Improved calibration method for displacement transformation coefficient in optical and visual measurements

被引:0
|
作者
Li, Haopeng [1 ]
Qiu, Zurong [1 ]
机构
[1] Tianjin Univ, Sch Precis Instruments & Optoelect Engn, State Key Lab Precis Measuring Technol & Instrumen, Tianjin 300072, Peoples R China
基金
中国国家自然科学基金;
关键词
Displacement transformation coefficient (DTC); Laser and vision-based displacement measuring module (LVDMM); Displacement measurement; Locally weighted scatterplot smoothing (LOWESS); Calibration method;
D O I
10.1063/10.0016714
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
Optical and visual measurement technology is used widely in fields that involve geometric measurements, and among such technology are laser and vision-based displacement measuring modules (LVDMMs). The displacement transformation coefficient (DTC) of an LVDMM changes with the coordinates in the camera image coordinate system during the displacement measuring process, and these changes affect the displacement measurement accuracy of LVDMMs in the full field of view (FFOV). To give LVDMMs higher accuracy in the FFOV and make them adaptable to widely varying measurement demands, a new calibration method is proposed to improve the displacement measurement accuracy of LVDMMs in the FFOV. First, an image coordinate system, a pixel measurement coordinate system, and a displacement measurement coordinate system are established on the laser receiving screen of the LVDMM. In addition, marker spots in the FFOV are selected, and the DTCs at the marker spots are obtained from calibration experiments. Also, a fitting method based on locally weighted scatterplot smoothing (LOWESS) is selected, and with this fitting method the distribution functions of the DTCs in the FFOV are obtained based on the DTCs at the marker spots. Finally, the calibrated distribution functions of the DTCs are applied to the LVDMM, and experiments conducted to verify the displacement measurement accuracies are reported. The results show that the FFOV measurement accuracies for horizontal and vertical displacements are better than +/- 15 mu m and +/- 19 mu m, respectively, and that for oblique displacement is better than +/- 24 mu m. Compared with the traditional calibration method, the displacement measurement error in the FFOV is now 90% smaller. This research on an improved calibration method has certain significance for improving the measurement accuracy of LVDMMs in the FFOV, and it provides a new method and idea for other vision-based fields in which camera parameters must be calibrated.
引用
收藏
页数:14
相关论文
共 50 条
  • [31] An Improved Method For Camera Calibration
    Zeng, Jie-xian
    Fu, Xiang
    Zeng, Jie-xian
    Li, Si-xin
    Tian, Bing
    Yu, De-sheng
    PROCEEDINGS OF THE 2009 2ND INTERNATIONAL CONGRESS ON IMAGE AND SIGNAL PROCESSING, VOLS 1-9, 2009, : 2331 - 2335
  • [32] CALIBRATION OF PHOTODIODE MEASUREMENTS OF CELL MOTION BY A TRANSMISSION OPTICAL-LEVER METHOD
    CLARK, BA
    HALLWORTH, R
    EVANS, BN
    PFLUGERS ARCHIV-EUROPEAN JOURNAL OF PHYSIOLOGY, 1990, 415 (04): : 490 - 493
  • [33] Transmission and division of total optical depth method: A universal calibration method for Sun photometric measurements
    Zhang, Ming
    Gong, Wei
    Ma, Yingying
    Wang, Lunche
    Chen, Zhongyong
    GEOPHYSICAL RESEARCH LETTERS, 2016, 43 (06) : 2974 - 2980
  • [34] A Fiber-Optic Angular Displacement Sensor Based on Optical Microfibers and Its Calibration Method
    Egorov, F. A.
    BULLETIN OF THE LEBEDEV PHYSICS INSTITUTE, 2024, 51 (SUPPL 7) : S572 - S580
  • [35] Optical-interference calibration of inductive displacement sensors
    Yu. D. Bogunenko
    G. A. Bondarenko
    L. V. Edunov
    I. N. Nikitin
    V. A. Starodubtsev
    Instruments and Experimental Techniques, 2000, 43 : 410 - 412
  • [36] Optical-interference calibration of inductive displacement sensors
    Bogunenko, YD
    Bondarenko, GA
    Edunov, LV
    Nikitin, IN
    Starodubtsev, VA
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 2000, 43 (03) : 410 - 412
  • [37] Calibration method for the coefficient of thermal conductivity instruments
    Chen, Ru-Bing
    Wu, Xiang-Dong
    Jiliang Xuebao/Acta Metrologica Sinica, 2007, 28 (SUPPL.): : 221 - 223
  • [38] An improved kinematic calibration method of industrial robots based on double ball bar measurements
    Jiang, Xiaogeng
    Li, Mingxu
    Wang, Hao
    Wang, Yan
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART B-JOURNAL OF ENGINEERING MANUFACTURE, 2025,
  • [39] AN IMPROVED CALIBRATION FOR TETHERED BALLOON WIND MEASUREMENTS
    MORIARTY, WW
    BOUNDARY-LAYER METEOROLOGY, 1993, 63 (1-2) : 183 - 196
  • [40] Iterative mapping for high-precision calibration and displacement measurements
    Liu, Cong
    Wang, Licheng
    Liu, Xiaopeng
    Xu, Zhihong
    Optik, 2021, 248