Comparison of Radiation Dose and Image Quality of Pediatric High-Resolution Chest CT Between Photon-Counting Detector CT and Energy-Integrated Detector CT: A Matched Study

被引:8
|
作者
Siegel, Marilyn J. [1 ]
Bugenhagen, Scott M. [1 ]
Sanchez, Adrian [1 ]
Kim, Stacy [1 ]
Abadia, Andres [1 ]
Ramirez-Giraldo, Juan C. [1 ]
机构
[1] Washington Univ, Mallinckrodt Inst Radiol, Sch Med, 510 S Kingshighway Blvd, St Louis, MO 63110 USA
关键词
CT; high-resolution CT; pediatrics; photon-counting CT; radiation dose; COMPUTED-TOMOGRAPHY; 100; KVP; REDUCTION; AGREEMENT; SELECTION;
D O I
10.2214/AJR.23.29077
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BACKGROUND. Photon-counting detector (PCD) CT has been shown to reduce radiation dose and improve image quality in adult chest CT examinations; its potential impact in pediatric CT is not well documented. OBJECTIVE. The purpose of our study was to compare radiation dose, objective image quality, and subjective image quality of PCD CT and energy-integrating detector (EID) CT in children undergoing high-resolution CT (HRCT) of the chest. METHODS. This retrospective study included 27 children (median age, 3.9 years; 10 girls, 17 boys) who underwent PCD CT between March 1, 2022, and August 31, 2022, and 27 children (median age, 4.0 years; 13 girls, 14 boys) who underwent EID CT between August 1, 2021, and January 31, 2022; all examinations comprised clinically indicated chest HRCT. The patients in the two groups were matched by age and water-equivalent diameter. Radiation dose parameters were recorded. One observer placed ROIs to measure objective parameters (lung attenuation, image noise, and SNR). Two radiologists independently assessed subjective measures (overall image quality and motion artifacts) using 5-point Likert scales (1 = highest quality). Groups were compared. RESULTS. PCD CT, in comparison with EID CT, showed lower median CTDI vol (0.41 vs 0.71 mGy, p < .001), DLP (10.2 vs 13.7 mGy x cm, p = .008), size-specific dose estimate (0.82 vs 1.34 mGy, p < .001), and tube current-exposure time product (48.0 vs 202.0 mAs, p < .001). PCD CT and EID CT showed no significant difference in right upper lobe (RUL) lung attenuation (mean, -793 vs -750 HU; p = .09), right lower lobe (RLL) lung attenuation (mean, -745 vs -716 HU; p = .23), RUL image noise (mean, 55 vs 51 HU; p = .27), RLL image noise (mean, 59 vs 57 HU; p = .48), RUL SNR (mean, -14.9 vs -15.8; p = .89), or RLL SNR (mean, -13.1 vs -13.6; p = .79). PCD CT and EID CT showed no significant difference in median overall image quality for reader 1 (1.0 vs 1.0, p = .28) or reader 2 (1.0 vs 1.0, p = .17) or median motion artifacts for reader 1 (1.0 vs 1.0, p = .07) or reader 2 (1.0 vs 1.0, p = .22). CONCLUSION. PCD CT showed significantly reduced dose levels without a significant difference in objective or subjective image quality compared with EID CT.
引用
收藏
页码:363 / 371
页数:9
相关论文
共 50 条
  • [1] Pediatric contrast-enhanced chest CT on a photon-counting detector CT: radiation dose and image quality compared to energy-integrated detector CT
    El-Ali, Alexander M.
    Strubel, Naomi
    Pinkney, Lynne
    Xue, Christine
    Dane, Bari
    Lala, Shailee V.
    PEDIATRIC RADIOLOGY, 2024, 54 (12) : 1984 - 1995
  • [2] Comparison of Image Quality and Radiation Dose in Pediatric Temporal Bone CT Using Photon-Counting Detector CT and Energy-Integrating Detector CT
    Lee, Jeong Sub
    Kim, John
    Bapuraj, Jayapalli R.
    Srinivasan, Ashok
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2024, 45 (09) : 1322 - 1326
  • [3] High-Resolution Head CTA: A Prospective Patient Study Comparing Image Quality of Photon-Counting Detector CT and Energy-Integrating Detector CT
    Diehn, Felix E.
    Zhou, Zhongxing
    Thorne, Jamison E.
    Campeau, Norbert G.
    Nagelschneider, Alex A.
    Eckel, Laurence J.
    Benson, John C.
    Madhavan, Ajay A.
    Bathla, Girish
    Lehman, Vance T.
    Huber, Nathan R.
    Baffour, Francis
    Fletcher, Joel G.
    McCollough, Cynthia H.
    Yu, Lifeng
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2024, 45 (10) : 1441 - 1449
  • [4] A new era of high-resolution CT diagnostics of the lung: improved image quality, detailed morphology, and reduced radiation dose with high-resolution photon-counting CT of the lungs compared to high-resolution energy-integrated CT
    Aurumskjold, Marie-Louise
    Sjunnesson, Lotta
    Pistea, Adrian
    Asbjornsson, Gylfi
    Wellman, Fredrik
    Bozovic, Gracijela
    ACTA RADIOLOGICA, 2024, 65 (10) : 1211 - 1221
  • [5] Low-dose high-resolution chest CT in adults with cystic fibrosis: intraindividual comparison between photon-counting and energy-integrating detector CT
    Frings, Marko
    Welsner, Matthias
    Mousa, Christin
    Zensen, Sebastian
    Salhoefer, Luca
    Meetschen, Mathias
    Beck, Nikolas
    Bos, Denise
    Westhoelter, Dirk
    Wienker, Johannes
    Taube, Christian
    Umutlu, Lale
    Schaarschmidt, Benedikt M.
    Forsting, Michael
    Haubold, Johannes
    Sutharsan, Sivagurunathan
    Opitz, Marcel
    EUROPEAN RADIOLOGY EXPERIMENTAL, 2024, 8 (01)
  • [6] Optimizing Chest CT With Photon-counting Detector Technology: Comparison to Established Energy-integrating Detector CT Protocols
    Alarab, N.
    Guo, J.
    Atha, J.
    Sieren, J. C.
    Puliyakote, A. S. Kizhakke
    Hoffman, E. A.
    Abadi, E.
    Fain, S. B.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2024, 209
  • [7] Photon- Counting Detector CT Versus Energy- Integrating Detector CT of the Lumbar Spine: Comparison of Radiation Dose and Image Quality
    Marth, Adrian A.
    Marcus, Roy P.
    Feuerriegel, Georg C.
    Nanz, Daniel
    Sutter, Reto
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2024, 222 (01)
  • [8] Pediatric Applications of Photon-Counting Detector CT
    Cao, Joseph
    Bache, Steve
    Schwartz, Fides R.
    Frush, Donald
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2023, 220 (04) : 580 - 589
  • [9] Contrast-Enhanced Abdominal CT with Clinical Photon-Counting Detector CT: Assessment of Image Quality and Comparison with Energy-Integrating Detector CT
    Higashigaito, Kai
    Euler, Andre
    Eberhard, Matthias
    Flohr, Thomas G.
    Schmidt, Bernhard
    Alkadhi, Hatem
    ACADEMIC RADIOLOGY, 2022, 29 (05) : 689 - 697
  • [10] Low-dose CT of the abdomen: Initial experience on a novel photon-counting detector CT and comparison with energy-integrating detector CT
    Decker, Josua A.
    Bette, Stefanie
    Lubina, Nora
    Rippel, Katharina
    Braun, Franziska
    Risch, Franka
    Wollny, Claudia
    Scheurig-Muenkler, Christian
    Kroencke, Thomas J.
    Schwarz, Florian