An electrochemical biosensor for the determination of hormone Human Chorionic Gonadotropin (hCG) in human serum

被引:2
|
作者
Koterwa, Adrian [1 ]
Bojko, Magdalena [2 ]
Ryl, Jacek [3 ]
Lukaszuk, Krzysztof [4 ,5 ]
Kozlowska, Kornelia [1 ]
Sieklicki, Wiktor [6 ]
Rodziewicz-Motowidlo, Sylwia [2 ]
Niedzialkowski, Pawel [1 ]
机构
[1] Univ Gdansk, Fac Chem, Dept Analyt Chem, Wita Stwosza 63, PL-80308 Gdansk, Poland
[2] Univ Gdansk, Fac Chem, Dept Biomed Chem, Gdansk, Poland
[3] Gdansk Univ Technol, Inst Nanotechnol & Mat Engn, Adv Mat Ctr, Gdansk, Poland
[4] INVICTA Res & Dev Ctr, Fertil & Reprod Ctr, Gdansk, Poland
[5] Med Univ Gdansk, Fac Hlth Sci, Dept Obstet & Gynecol Nursing, Gdansk, Poland
[6] Gdansk Univ Technol, Fac Mech Engn, Gdansk, Poland
关键词
electrochemical detection; gold electrode functionalization; human Chorionic Gonadotropin (hCG); human serum; SILVER NANOPARTICLES; GOLD NANOPARTICLES; IONIC LIQUID; CLINICAL-USE; IMMUNOSENSOR; NANOCOMPOSITE; ELECTRODES; GRAPHENE; SENSITIVITY; CHITOSAN;
D O I
10.1002/elan.202300095
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This work describes the modification of a gold electrode to create an electrochemical biosensor capable of detecting human chorionic gonadotropin (hCG). The biosensor was obtained by modifying the gold electrode with cysteamine and oligopeptide (PPLRINRHILTR). The modification steps of the gold electrode were confirmed by cyclic voltammetry (CV) and impedance electrochemical spectroscopy (EIS) measurements. The conducted EIS experiments in 0.01 M PBS, pH 7.4 confirm that the biosensor exhibits sensitivity towards hCG in a range of concentrations from 1x10(-12) to 1x10(-7) M (0.5 mIU/mL - 50 000 mIU/mL) to solutions with a detection limit of 1. 91x10(-14) M (0.0095 mIU/mL). The effectiveness of the investigated biosensor was also investigated in human serum. The EIS comparative investigations were performed in human serum containing a concentration of 1x10(-12) M (0.5 mIU/mL) hCG and in human serum where the hCG was added. The obtained results indicate that the investigated biosensor is selective for the presence of hCG hormone in the human serum.