Quantitative ultrasound techniques for assessing thermal ablation: Measurement of the backscatter coefficient from ex vivo human liver

被引:2
|
作者
Rohfritsch, Adrien [1 ,3 ]
Franceschini, Emilie [2 ]
Dupre, Aurelien [1 ]
Melodelima, David [1 ]
机构
[1] Univ Lyon, Univ Lyon 1, Ctr Leon Berard, LabTAU,INSERM, Lyon, France
[2] Aix Marseille Univ, CNRS, Cent Marseille LMA, Marseille, France
[3] Univ Lyon, Univ Lyon 1, LabTAU, INSERM,Ctr Leon Berard, F-69003 Lyon, France
关键词
histology; human liver; quantitative ultrasound; INTENSITY FOCUSED ULTRASOUND; ATTENUATION; SCATTERING; BIOPHANTOMS; PARAMETERS; STEATOSIS; TUMORS;
D O I
10.1002/mp.16762
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundUnderstanding the changes occurring in biological tissue during thermal ablation is at the heart of many current challenges in both therapy and medical imaging research.PurposeThe objective of this work is to quantitatively interpret the scattering response of human liver samples, before and after thermal ablation. We report acoustic measurements performed involving n = 21 human liver samples. Thermal ablation is achieved at temperatures between 45 and 80 & DEG;C and quantification of the irreversible changes in acoustic attenuation and Backscattering Coefficient (BSC) is reported, with a particular attention to the latter.MethodsBoth attenuation coefficient and BSCs were measured in the frequency range from 10 to 52 MHz. Scans were performed before heating and after cooling down. Attenuation coefficients were calculated using spectral difference method and BSC estimated using the reference phantom method.ResultsStrong increases of attenuation coefficients and BSCs with heating temperature were observed. Quantitative ultrasonic parameters obtained with the polydisperse structure factor model (poly-SFM)are compared to histological observations and seen to be close to hepatocyte mean diameter (HMD).ConclusionsThe results presented in this study provide a description of the impact of thermal ablation in human liver tissue on acoustic attenuation and the BSC. For the first time, quantitative agreement between the Effective Scatterer Diameter (ESD) estimated from BSC and HMD was shown, highlighting the important role of cellular network in the scattering response of the medium. This core result is an important step toward the determination of the nature of scattering sources in biological tissues.
引用
收藏
页码:6908 / 6919
页数:12
相关论文
共 50 条
  • [21] Measurement of Ex Vivo Liver, Brain and Pancreas Thermal Properties as Function of Temperature
    Mohammadi, Ahad
    Bianchi, Leonardo
    Asadi, Somayeh
    Saccomandi, Paola
    SENSORS, 2021, 21 (12)
  • [22] Comparison of Ultrasound Strain images with Multi-modality Imaging Techniques in Liver RF Ablation Assessment: Initial Ex vivo and Clinical results
    Fernandez, Anna T.
    Kolokythas, Orpheus
    Gauthier, Thomas
    Herzka, Daniel A.
    Patil, Abhay V.
    Xie, Hua
    2008 IEEE ULTRASONICS SYMPOSIUM, VOLS 1-4 AND APPENDIX, 2008, : 313 - +
  • [23] Microwave ablation of ex vivo human liver and colorectal liver metastases with a novel 14.5 GHz generator
    Jones, Robert P.
    Kitteringham, Neil R.
    Terlizzo, Monica
    Hancock, Christopher
    Dunne, Declan
    Fenwick, Stephen W.
    Poston, Graeme J.
    Ghaneh, Paula
    Malik, Hassan Z.
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2012, 28 (01) : 43 - 54
  • [24] Microwave Ablation of Ex Vivo Human Liver and Colorectal Liver Metastases With a Novel 14.5 GHz Generator
    Jones, R.
    Dunne, D.
    Terlizzo, M.
    Fenwick, S.
    Malik, H.
    Poston, G.
    Kitteringham, N.
    Ghaneh, P.
    EUROPEAN JOURNAL OF CANCER, 2011, 47 : S217 - S218
  • [25] Discontinuous moving shot technique for conformal thermal ablation in an ex vivo porcine liver model
    Ma, Ge
    Chen, Hao
    Xu, Jin
    Pan, Hong
    Yu, Muxin
    Wang, Yue
    Xie, Hui
    Zhou, Wenbin
    Wang, Shui
    DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY, 2021, 27 (03) : 418 - 423
  • [26] Image-guided ex vivo liver ablation by unfocused ultrasound using passive cavitation detection
    Salgaonkar, Vasant A.
    Karunakaran, Chandra Priya
    Besse, John A.
    Heinlein, Grace
    Datta, Saurabh
    Holland, Christy K.
    Mast, T. Douglas
    THERMAL TREATMENT OF TISSUE: ENERGY DELIVERY AND ASSESSMENT IV, 2007, 6440
  • [27] Ultrasound interstitial mini invasive probes for thermal ablation in liver: Feasibility study in vivo
    Delabrousse, E.
    Mithieux, F.
    Birer, A.
    Salomir, R.
    Chapelon, J. Y.
    Lafon, C.
    JOURNAL DE RADIOLOGIE, 2007, 88 (11): : 1817 - 1822
  • [28] Dependence of Ultrasound Echo Decorrelation On Tissue Temperature During Radiofrequency Ablation of Ex Vivo Bovine Liver
    Subramanian, S.
    Schmidt, D. T.
    Fosnight, T. R.
    Rao, M. B.
    Mast, T. D.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [29] Potential of Determining Thermal Dose for Ablation Therapies Using Ultrasound Elastography: An Ex Vivo Feasibility Study
    Kling, Ami
    Jiang, Jingfeng
    2018 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2018,
  • [30] Biomechanical testing of ex vivo porcine tendons following high intensity focused ultrasound thermal ablation
    Kwan, William Chu
    Partanen, Ari
    Narayanan, Unni
    Waspe, Adam C.
    Drake, James M.
    PLOS ONE, 2024, 19 (05):