Developing a medical device-grade T2 phantom optimized for myocardial T2 mapping by cardiovascular magnetic resonance

被引:2
|
作者
Topriceanu, Constantin-Cristian [1 ,2 ,5 ]
Fornasiero, Massimiliano [3 ]
Seo, Han [4 ]
Webber, Matthew [1 ,2 ,4 ,5 ]
Keenan, Kathryn E. E. [6 ]
Stupic, Karl F. F. [6 ]
Bruehl, Ruediger [7 ]
Ittermann, Bernd [7 ]
Price, Kirsty [8 ]
McGrath, Louise [8 ]
Pang, Wenjie [9 ]
Hughes, Alun D. D. [2 ,5 ]
Nezafat, Reza [10 ]
Kellman, Peter [11 ]
Pierce, Iain [1 ,2 ]
Moon, James C. C. [1 ,2 ]
Captur, Gabriella [2 ,4 ,5 ,12 ]
机构
[1] Barts Hlth NHS Trust, Barts Heart Ctr, ECIA 7BE, London, England
[2] UCL, Inst Cardiovasc Sci, Huntley St, London WC1E 6DD, England
[3] UCL Med Sch, 74 Huntley St, London WC1E 6DE, England
[4] Royal Free London NHS Fdn Trust, Ctr Inherited Heart Muscle Condit, Dept Cardiol, Pond St, London NW3 2QG, England
[5] UCL, Med Res Council Unit Lifelong Hlth & Ageing, 1-19 Torrington Pl, London WC1E 7HB, England
[6] NIST, 325 Broadway, Boulder, CO 80305 USA
[7] Phys Tech Bundesanstalt PTB, Abbestr 2-12, D-10587 Berlin, Germany
[8] UCL Bloomsbury Ctr Clin Phenotyping, London WC1E 6HX, England
[9] Resonance Hlth RH, 141 Burswood Rd, Burswood, WA 6100, Australia
[10] Harvard Med Sch, Beth Israel Deaconess Med Ctr, Dept Med, Cardiovasc Div, 330 Brookline Ave, Boston, MA 02215 USA
[11] NHLBI, NIH, Bethesda, MD 20892 USA
[12] UCL, Inst Cardiovasc Sci, Inherited Heart Muscle Condit, Gower St, London WC1E 6BT, England
基金
英国经济与社会研究理事会; 欧盟地平线“2020”; 英国医学研究理事会;
关键词
T-1; mapping; T-2; Phantom; Quality control; MRI; FIELD; ECHO; SEQUENCES; VALUES; 3T;
D O I
10.1186/s12968-023-00926-z
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
IntroductionA long T-2 relaxation time can reflect oedema, and myocardial inflammation when combined with increased plasma troponin levels. Cardiovascular magnetic resonance (CMR) T-2 mapping therefore has potential to provide a key diagnostic and prognostic biomarkers. However, T-2 varies by scanner, software, and sequence, highlighting the need for standardization and for a quality assurance system for T-2 mapping in CMR.AimTo fabricate and assess a phantom dedicated to the quality assurance of T-2 mapping in CMR.MethodA T-2 mapping phantom was manufactured to contain 9 T-1 and T-2 (T-1|T-2) tubes to mimic clinically relevant native and post-contrast T-2 in myocardium across the health to inflammation spectrum (i.e., 43-74 ms) and across both field strengths (1.5 and 3 T). We evaluated the phantom's structural integrity, B-0 and B-1 uniformity using field maps, and temperature dependence. Baseline reference T-1|T-2 were measured using inversion recovery gradient echo and single-echo spin echo (SE) sequences respectively, both with long repetition times (10 s). Long-term reproducibility of T-1|T-2 was determined by repeated T-1|T-2 mapping of the phantom at baseline and at 12 months.ResultsThe phantom embodies 9 internal agarose-containing T-1|T-2 tubes doped with nickel di-chloride (NiCl2) as the paramagnetic relaxation modifier to cover the clinically relevant spectrum of myocardial T-2. The tubes are surrounded by an agarose-gel matrix which is doped with NiCl2 and packed with high-density polyethylene (HDPE) beads. All tubes at both field strengths, showed measurement errors up to <= 7.2 ms [< 14.7%] for estimated T-2 by balanced steady-state free precession T-2 mapping compared to reference SE T-2 with the exception of the post-contrast tube of ultra-low T-1 where the deviance was up to 16 ms [40.0%]. At 12 months, the phantom remained free of air bubbles, susceptibility, and off-resonance artifacts. The inclusion of HDPE beads effectively flattened the B-0 and B-1 magnetic fields in the imaged slice. Independent temperature dependency experiments over the 13-38 degrees C range confirmed the greater stability of shorter vs longer T-1|T-2 tubes. Excellent long-term (12-month) reproducibility of measured T-1|T-2 was demonstrated across both field strengths (all coefficients of variation < 1.38%).ConclusionThe T-2 mapping phantom demonstrates excellent structural integrity, B-0 and B-1 uniformity, and reproducibility of its internal tube T-1|T-2 out to 1 year. This device may now be mass-produced to support the quality assurance of T-2 mapping in CMR.
引用
收藏
页数:14
相关论文
共 50 条
  • [41] Black blood myocardial T2 mapping
    Wang, Chengyan
    Jang, Jihye
    Neisius, Ulf
    Nezafat, Maryam
    Fahmy, Ahmed
    Kang, Jinkyu
    Rodriguez, Jennifer
    Goddu, Beth
    Pierce, Patrick
    Berg, Sophie
    Zhang, Jue
    Wang, Xiaoying
    Nezafat, Reza
    MAGNETIC RESONANCE IN MEDICINE, 2019, 81 (01) : 153 - 166
  • [42] T2 mapping and T2* imaging in heart failure
    A.S. Lota
    P.D. Gatehouse
    R.H. Mohiaddin
    Heart Failure Reviews, 2017, 22 : 431 - 440
  • [43] Application of quantitative T1, T2 and T2*mapping magnetic resonance imaging in cartilage degeneration of the shoulder joint
    Cao, Guijuan
    Gao, Shubo
    Xiong, Bin
    SCIENTIFIC REPORTS, 2023, 13 (01):
  • [44] Application of quantitative T1, T2 and T2* mapping magnetic resonance imaging in cartilage degeneration of the shoulder joint
    Guijuan Cao
    Shubo Gao
    Bin Xiong
    Scientific Reports, 13 (1)
  • [45] T1 and T2 mapping cardiovascular magnetic resonance imaging techniques reveal unapparent myocardial injury in patients with myocarditis
    Radunski, Ulf K.
    Lund, Gunnar K.
    Saering, Dennis
    Bohnen, Sebastian
    Stehning, Christian
    Schnackenburg, Bernhard
    Avanesov, Maxim
    Tahir, Enver
    Adam, Gerhard
    Blankenberg, Stefan
    Muellerleile, Kai
    CLINICAL RESEARCH IN CARDIOLOGY, 2017, 106 (01) : 10 - 17
  • [46] Tissue characterization by T1 and T2 mapping cardiovascular magnetic resonance imaging to monitor myocardial inflammation in healing myocarditis
    Bohnen, S.
    Radunski, U. K.
    Lund, G. K.
    Ojeda, F.
    Looft, Y.
    Senel, M.
    Radziwolek, L.
    Avanesov, M.
    Tahir, E.
    Stehning, C.
    Schnackenburg, B.
    Adam, G.
    Blankenberg, S.
    Muellerleile, K.
    EUROPEAN HEART JOURNAL-CARDIOVASCULAR IMAGING, 2017, 18 (07) : 744 - 751
  • [47] Left ventricular function recovery in peripartum cardiomyopathy: a cardiovascular magnetic resonance study by myocardial T1 and T2 mapping
    Liang, Yao-Dan
    Xu, Yuan-Wei
    Li, Wei-Hao
    Wan, Ke
    Sun, Jia-Yu
    Lin, Jia-Yi
    Zhang, Qing
    Zhou, Xiao-Yue
    Chen, Yu-Cheng
    JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2020, 22 (01)
  • [48] Left ventricular function recovery in peripartum cardiomyopathy: a cardiovascular magnetic resonance study by myocardial T1 and T2 mapping
    Yao-Dan Liang
    Yuan-Wei Xu
    Wei-Hao Li
    Ke Wan
    Jia-Yu Sun
    Jia-Yi Lin
    Qing Zhang
    Xiao-Yue Zhou
    Yu-Cheng Chen
    Journal of Cardiovascular Magnetic Resonance, 22
  • [49] Parametric methods for characterizing myocardial tissue by magnetic resonance imaging (part 2): T2 mapping
    Perea Palazon, R. J.
    Sole Arques, M.
    Prat Gonzalez, S.
    de Caralt Robira, T. M.
    Cibeira Lopez, M. T.
    Ortiz Perez, J. T.
    RADIOLOGIA, 2015, 57 (06): : 471 - 479
  • [50] Left Ventricular Function Recovery In Peripartum Cardiomyopathy: A Cardiovascular Magnetic Resonance Study By Myocardial T1 And T2 Mapping
    Liang, Yaodan
    Xu, Yuanwei
    Li, Weihao
    Wan, Ke
    Zhang, Qing
    Chen, Yucheng
    CIRCULATION, 2019, 140