Insole Pressure Sensors to Assess Post-Stroke Gait

被引:0
|
作者
Nam, Hyung Seok [1 ,2 ,3 ]
Clancy, Caitlin [4 ]
Smuck, Matthew [3 ]
Lansberg, Maarten G. [4 ,5 ]
机构
[1] Sheikh Khalifa Specialty Hosp, Dept Rehabil Med, Ras Al Khaymah, U Arab Emirates
[2] Seoul Natl Univ Hosp, Dept Rehabil Med, Seoul, South Korea
[3] Stanford Univ, Dept Orthopaed Surg, Wearable Hlth Lab, Redwood City, CA 94063 USA
[4] Stanford Univ, Stanford Stroke Ctr, Palo Alto, CA 94304 USA
[5] Stanford Univ, Stanford Stroke Ctr, 780 Welch Rd, Palo Alto, CA 94304 USA
来源
ANNALS OF REHABILITATION MEDICINE-ARM | 2024年 / 48卷 / 01期
关键词
Stroke; Gait analysis; Insole pressure sensor; Outcome measure; PARAMETERS; STROKE; INDIVIDUALS; SYMMETRY; WALKING;
D O I
10.5535/arm.23064
中图分类号
R49 [康复医学];
学科分类号
100215 ;
摘要
Objective: To confirm that the simplified insole does not affect the gait speed and to identify objective sensor-based gait parameters that correlate strongly with existing clinical gait assessment scales. Methods: Ten participants with gait impairment due to hemiplegic stroke were enrolled in this study. Pairs of insoles with four pressure sensors on each side were manufactured and placed in each shoe. Data were extracted during the 10-Meter Walk Test. Several sensor-derived parameters (for example stance time, heel_on-to-toe_peak time, and toe_peak pressure) were calculated and correlated with gait speed and lower extremity Fugl-Meyer (F-M) score. Results: The insole pressure sensor did not affect gait, as indicated by a strong correlation (rho=0.988) and high agreement (ICC=0.924) between the gait speeds with and without the insole. The parameters that correlated most strongly with highest beta coefficients against the clinical measures were stance time of the non-hemiplegic leg (beta=-0.87 with F-M and beta=0.95 with gait speed) and heel_on-to-toe_peak time of the non-hemiplegic leg (beta=-0.86 with F-M and -0.94 with gait speed). Conclusion: Stance time of the non-hemiparetic leg correlates most strongly with clinical measures and can be assessed using a non-obtrusive insole pressure sensor that does not affect gait function. These results suggest that an insole pressure sensor, which is applicable in a home environment, may be useful as a clinical endpoint in post-stroke gait therapy trials.
引用
收藏
页码:42 / 49
页数:8
相关论文
共 50 条
  • [41] Steering of human gait network and changes induced post-stroke
    Starrs, F.
    Frias, I
    Soucy, J-P
    Paquette, C.
    INTERNATIONAL JOURNAL OF STROKE, 2015, 10 : 101 - 101
  • [42] Use of virtual reality in balance and gait training post-stroke
    Pang, Marco Y. C.
    HONG KONG PHYSIOTHERAPY JOURNAL, 2014, 32 (02) : 49 - 49
  • [43] Design of a Biofeedback Device for Gait Rehabilitation in Post-Stroke Patients
    Khoo, I-Hung
    Marayong, Panadda
    Krishnan, Vennila
    Balagtas, Michael Nico
    Rojas, Omar
    2015 IEEE 58TH INTERNATIONAL MIDWEST SYMPOSIUM ON CIRCUITS AND SYSTEMS (MWSCAS), 2015,
  • [44] Effects of hip abduction and adduction accuracy on post-stroke gait
    Dean, Jesse C.
    Embry, Aaron E.
    Stimpson, Katy H.
    Perry, Lindsay A.
    Kautz, Steven A.
    CLINICAL BIOMECHANICS, 2017, 44 : 14 - 20
  • [45] Neuroplasticity in post-stroke gait recovery and noninvasive brain stimulation
    Yi Xu
    Qing-hua Hou
    Shawn D.Russell
    Bradford C.Bennett
    Andrew J.Sellers
    Qiang Lin
    Dong-feng Huang
    Neural Regeneration Research, 2015, 10 (12) : 2072 - 2080
  • [46] Characteristics of limb kinematics in the gait disorders of post-stroke patients
    Naomichi Mizuta
    Naruhito Hasui
    Takumi Kai
    Yasuhiro Inui
    Masahiro Sato
    Sora Ohnishi
    Junji Taguchi
    Tomoki Nakatani
    Scientific Reports, 14
  • [47] Predictive simulation of post-stroke gait with functional electrical stimulation
    Gilmar F. Santos
    Eike Jakubowitz
    Nicolas Pronost
    Thomas Bonis
    Christof Hurschler
    Scientific Reports, 11
  • [48] Anterior-posterior displacement of center of pressure measured by insole foot pressure measurement system in subacute recovery stage of post-stroke hemiplegia
    Choi, Hanboram
    Kim, Woo-Sub
    TECHNOLOGY AND HEALTH CARE, 2018, 26 (04) : 649 - 657
  • [49] Systematic review of automatic post-stroke gait classification systems
    Jiao, Yiran
    Hart, Rylea
    Reading, Stacey
    Zhang, Yanxin
    GAIT & POSTURE, 2024, 109 : 259 - 270
  • [50] Insole Plantar Pressure Measurement During Quiet Stance Post Stroke
    Hillier, Susan
    Lai, Myra S.
    TOPICS IN STROKE REHABILITATION, 2009, 16 (03) : 189 - 195