Craniospinal Irradiation: A Dosimetric Comparison Between O-Ring Linac and Conventional C-arm Linac

被引:3
|
作者
Stroubinis, Theodoros [1 ]
Psarras, Michalis [1 ]
Zygogianni, Anna [2 ]
Protopapa, Maria [1 ]
Kouloulias, Vassilis [3 ]
Platoni, Kalliopi [4 ]
机构
[1] Mediterraneo Hosp, Dept Radiat Oncol & Stereotact Radiosurg, Athens, Greece
[2] Natl & Kapodistrian Univ Athens, Aretaieion Univ Hosp, Sch Med, Dept Radiol 1,Radiat Oncol Unit, Athens, Greece
[3] Natl & Kapodistrian Univ Athens, Attikon Univ Hosp, Sch Med, Dept Radiol 2,Radiat Oncol Unit, Athens, Greece
[4] Natl & Kapodistrian Univ Athens, Attikon Univ Hosp, Sch Med, Dept Radiol 2,Med Phys Unit, Athens, Greece
关键词
INTENSITY-MODULATED RADIOTHERAPY; GRADIENT JUNCTION TECHNIQUE; RADIATION-THERAPY;
D O I
10.1016/j.adro.2022.101139
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: The aim of this study was to perform a dosimetric evaluation between craniospinal irradiation volumetric modulated arc therapy plans designed for an O-Ring and a conventional C-arm Linac.Methods and Materials: Two adult patients were selected for this study. Two plans were designed one for a TrueBeam Edge and one for Halcyon O-ring Linac for each patient. The evaluation of the plans was conducted in terms of dose volume histogram analysis of the target volume and organs at risk (OARs) along with total plan monitor units and beam-on time. Paired sample t test was performed to compare Dmax and Dmean of OARs for each plan's comparison. The delivery of volumetric modulated arc therapy plans was evaluated using Octavius 4D phantom.Results: All plans demonstrated dose distributions with sufficient planned target volume conformity and homogeneity. The Homogeneity Index and Conformity Index for all plans were found comparable. The paired sample t test did not demonstrate significant difference in terms of Dmax and Dmean of OARs between plans for both patients. All plans met the threshold of 90%, withConclusions: Craniospinal irradiation plans generated for Halcyon and Edge are equivalent in terms of plan quality and dose sparing at OARs. The small variations may have originated from the differences in beam profile or mean energy, the degree of the modulation for each plan and characteristics of multi leaf collimators for each treatment unit. Halcyon is able to deliver a distinctly faster treatment, but Edge provides an automatic rotational correction for patient positioning.& COPY; 2022 The Authors. Published by Elsevier Inc. on behalf of American Society for Radiation Oncology. This is an open access
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Volumetric Modulated Arc Therapy Planning for Craniospinal Irradiation With a New O-ring Linac
    Matsumoto, Kenji
    Monzen, Hajime
    Kubo, Kazuki
    Otsuka, Masakazu
    Nambu, Hidekazu
    Nishimura, Yasumasa
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2023, 15 (03)
  • [2] Volumetric modulated arc therapy of head-and-neck cancer on a fast-rotating O-ring linac: Plan quality and delivery time comparison with a C-arm linac
    Michiels, Steven
    Poels, Kenneth
    Crijns, Wouter
    Delombaerde, Laurence
    De Roover, Robin
    Vanstraelen, Bianca
    Haustermans, Karin
    Nuyts, Sandra
    Depuydt, Tom
    RADIOTHERAPY AND ONCOLOGY, 2018, 128 (03) : 479 - 484
  • [3] Comparative analysis of fetal dose sparing between a C-arm linac and an O-ring linac in a SIB-VMAT sarcoma treatment for a pregnant patient: A technical note and case report
    Rivais, Wesley
    Constine, Louis
    Pacella, Matthew
    Joyce, Neil
    Nagey, Maimuna
    Webster, Matthew
    Yoon, Jihyung
    Jung, Hyunuk
    Tanny, Sean
    Lemus, Olga Maria Dona
    Zheng, Dandan
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2025, 26 (01):
  • [4] Mechanical and dosimetric accuracy of Dynamic Trajectory Radiotherapy delivery on a C-arm linac
    Bertholet, J.
    Mackeprang, P.
    Loebner, H. L.
    Mueller, S.
    Guyer, G.
    Wyss, Y.
    Frei, D.
    Volken, W.
    Elicin, O.
    Aebersold, D. M.
    Fix, M. K.
    Manser, P.
    RADIOTHERAPY AND ONCOLOGY, 2022, 170 : S1482 - S1483
  • [5] Dosimetric Comparison of O-Ring and C-Arm Linac for Volumetric Modulated Arc Therapy with Simultaneous Integrated Boost of Left-Sided Breast Cancer under Deep Inspiration Breath Hold
    Hui, T.
    Ho, Y. K.
    Ko, C. C.
    Wong, M. K.
    Ngan, Y. T.
    Mok, T. W.
    Chan, L. C.
    Ho, C. H.
    Lok, K. H.
    Chan, S. N.
    Tang, H. L.
    Cheng, H. C.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2024, 120 (02): : E141 - E141
  • [6] SUPINE LINAC TREATMENT VERSUS TOMOTHERAPY IN CRANIOSPINAL IRRADIATION: PLANNING COMPARISON AND DOSIMETRIC EVALUATION
    Hong, J. Y.
    Kim, G. W.
    Kim, C. U.
    Cheon, G. S.
    Son, S. H.
    Lee, J. Y.
    Lee, Y. H.
    Lee, J. H.
    Choi, B. O.
    Kim, Y. S.
    Lee, S. N.
    Jang, H. S.
    Kang, Y. N.
    Yoon, S. C.
    RADIATION PROTECTION DOSIMETRY, 2011, 146 (1-3) : 364 - 366
  • [7] Clinical validation of ring-mounted halcyon linac for lung SBRT: comparison to SBRT-dedicated C-arm linac treatments
    Pokhrel, Damodar
    Visak, Justin
    Critchfield, Lana C.
    Stephen, Joseph
    Bernard, Mark E.
    Randall, Marcus
    Kudrimoti, Mahesh
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (01): : 261 - 270
  • [8] Automated treatment planning of prostate stereotactic body radiotherapy with focal boosting on a fast-rotating O-ring linac: Plan quality comparison with C-arm linacs
    De Roover, Robin
    Crijns, Wouter
    Poels, Kenneth
    Dewit, Bertrand
    Draulans, Cedric
    Haustermans, Karin
    Depuydt, Tom
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (09): : 59 - 72
  • [9] Craniospinal Irradiation: A Dosimetric Analysis between Halcyon and Truebeam Linac Accelerators Using Mlc Models
    Sundaram, V.
    Khanna, D.
    Palanisamy, M.
    Chakravarty, N.
    Arya, R.
    MEDICAL PHYSICS, 2024, 51 (10) : 7827 - 7827
  • [10] Comparison between Midline and Mid-Breast Isocenter Placement for Prone Breast Irradiation Using an O-Ring Jawless Linac
    Apinorasethkul, O.
    Li, T.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2019, 105 (01): : E698 - E698