This study aims to evaluate the effect of pre-reconstruction process for low tube voltage computed tomography (CT) on image quality of filtered back projection (FBP) reconstruction. Small and large quality assurance water phantoms (19 and 33 cm diameter) were scanned on a third-generation dual-source CT with 70 kVp and 120 kVp at various dose levels. Image quality was assessed in terms of the noise power spectrum (NPS) and task-based transfer function (TTF). NPSs and TTFs in the small phantom were comparable between 70 and 120 kVp protocols. In the large phantom, the curves of the NPS changed and the TTF decreased even at the high-dose levels for 70 kVp protocol compared to 120 kVp protocol. Our results indicated that the pre-reconstruction process is performed in low tube voltage CT for large objects even for the FBP reconstruction and has an effect on the image quality.
引用
收藏
页码:306 / 314
页数:9
相关论文
共 50 条
  • [21] Effects of Model-Based Iterative Reconstruction in Low-Dose Paranasal Computed Tomography: A Comparison with Filtered Back Projection and Hybrid Iterative Reconstruction
    Tomita, Hayato
    Kuramochi, Kenji
    Fujikawa, Atsuko
    Ikeda, Hirotaka
    Komita, Midori
    Kurihara, Yoshiko
    Kobayashi, Yasuyuki
    Mimura, Hidefumi
    ACTA MEDICA OKAYAMA, 2022, 76 (05) : 511 - 517
  • [22] Rapid implementation of filtered back-projection and analysis of influencing factors on image reconstruction in optical projection tomography
    State Key Laboratory of Modern Optical Instrumentation, Zhejiang University, Hangzhou 310027, China
    Guangzi Xuebao, 2008, 7 (1488-1491): : 1488 - 1491
  • [23] Computed tomography of the cervical spine: comparison of image quality between a standard-dose and a low-dose protocol using filtered back-projection and iterative reconstruction
    Becce, Fabio
    Ben Salah, Yosr
    Verdun, Francis R.
    Vande Berg, Bruno C.
    Lecouvet, Frederic E.
    Meuli, Reto
    Omoumi, Patrick
    SKELETAL RADIOLOGY, 2013, 42 (07) : 937 - 945
  • [24] Computed tomography of the cervical spine: comparison of image quality between a standard-dose and a low-dose protocol using filtered back-projection and iterative reconstruction
    Fabio Becce
    Yosr Ben Salah
    Francis R. Verdun
    Bruno C. Vande Berg
    Frederic E. Lecouvet
    Reto Meuli
    Patrick Omoumi
    Skeletal Radiology, 2013, 42 : 937 - 945
  • [25] Review of the Effect of Computed Tomography Projection Sampling Strategy on Reconstruction Quality
    Yang F.-Q.
    Zhang D.-H.
    Huang K.-D.
    Gao Z.-Z.
    Liao J.-M.
    Ruan Jian Xue Bao/Journal of Software, 2018, 29 (07): : 2133 - 2151
  • [26] The evolution of image reconstruction for CTfrom filtered back projection to artificial intelligence
    Willemink, Martin J.
    Noeel, Peter B.
    EUROPEAN RADIOLOGY, 2019, 29 (05) : 2185 - 2195
  • [27] Tomographic Image Reconstruction Using Filtered Back Projection (FBP) And Algebraic Reconstruction Technique (ART)
    Chetih, Nabil
    Messali, Zoubeida
    3RD INTERNATIONAL CONFERENCE ON CONTROL, ENGINEERING & INFORMATION TECHNOLOGY (CEIT 2015), 2015,
  • [28] Knowledge-based iterative model reconstruction in coronary computed tomography angiography: comparison with hybrid iterative reconstruction and filtered back projection
    Cha, Min Jae
    Seo, Jae Seung
    Yoo, Dong Soo
    Chong, Semin
    ACTA RADIOLOGICA, 2018, 59 (03) : 280 - 286
  • [29] Improved Image Quality in Dual-Energy Abdominal CT: Comparison of Iterative Reconstruction in Image Space and Filtered Back Projection Reconstruction
    Wang, Rui
    Yu, Wei
    Wu, Runze
    Yang, Hua
    Lu, Dongxu
    Liu, Jiayi
    Zhang, Zhaoqi
    Zhang, Chuanchen
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2012, 199 (02) : 402 - 406
  • [30] Comparison of image quality from filtered back projection, statistical iterative reconstruction, and model-based iterative reconstruction algorithms in abdominal computed tomography (vol 95, pg 31, 2016)
    Kuo, Y.
    Lin, Y-Y
    Lee, R-C
    MEDICINE, 2016, 95 (49)