Accuracy of implant site preparation in robotic navigated dental implant surgery

被引:21
|
作者
Linn, Thu Ya [1 ]
Salamanca, Eisner [1 ]
Aung, Lwin Moe [1 ]
Huang, Ta-Ko [2 ,3 ]
Wu, Yi-Fan [1 ]
Chang, Wei-Jen [1 ,4 ]
机构
[1] Taipei Med Univ, Coll Oral Med, Sch Dent, Taipei, Taiwan
[2] Kaohsiung Med Univ, Sch Dent, Kaohsiung, Taiwan
[3] EPED Inc, Kaohsiung, Taiwan
[4] Taipei Med Univ, Shuang Ho Hosp, Dent Dept, New Taipei City, Taiwan
关键词
digital dentistry; implant drilled socket; implantology; radiographs; surgical; techniques; DYNAMIC NAVIGATION; PLACEMENT; SYSTEM;
D O I
10.1111/cid.13224
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
BackgroundModern technological advancements have led to increase in the development of surgical robots in dentistry, resulting in excellent clinical treatment outcomes. PurposeThis study aimed to determine the accuracy of automatic robotic implant site preparation for different implant sizes by correlating planned and posttreatment positions, and to compare the performance of robotic and human freehand drilling. MethodSeventy-six drilling sites on partially edentulous models were used, with three different implant sizes (o = 3.5 x 10 mm, 4.0 x 10 mm, 5.0 x 10 mm). The robotic procedure was performed using software for calibration and step-by-step drilling processes. After robotic drilling, deviations in the implant position from the planned position were determined. The angulation, depth, and coronal and apical diameters on the sagittal plane of sockets created by human and robotic drilling were measured. ResultsThe deviation of the robotic system was 3.78 degrees +/- 1.97 degrees (angulation), 0.58 +/- 0.36 mm (entry point), and 0.99 +/- 0.56 mm (apical point). Comparison of implant groups showed the largest deviation from the planned position for 5 mm implants. On the sagittal plane, there were no significant differences between robotic and human surgery except for the 5-mm implant angulation, indicating similar quality between human and robotic drilling. Based on standard implant measurements, robotic drilling exhibited comparable performance to freehand human drilling. ConclusionsA robotic surgical system can provide the greatest accuracy and reliability regarding the preoperative plan for small implant diameters. In addition, the accuracy of robotic drilling for anterior implant surgery can also be comparable to that of human drilling.
引用
收藏
页码:881 / 891
页数:11
相关论文
共 50 条
  • [21] DENTAL AND CRANIOMAXILLOFACIAL IMPLANT SURGERY
    Triplett, R. Gilbert
    Berger, Joel
    Jensen, Ole
    Louis, Patrick
    JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2017, 75 (08) : E74 - E93
  • [22] Accuracy of robotic computer-assisted implant surgery for immediate implant placement: A retrospective case series study
    Li, Ping
    Zhao, Chunhui
    Chen, Jiahao
    Xu, Shulan
    Yang, Shuo
    Li, An
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2024, 26 (06) : 1279 - 1288
  • [23] Flapless dental implant surgery enabled by haptic robotic guidance: A case report
    Ali, Mohamed
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2024, 26 (02) : 251 - 257
  • [24] A method to evaluate the accuracy of dental implant placement without postoperative radiography after computer-guided implant surgery: A dental technique
    Son, KeunBaDa
    Huang, Mei-Yang
    Lee, Kyu-Bok
    JOURNAL OF PROSTHETIC DENTISTRY, 2020, 123 (05): : 661 - 666
  • [25] Entirely robotic cochlear implant surgery
    Abari, Jaouad
    Heuninck, Emilie
    Topsakal, Vedat
    AMERICAN JOURNAL OF OTOLARYNGOLOGY, 2024, 45 (05)
  • [26] Accuracy of an autonomous dental implant robotic system in partial edentulism: A pilot clinical study
    Wang, MiaoZhen
    Liu, Feng
    Yu, Tao
    Zhan, Yalin
    Ma, Feifei
    Rausch-Fan, Xiaohui
    CLINICAL ORAL INVESTIGATIONS, 2024, 28 (07)
  • [27] A Robotic System Integrated With CBCT for Cochlear Implant Surgery: Accuracy Improvement and Validation
    Liu, Hengjia
    Liu, Feng
    Yu, Hongjian
    Du, Zhijiang
    IEEE ROBOTICS AND AUTOMATION LETTERS, 2023, 8 (12): : 8010 - 8017
  • [28] Does implant drill design influence the accuracy of dental implant placement using static computer-assisted implant surgery? An in vitro study
    Anna Takács
    Gyula Marada
    Kinga Turzó
    Ákos Nagy
    Orsolya Németh
    Eitan Mijiritsky
    Márton Kivovics
    Attila Mühl
    BMC Oral Health, 23
  • [29] Does implant drill design influence the accuracy of dental implant placement using static computer-assisted implant surgery? An in vitro study
    Takacs, Anna
    Marada, Gyula
    Turzo, Kinga
    Nagy, Akos
    Nemeth, Orsolya
    Mijiritsky, Eitan
    Kivovics, Marton
    Muhl, Attila
    BMC ORAL HEALTH, 2023, 23 (01)
  • [30] Accuracy and Deviation Analysis of Static and Robotic Guided Implant Surgery: A Case Study
    Mozer, Paul S.
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2020, 35 (05) : E86 - E90