Evaluation on the Uncertainty Propagation With the Replacement of the Mercury Triple Point of ITS-90

被引:0
|
作者
Zhao, Yikun [1 ,2 ,3 ]
Zhang, Jintao [2 ]
Liang, Yu [2 ]
Feng, Xiaojuan [2 ]
He, Yongdong [1 ]
机构
[1] Xinjiang Univ, Sch Phys Sci & Technol, Urumqi 830017, Peoples R China
[2] Natl Inst Metrol, Div Thermophys Metrol, Beijing 100029, Peoples R China
[3] Xinjiang Uygur Autonomous Reg Res Inst Measurement, Urumqi 830011, Peoples R China
基金
中国国家自然科学基金;
关键词
Carbon dioxide triple point (CO2 TP); international temperature scale; mercury triple point (Hg TP); standard platinum resistance thermometer (SPRT); uncertainty propagation; xenon triple point (Xe TP); XENON;
D O I
10.1109/JSEN.2023.3294260
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The mercury triple point (Hg TP), one of the defining points of the international temperature scale of 1990 (ITS-90), has the distinguishing role for reducing Type 1 and 3 nonuniqueness with the subrange between the argon triple point (Ar TP) and the water triple point (TPW). However, mercury is highly toxic and will be banned from global commercial trade by the action of Minamata Convention on Mercury. The triple points of xenon (Xe TP), carbon dioxide (CO2 TP), and sulfur hexafluoride (SF6 TP) are the candidates for the replacement of the Hg TP. We presented in this study our analytical investigation of the propagation of uncertainties with the fixed points, the propagation of calibration errors for standard platinum resistance thermometers (SPRTs), and Type 3 nonuniqueness. We observed that, given the Hg TP as the reference, the interpolation formula using the single Xe TP for the replacement is capable for the reduction of 57% of the uncertainties of propagation and 48% of the errors of calibration for SPRTs. The formula using the single CO2 TP for the replacement is capable of the reduction of 31% of the uncertainties of propagation and 26% of the errors of calibration for SPRTs. Because of the lack of calibration data, we failed to get an entire insight of the effect on Type 1 and Type 3 nonuniqueness with the replacements of the typical interpolation formulas. Instead, we conducted a limit evaluation on Type 3 nonuniqueness only with three SPRTs calibrated at the CO2 TP. We observed no general effect on Type 3 nonuniqueness for the interpolation formula using the single CO2 TP.
引用
收藏
页码:17951 / 17957
页数:7
相关论文
共 50 条
  • [21] Comparison measurement of mercury triple point cells for uncertainty evaluation
    Sato, K
    Kishimoto, I
    Arai, M
    SICE 2003 ANNUAL CONFERENCE, VOLS 1-3, 2003, : 2598 - 2602
  • [22] Realization at IMGC of the ITS-90 fixed points from the argon triple point upwards
    Marcarino, P
    Steur, PPM
    Dematteis, R
    TEMPERATURE: ITS MEASUREMENT AND CONTROL IN SCIENCE AND INDUSTRY, VOL 7, PTS 1 AND 2, 2003, 684 : 65 - 70
  • [23] A new interpolation equation in the ITS-90 subrange from the triple point of water to the freezing point of indium
    Lan, Kang
    Fang, Lide
    Wang, Ying
    Kang, Zhiru
    Sang, Suli
    METROLOGIA, 2024, 61 (02)
  • [24] Apparatus for realizing the ITS-90 in the temperature range from the triple point of argon to the melting point of gallium
    Ya. E. Razhba
    I. A. Razhba
    Measurement Techniques, 2007, 50 : 1165 - 1173
  • [25] Apparatus for realizing the ITS-90 in the temperature range from the triple point of argon to the melting point of gallium
    Razhba, Ya. E.
    Razhba, I. A.
    MEASUREMENT TECHNIQUES, 2007, 50 (11) : 1165 - 1173
  • [26] Contribution of uncertainties in resistance measurements to uncertainty in ITS-90
    White, DR
    TEMPERATURE: ITS MEASUREMENT AND CONTROL IN SCIENCE AND INDUSTRY, VOL 7, PTS 1 AND 2, 2003, 684 : 321 - 326
  • [27] Realization of the triple point of carbon dioxide evaluated by the ITS-90 (vol 57, 015004, 2020)
    Kawamura, Yasuki
    Matsumoto, Nobuhiro
    Nakano, Tohru
    METROLOGIA, 2021, 58 (02)
  • [28] Uncertainty budgets for realization of ITS-90 by radiation thermometry
    Fischer, J
    Battuello, M
    Sadli, M
    Ballico, M
    Park, SN
    Saunders, P
    Yuan, ZD
    Johnson, BC
    van der Ham, E
    Sakuma, F
    Machin, G
    Fox, N
    Li, W
    Ugur, S
    Matveyev, M
    TEMPERATURE: ITS MEASUREMENT AND CONTROL IN SCIENCE AND INDUSTRY, VOL 7, PTS 1 AND 2, 2003, 684 : 631 - 638
  • [29] Realization of ITS-90 Above the Silver Point at the NIM
    Yuan, Z.
    Lu, X.
    Wang, J.
    Bai, C.
    Dong, W.
    Hao, X.
    Wang, T.
    INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2011, 32 (7-8) : 1611 - 1621
  • [30] The β-γ Transition of Oxygen as a Secondary Fixed Point of the ITS-90
    Szmyrka-Grzebyk, A.
    Kowal, A.
    Lipinski, L.
    Manuszkiewicz, H.
    Steur, P. P. M.
    Pavese, F.
    TEMPERATURE: ITS MEASUREMENT AND CONTROL IN SCIENCE AND INDUSTRY, VOL 8, 2013, 1552 : 204 - 208