • [33] The effect of posterior tibial slope on range of motion after total knee arthroplasty
    Kansara, Devanshu
    Markel, David C.
    JOURNAL OF ARTHROPLASTY, 2006, 21 (06): : 809 - 813
  • [34] Higher survival rate in total knee arthroplasty after high tibial osteotomy than that after unicompartmental knee arthroplasty
    Lee, Sun-Ho
    Seo, Hyoung-Yeon
    Lim, Jun-Hyuk
    Kim, Min-Gwang
    Seon, Jong-Keun
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2023, 31 (03) : 1132 - 1142
  • [35] Higher survival rate in total knee arthroplasty after high tibial osteotomy than that after unicompartmental knee arthroplasty
    Sun-Ho Lee
    Hyoung-Yeon Seo
    Jun-Hyuk Lim
    Min-Gwang Kim
    Jong-Keun Seon
    Knee Surgery, Sports Traumatology, Arthroscopy, 2023, 31 : 1132 - 1142
  • [36] The effect of unicompartmental knee arthroplasty on lower extremity axis, tibiofemoral subluxation, and lateral knee joint distance
    Aslan, Hakan
    Duman, Evrim
    Atli, Osman Yagiz
    Cevik, Huseyin Bilgehan
    CUKUROVA MEDICAL JOURNAL, 2021, 46 (02): : 699 - 705
  • [37] The Coverage Characteristic of Tibial Component for Unicompartmental Knee Arthroplasty in Chinese Patients - a Morphological Analysis
    Chang, T. W.
    Wei, H. W.
    Chen, W. C.
    Lin, K. J.
    Cheng, C. K.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 4: IMAGE PROCESSING, BIOSIGNAL PROCESSING, MODELLING AND SIMULATION, BIOMECHANICS, 2010, 25 : 677 - 679
  • [38] Tibial component alignment and risk of loosening in unicompartmental knee arthroplasty: a radiographic and radiostereometric study
    Barbadoro, P.
    Ensini, A.
    Leardini, A.
    d'Amato, M.
    Feliciangeli, A.
    Timoncini, A.
    Amadei, F.
    Belvedere, C.
    Giannini, S.
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2014, 22 (12) : 3157 - 3162
  • [39] Tibial component alignment and risk of loosening in unicompartmental knee arthroplasty: a radiographic and radiostereometric study
    P. Barbadoro
    A. Ensini
    A. Leardini
    M. d’Amato
    A. Feliciangeli
    A. Timoncini
    F. Amadei
    C. Belvedere
    S. Giannini
    Knee Surgery, Sports Traumatology, Arthroscopy, 2014, 22 : 3157 - 3162
  • [40] Improved positioning of the tibial component in unicompartmental knee arthroplasty with patient-specific cutting
    Mai Lan Dao Trong
    Diezi, Christian
    Goerres, Gerhard
    Helmy, Nader
    SWISS MEDICAL WEEKLY, 2013, 143 : 4S - 4S