Craniospinal Irradiation: A Dosimetric Comparison Between O-Ring Linac and Conventional C-arm Linac

被引:3
|
作者
Stroubinis, Theodoros [1 ]
Psarras, Michalis [1 ]
Zygogianni, Anna [2 ]
Protopapa, Maria [1 ]
Kouloulias, Vassilis [3 ]
Platoni, Kalliopi [4 ]
机构
[1] Mediterraneo Hosp, Dept Radiat Oncol & Stereotact Radiosurg, Athens, Greece
[2] Natl & Kapodistrian Univ Athens, Aretaieion Univ Hosp, Sch Med, Dept Radiol 1,Radiat Oncol Unit, Athens, Greece
[3] Natl & Kapodistrian Univ Athens, Attikon Univ Hosp, Sch Med, Dept Radiol 2,Radiat Oncol Unit, Athens, Greece
[4] Natl & Kapodistrian Univ Athens, Attikon Univ Hosp, Sch Med, Dept Radiol 2,Med Phys Unit, Athens, Greece
关键词
INTENSITY-MODULATED RADIOTHERAPY; GRADIENT JUNCTION TECHNIQUE; RADIATION-THERAPY;
D O I
10.1016/j.adro.2022.101139
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: The aim of this study was to perform a dosimetric evaluation between craniospinal irradiation volumetric modulated arc therapy plans designed for an O-Ring and a conventional C-arm Linac.Methods and Materials: Two adult patients were selected for this study. Two plans were designed one for a TrueBeam Edge and one for Halcyon O-ring Linac for each patient. The evaluation of the plans was conducted in terms of dose volume histogram analysis of the target volume and organs at risk (OARs) along with total plan monitor units and beam-on time. Paired sample t test was performed to compare Dmax and Dmean of OARs for each plan's comparison. The delivery of volumetric modulated arc therapy plans was evaluated using Octavius 4D phantom.Results: All plans demonstrated dose distributions with sufficient planned target volume conformity and homogeneity. The Homogeneity Index and Conformity Index for all plans were found comparable. The paired sample t test did not demonstrate significant difference in terms of Dmax and Dmean of OARs between plans for both patients. All plans met the threshold of 90%, withConclusions: Craniospinal irradiation plans generated for Halcyon and Edge are equivalent in terms of plan quality and dose sparing at OARs. The small variations may have originated from the differences in beam profile or mean energy, the degree of the modulation for each plan and characteristics of multi leaf collimators for each treatment unit. Halcyon is able to deliver a distinctly faster treatment, but Edge provides an automatic rotational correction for patient positioning.& COPY; 2022 The Authors. Published by Elsevier Inc. on behalf of American Society for Radiation Oncology. This is an open access
引用
收藏
页数:10
相关论文
共 50 条
  • [41] Increase in Superficial Dose in Whole-Breast Irradiation With Halcyon Straight-Through Linac Compared With Traditional C-arm Linac With Flattening Filter: In vivo Dosimetry and Planning Study
    O'Grady, Fionnbarr
    Barsky, Andrew R.
    Anamalayil, Shibu
    Freedman, Gary M.
    Kennedy, Christopher
    Cai, Bin
    Laugeman, Eric
    Dong, Lei
    Hugo, Geoffrey D.
    Metz, James M.
    Mutic, Sasa
    Taunk, Neil K.
    Li, Taoran
    ADVANCES IN RADIATION ONCOLOGY, 2020, 5 (01) : 120 - 126
  • [42] Prostate SBRT using O-Ring Halcyon Linac - Plan quality, delivery efficiency, and accuracy
    Pokhrel, Damodar
    Tackett, Tanner
    Stephen, Joseph
    Visak, Justin
    Amin-Zimmerman, Falguni
    McGregor, Andrew
    Strup, Stephen E.
    St Clair, William
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (01): : 68 - 75
  • [43] Surface scanner camera position optimization on the Varian Halcyon™ O-ring gantry linac system
    Delombaerde, L.
    Petillion, S.
    Depuydt, T.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S1122 - S1123
  • [44] Impact of Target Volume on the Quality of Automatically Generated O-Ring linac SBRT Treatment Plans
    Hernandez, Katerine Viviana Diaz
    Unterkirhers, Sergejs
    Schneider, Uwe
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S3559 - S3560
  • [45] Investigating In Vivo Image Quality of Pelvic Patients with a Novel C-Arm Linac CBCT Feature
    Woods, R. C.
    Gregg, K. W.
    Guevara, B.
    Rezaei, A.
    Jia, A.
    Spratt, D. E.
    Robar, J. L.
    Kashani, R.
    Henke, L.
    Price, A. T.
    MEDICAL PHYSICS, 2024, 51 (10) : 7763 - 7763
  • [46] A generic multimodule phantom for testing geometry of a linac c-arm as a part of quality control in radiotherapy
    Tabor, Zbislaw
    Kabat, Damian
    Tulik, Monika
    Kycia, Radoslaw
    Latala, Zbigniew
    MEDICAL PHYSICS, 2017, 44 (10) : 4989 - 5000
  • [47] Comparison of Skin Dose for Brain Cancer Patients Between a 1.5 T MR-Linac and a Conventional Linac
    Kim, A.
    McCann, C.
    Ruschin, M. E.
    Tseng, C. L.
    Sahgal, A.
    Keller, B. M.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2020, 108 (03): : E279 - E279
  • [48] Dosimetric Comparison in Malignant Glioma Patients Clinically Treated on Hybrid Magnetic Resonance Imaging (MRI)-Linac (MRL) vs. Conventional Linac
    Wang, M. H.
    Kim, A.
    Ruschin, M. E.
    Tan, H.
    Soliman, H.
    Myrehaug, S. D.
    Detsky, J.
    Husain, Z. A.
    Atenafu, E.
    Keller, B. M.
    Sahgal, A.
    Tseng, C. L.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2021, 111 (03): : E517 - E518
  • [49] Focal boost dose escalated prostate SBRT on the Halcyon fast-rotating O-ring linac
    De Roover, R.
    Crijns, W.
    Michiels, S.
    Draulans, C.
    Poels, K.
    Haustermans, K.
    Depuydt, T.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S499 - S499
  • [50] Validation and IMRT/VMAT delivery quality of a preconfigured fast-rotating O-ring linac system
    De Roover, Robin
    Crijns, Wouter
    Poels, Kenneth
    Michiels, Steven
    Nulens, An
    Vanstraelen, Bianca
    Petillion, Saskia
    De Brabandere, Marisol
    Haustermans, Karin
    Depuydt, Tom
    MEDICAL PHYSICS, 2019, 46 (01) : 328 - 339