Temperature Rise in Curing Modes of Two Different Dental Light-Curing Units: The Importance of Heating Rate

被引:1
|
作者
Soori, Ahmad [1 ,2 ]
Soori, Faezeh [3 ]
Kowsary, Farshad [1 ]
Kasraei, Shahin [4 ]
机构
[1] Univ Tehran, Coll Engn, Sch Mech Engn, North Kargar St, Tehran, Iran
[2] Islamic Azad Univ Dent Branch Tehran, Tehran, Iran
[3] Hamadan Univ Med Sci, Dent Sch, Hamadan, Iran
[4] Shahid Beheshti Univ Med Sci, Dent Sch, Dept Restorat Dent, Tehran, Iran
关键词
Dentistry; Heating rate; Heat transfer; Infrared thermography; Light cure; INFRARED THERMOGRAPHY; EMITTING DIODE; PULP CHAMBER; POLYMERIZATION; THRESHOLDS; GENERATION; DENTISTRY; INJURY;
D O I
10.1007/s10765-023-03166-6
中图分类号
O414.1 [热力学];
学科分类号
摘要
The heat generated in the process of light curing may cause irreversible damage to the dental pulp. This study aimed to assess heat transfer in the process of irradiation of a dentin disk by using different modes of quartz-tungsten halogen (QTH) and light-emitting diode (LED) dental curing units. The surface temperature was measured by an infrared (IR) thermometer following light curing with standard, ramp, boost, and bleaching modes of a QTH curing unit, and high, soft, and pulse modes of a LED curing unit. Since the maximum temperature rise alone is not sufficient to study the thermal damage to the pulp tissue, we also assessed the temperature rise profile and heating rate, in addition to the maximum temperature rise in this study. The results showed that decreasing the duration of light curing effectively decreased the risk of pulpal damage. Shorter radiation modes, such as the 10 s standard mode and the 10 s boost curing mode of the QTH curing unit, and the 5 s high mode and the 10 s pulse mode of the LED curing unit, all had the lowest risk of thermal damage to the pulp.
引用
收藏
页数:17
相关论文
共 50 条
  • [31] Evaluation of different light-curing lamps
    Ferrante, Maurizio
    Dottore, Paolo
    Petrini, Morena
    Trentini, Paolo
    Spoto, Giuseppe
    JOURNAL OF THERMAL ANALYSIS AND CALORIMETRY, 2012, 109 (02) : 939 - 944
  • [32] The influence of different light-curing modes on microleakage of posterior resin composites
    Marghalani, H.Y. (hmarghalani@kau.edu.sa), 1600, Taylor and Francis Ltd. (28):
  • [33] The influence of different light-curing modes on microleakage of posterior resin composites
    Marghalani, Hanadi Yousif
    JOURNAL OF ADHESION SCIENCE AND TECHNOLOGY, 2014, 28 (02) : 136 - 150
  • [34] Assessing dental light-curing units' output using radiometers: A narrative review
    Assaf, Cendrella
    Fahd, Jean-Claude
    Sabbagh, Joseph
    JOURNAL OF INTERNATIONAL SOCIETY OF PREVENTIVE AND COMMUNITY DENTISTRY, 2020, 10 (01): : 1 - 8
  • [35] Comparison of curing depth of a colored polyacid-modified composite resin with different light-curing units
    Vandenbulcke, Jeroen D. E.
    Marks, Luc A. M.
    Martens, Luc C.
    Verbeeck, Ronald M. H.
    QUINTESSENCE INTERNATIONAL, 2010, 41 (09): : 787 - 794
  • [36] Pulp Temperature Rise Induced by Light-Emitting Diode Light-Curing Units Using an Ex Vivo Model
    Vinagre, Alexandra
    Ramos, Joao C.
    Rebelo, Clara
    Basto, Jose Francisco
    Messias, Ana
    Alberto, Nelia
    Nogueira, Rogerio
    MATERIALS, 2019, 12 (03)
  • [37] Evaluation of light-curing units in rural and urban areas
    AlShaafi, Maan M.
    SAUDI DENTAL JOURNAL, 2012, 24 (3-4) : 163 - 167
  • [38] A SURVEY OF THE EFFECTIVENESS OF LIGHT-CURING UNITS AND TESTING DEVICES
    DAVIES, BR
    DUNNE, SM
    MILLAR, BJ
    JOURNAL OF DENTAL RESEARCH, 1995, 74 (03) : 896 - 896
  • [39] Shedding light on the problem: Proficiency and maintenance practices of light-curing units among dental assistants
    Ismail, Eman
    Al-Zain, Afnan
    Alghaith, Lamya
    Alsenan, Dana
    Ageel, Fatin
    JOURNAL OF ESTHETIC AND RESTORATIVE DENTISTRY, 2024, 36 (04) : 680 - 689
  • [40] Influence of light intensity from different curing units upon composite temperature rise
    Knezevic, A
    Tarle, Z
    Meniga, A
    Sutalo, J
    Pichler, G
    JOURNAL OF ORAL REHABILITATION, 2005, 32 (05) : 362 - 367