Opportunistic AI-enabled automated bone mineral density measurements in lung cancer screening and coronary calcium scoring CT scans are equivalent

被引:6
|
作者
Naghavi, Morteza [1 ]
De Oliveira, Isabel [1 ]
Mao, Song Shou [2 ]
Jaberzadeh, Amirhossein [1 ]
Montoya, Juan [1 ]
Zhang, Chenyu [1 ]
Atlas, Kyle [1 ]
Manubolu, Venkat [2 ]
Montes, Marlon [1 ]
Li, Dong [3 ]
Atlas, Thomas [1 ]
Reeves, Anthony [5 ]
Henschke, Claudia [4 ]
Yankelevitz, David [4 ]
Budoff, Matthew [2 ]
机构
[1] TMC Innovat, HeartLung AI Technol, 2450 Holcomb Blvd, Houston, TX 77021 USA
[2] Harbor UCLA Med Ctr, Lundquist Inst, 1124 W Carson St, Torrance, CA 90502 USA
[3] Emory Univ, 201 Dowman Dr, Atlanta, GA 30322 USA
[4] Mt Sinai, 1176 5th Ave,MC Level, New York, NY 10029 USA
[5] Cornell Univ, Ithaca, NY 14850 USA
基金
美国国家卫生研究院;
关键词
Bone mineral density; Artificial intelligence; Deep learning; Opportunistic; Coronary artery calcium score; Quantitative computed tomography; Osteoporosis; Osteopenia; Cardiovascular Screening; Lung cancer screening; QUANTITATIVE COMPUTED-TOMOGRAPHY; X-RAY ABSORPTIOMETRY; OSTEOPOROSIS; QCT;
D O I
10.1016/j.ejro.2023.100492
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and objectives: We previously reported a novel manual method for measuring bone mineral density (BMD) in coronary artery calcium (CAC) scans and validated our method against Dual X-Ray Absorptiometry (DEXA). Furthermore, we have developed and validated an artificial intelligence (AI) based automated BMD (AutoBMD) measurement as an opportunistic add-on to CAC scans that recently received FDA approval. In this report, we present evidence of equivalency between AutoBMD measurements in cardiac vs lung CT scans. Materials and methods: AI models were trained using 132 cases with 7649 (3 mm) slices for CAC, and 37 cases with 21918 (0.5 mm) slices for lung scans. To validate AutoBMD against manual measurements, we used 6776 cases of BMD measured manually on CAC scans in the Multi-Ethnic Study of Atherosclerosis (MESA). We then used 165 additional cases from Harbor UCLA Lundquist Institute to compare AutoBMD in patients who underwent both cardiac and lung scans on the same day. Results: Mean +/- SD for age was 69 +/- 9.4 years with 52.4% male. AutoBMD in lung and cardiac scans, and manual BMD in cardiac scans were 153.7 +/- 43.9, 155.1 +/- 44.4, and 163.6 +/- 45.3 g/cm3, respectively (p = 0.09). BlandAltman agreement analysis between AutoBMD lung and cardiac scans resulted in 1.37 g/cm3 mean differences. Pearson correlation coefficient between lung and cardiac AutoBMD was R2 = 0.95 (p < 0.0001). Conclusion: Opportunistic BMD measurement using AutoBMD in CAC and lung cancer screening scans is promising and yields similar results. No extra radiation plus the high prevalence of asymptomatic osteoporosis makes AutoBMD an ideal screening tool for osteopenia and osteoporosis in CT scans done for other reasons.
引用
收藏
页数:7
相关论文
共 37 条
  • [31] Automated deep learning-based bone mineral density assessment for opportunistic osteoporosis screening using various CT protocols with multi-vendor scanners
    Park, Heejun
    Kang, Woo Young
    Woo, Ok Hee
    Lee, Jemyoung
    Yang, Zepa
    Oh, Sangseok
    SCIENTIFIC REPORTS, 2024, 14 (01):
  • [32] Reliability analysis of visual ranking of coronary artery calcification on low-dose CT of the thorax for lung cancer screening: comparison with ECG-gated calcium scoring CT
    Yoon Kyung Kim
    Yon Mi Sung
    So Hyun Cho
    Young Nam Park
    Hye-Young Choi
    The International Journal of Cardiovascular Imaging, 2014, 30 : 81 - 87
  • [33] RELIABILITY ANALYSIS OF VISUAL RANKING OF CORONARY ARTERY CALCIFICATION ON LOW-DOSE CT OF THE THORAX FOR LUNG CANCER SCREENING: COMPARISON WITH ECG-GATED CALCIUM SCORING CT
    Sung, Yon Mi
    Kim, Yoon Kyung
    JOURNAL OF THORACIC ONCOLOGY, 2013, 8 : S1267 - S1267
  • [34] Reliability analysis of visual ranking of coronary artery calcification on low-dose CT of the thorax for lung cancer screening: comparison with ECG-gated calcium scoring CT
    Kim, Yoon Kyung
    Sung, Yon Mi
    Cho, So Hyun
    Park, Young Nam
    Choi, Hye-Young
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2014, 30 : 81 - 87
  • [35] ACCURACY OF QUANTITATIVE CT AND L1 HOUNSFIELD UNIT METHOD IN DETECTING LOW BONE MINERAL DENSITY IN A LUNG CANCER SCREENING COHORT
    Sarno, I
    Marshall, H.
    Hopcraft, K.
    RESPIROLOGY, 2019, 24 : 161 - 161
  • [36] AI-CVD: Artificial Intelligence-Enabled Opportunistic Screening of Coronary Artery Calcium Computed Tomography Scans for Predicting CVD Events and All-Cause Mortality: The Multi-Ethnic Study of Atherosclerosis (MESA)
    Naghavi, Morteza
    Reeves, Anthony
    Atlas, Kyle
    Zhang, Chenyu
    Atlas, Thomas
    Henschke, Claudia
    Yankelevitz, David
    Budoff, Matthew
    Fan, Wenjun
    Yu, Ruilin
    Branch, Andrea
    Ma, Ning
    Roy, Sion
    Nasir, Khurram
    Molloi, Sabee
    Fayad, Zahi
    McConnell, Michael
    Kakadiaris, Ioannis
    Zulueta, Javier
    Shah, Prediman
    Williams, Kim
    Abela, George
    Levy, Daniel
    Wong, Nathan
    CIRCULATION, 2024, 150
  • [37] Assessing Thoracic Vertebral Bone Mineral Density (T8-T10) for Osteoporosis Diagnosis During CT Lung Cancer Screening in Older Adults
    Li, Song
    Hu, Nandong
    Wei, Zicheng
    Wang, Jiangchuan
    Wang, Rongzhou
    Gao, Xifa
    Qiu, Yingping
    Chen, Xiao
    INTERNATIONAL JOURNAL OF GENERAL MEDICINE, 2024, 17 : 3403 - 3410