High-pitch, high temporal resolution, multi-energy cardiac imaging on a dual-source photon-counting-detector CT

被引:14
|
作者
Ahmed, Zaki [1 ]
Campeau, David [2 ]
Gong, Hao [1 ]
Rajendran, Kishore [1 ]
Rajiah, Prabhakar [1 ]
McCollough, Cynthia
Leng, Shuai [3 ]
机构
[1] Mayo Clin, Dept Radiol, Rochester, MN USA
[2] Mayo Clin, Dept Physiol & Biomed Engn, Rochester, MN USA
[3] 200 First St SW, Rochester, MN 55905 USA
基金
美国国家卫生研究院;
关键词
cardiac CT; computed tomography; dual energy CT; iodine quantification; material decomposition; motion artifact; photon counting detector; COMPUTED-TOMOGRAPHY; ENERGY CT; ANGIOGRAPHY; RECONSTRUCTION; QUALITY;
D O I
10.1002/mp.16124
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectiveTo measure the accuracy of material decomposition using a dual-source photon-counting-detector (DS-PCD) CT operated in the high-pitch helical scanning mode and compare the results against dual-source energy-integrating-detector (DS-EID) CT, which requires use of a low-pitch value in dual-energy mode. MethodsA DS-PCD CT and a DS-EID CT were used to scan a cardiac motion phantom consisting of a 3-mm diameter iodine cylinder. Iodine maps were reconstructed using DS-PCD in high-pitch mode and DS-EID in low-pitch mode. Image-based circularity, diameter, and iodine concentration of the iodine cylinder were calculated and compared between the two scanners. With institutional review board approval, in vivo exams were performed with the DS-PCD CT in high-pitch mode. Images were qualitatively compared against patients with similar heart rates that were scanned with DS-EID CT in low-pitch dual-energy mode. ResultsOn iodine maps, the mean circularity was 0.97 +/- 0.02 with DS-PCD in high-pitch mode and 0.95 +/- 0.06 with DS-EID in low-pitch mode. The mean diameter was 2.9 +/- 0.2 mm with DS-PCD and 3.1 +/- 0.2 mm with DS-EID, both of which are close to the 3 mm ground truth. For DS-PCD, the mean iodine concentration was 9.6 +/- 0.8 mg/ml and this was consistent with the 9.4 +/- 0.6 mg/ml value obtained with the cardiac motion disabled. For DS-EID, the concentration was 12.7 +/- 1.2 mg/ml with motion enabled and 11.7 +/- 0.5 mg/ml disabled. The background noise in the iodine maps was 15.1 HU with DS-PCD and 14.4 HU with DS-EID, whereas the volume CT dose index (CTDIvol) was 3 mGy with DS-PCD and 11 mGy with DS-EID. On comparison of six patients (three on PCD, three on EID) with similar heart rates, DS-PCD provided iodine maps with well-defined coronaries even at a high heart rate of 86 beats per minute. Meanwhile, there were substantial motion artifacts in iodine maps obtained with DS-EID for patients with similar heart rates. ConclusionIn a cardiac motion phantom, DS-PCD CT can perform accurate material decomposition in high-pitch mode, providing iodine maps with excellent geometric accuracy and robustness to motion at approximately 38% of the dose for similar noise as DS-EID CT.
引用
收藏
页码:1428 / 1435
页数:8
相关论文
共 50 条
  • [41] Assessment of the Coronary Artery in High-Pitch, Dual-Source CT Aortography Without ECG Synchronization
    Choi, Hyeong Gi
    Park, Mi Jung
    Choi, Ho Cheol
    Choi, Hye Young
    Shin, Hwa Seon
    Na, Jae Boem
    Cho, Jae Min
    Choi, Dae Seob
    IRANIAN JOURNAL OF RADIOLOGY, 2017, 14 (03)
  • [42] Diagnostic accuracy of high-pitch dual-source CT for the assessment of coronary stenoses: first experience
    Sebastian Leschka
    Paul Stolzmann
    Lotus Desbiolles
    Stephan Baumueller
    Robert Goetti
    Thomas Schertler
    Hans Scheffel
    Andre Plass
    Volkmar Falk
    Gudrun Feuchtner
    Borut Marincek
    Hatem Alkadhi
    European Radiology, 2009, 19 : 2896 - 2903
  • [43] Diagnostic accuracy of high-pitch dual-source CT for the assessment of coronary stenoses: first experience
    Leschka, Sebastian
    Stolzmann, Paul
    Desbiolles, Lotus
    Baumueller, Stephan
    Goetti, Robert
    Schertler, Thomas
    Scheffel, Hans
    Plass, Andre
    Falk, Volkmar
    Feuchtner, Gudrun
    Marincek, Borut
    Alkadhi, Hatem
    EUROPEAN RADIOLOGY, 2009, 19 (12) : 2896 - 2903
  • [44] Accuracy of Combined Computed Tomography Colonography and Dual Energy Iiodine Map Imaging for Detecting Colorectal masses using High-pitch Dual-source CT
    Sun, Kai
    Han, Ruijuan
    Han, Yang
    Shi, Xuesen
    Hu, Jiang
    Lu, Bin
    SCIENTIFIC REPORTS, 2018, 8
  • [45] Accuracy of Combined Computed Tomography Colonography and Dual Energy Iiodine Map Imaging for Detecting Colorectal masses using High-pitch Dual-source CT
    Kai Sun
    Ruijuan Han
    Yang Han
    Xuesen Shi
    Jiang Hu
    Bin lu
    Scientific Reports, 8
  • [46] Multi-contrast CT imaging using a high energy resolution CdTe detector and a CZT photon-counting detector
    Clements, N.
    Richtsmeier, D.
    Hart, A.
    Bazalova-Carter, M.
    JOURNAL OF INSTRUMENTATION, 2022, 17 (01)
  • [47] Approaches, advantages, and challenges to photon counting detector and multi-energy CT
    Toia, Giuseppe V.
    Mileto, Achille
    Borhani, Amir A.
    Chen, Guang-Hong
    Ren, Liqiang
    Uyeda, Jennifer W.
    Marin, Daniele
    ABDOMINAL RADIOLOGY, 2024, 49 (09) : 3251 - 3260
  • [48] High-pitch dual-source CT angiography of the whole aorta without ECG synchronisation: Initial experience
    Beeres, Martin
    Schell, Boris
    Mastragelopoulos, Aristidis
    Herrmann, Eva
    Kerl, Josef Matthias
    Gruber-Rouh, Tatjana
    Lee, Clara
    Siebenhandl, Petra
    Bodelle, Boris
    Zangos, Stephan
    Vogl, Thomas J.
    Jacobi, Volkmar
    Bauer, Ralf W.
    EUROPEAN RADIOLOGY, 2012, 22 (01) : 129 - 137
  • [49] High-pitch dual-source CT angiography of the whole aorta without ECG synchronisation: Initial experience
    Martin Beeres
    Boris Schell
    Aristidis Mastragelopoulos
    Eva Herrmann
    Josef Matthias Kerl
    Tatjana Gruber-Rouh
    Clara Lee
    Petra Siebenhandl
    Boris Bodelle
    Stephan Zangos
    Thomas J. Vogl
    Volkmar Jacobi
    Ralf W. Bauer
    European Radiology, 2012, 22 : 129 - 137
  • [50] Dual-source chest CT angiography with high temporal resolution and high pitch modes: evaluation of image quality in 140 patients
    Tacelli, Nunzia
    Remy-Jardin, Martine
    Flohr, Thomas
    Faivre, Jean-Baptiste
    Delannoy, Valerie
    Duhamel, Alain
    Remy, Jacques
    EUROPEAN RADIOLOGY, 2010, 20 (05) : 1188 - 1196