Initial end-to-end testing of the ExacTrac dynamic deep inspiration breath hold workflow using a breath hold breast phantom

被引:1
|
作者
Goodall, Simon K. [1 ,2 ]
Rampant, Peter L. [2 ]
机构
[1] Univ Western Australia, Fac Engn & Math Sci, Sch Phys Math & Comp, Crawley, WA 6009, Australia
[2] GenesisCare, 24 Salvado Rd, Wembley, WA 6014, Australia
关键词
Deep inspiration breath hold; ExacTrac Dynamic; Surface guided Radiotherapy; Stereoscopic breast imagining; ACCURACY; GUIDANCE; RADIOTHERAPY;
D O I
10.1007/s13246-023-01291-y
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
ExacTrac Dynamic (ETD) provides a Deep Inspiration Breath Hold (DIBH) workflow for breast patients. Stereoscopic x-ray imaging combined with optical and thermal mapping allows localisation against simulation imaging, alongside surface guided breath hold monitoring. This work aimed to determine appropriate imaging parameters, the optimal Hounsfield Unit (HU) threshold for patient contour generation and workflow evaluation via end-to-end (E2E) positioning using a custom breast DIBH phantom. After localisation via existing Image Guidance (IG), stereoscopic imaging was performed with a range of parameters to determine best agreement. Similarly, residual errors in prepositioning were minimised using a range of HU threshold contours. E2E positioning was completed for clinical workflows allowing residual isocentre position error measurement and existing IG comparison. Parameters of 60 kV and 25mAs were determined appropriate for patient imaging and HU thresholds between -600 HU and -200 HU enabled adequate prepositioning. The average and standard deviation in residual isocentre position error was 1.0 & PLUSMN; 0.9 mm, 0.4 & PLUSMN; 1.0 mm and 0.1 & PLUSMN; 0.5 mm in the lateral, longitudinal and vertical directions, respectively. Errors measured using existing IG were -0.6 & PLUSMN; 1.1 mm, 0.5 & PLUSMN; 0.7 mm and 0.2 & PLUSMN; 0.4 mm in the lateral, longitudinal and vertical directions, and 0.0 & PLUSMN; 1.0(o), 0.5 & PLUSMN; 1.7(o) and -0.8 & PLUSMN; 1.8(o) for pitch roll and yaw. The use of bone weighted matching increased residual error, while simulated reduction of DIBH volume maintained isocentre positioning accuracy despite anatomical changes. This initial testing indicated suitability for clinical implementation during DIBH breast treatments.
引用
收藏
页码:1239 / 1247
页数:9
相关论文
共 50 条
  • [41] A review of factors contributing to failure of deep inspiration breath hold in left breast radiotherapy
    Doull, N.
    Stevens, R.
    Aitken, K.
    Sinclair, F.
    Armstrong, A.
    CLINICAL ONCOLOGY, 2025, 38
  • [42] Deep inspiration breath hold in prone photon or proton irradiation of breast and lymph nodes
    Speleers, B.
    Schoepen, M.
    Belosi, F.
    Vakaet, V.
    De Neve, W.
    Deseyne, P.
    Paelinck, L.
    Vercauteren, T.
    Parkes, M. J.
    Lomax, T.
    Van Greveling, A.
    Bolsi, A.
    Weber, D. C.
    Veldeman, L.
    De Gersem, W.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S922 - S923
  • [43] Heart Sparing Radiotherapy Techniques in Breast Cancer: A Focus on Deep Inspiration Breath Hold
    Stowe, Hayley B.
    Andruska, Neal D.
    Reynoso, Francisco
    Thomas, Maria
    Bergom, Carmen
    BREAST CANCER-TARGETS AND THERAPY, 2022, 14 : 175 - 186
  • [44] Investigating optimal modality for boost treatment of left breast with deep inspiration breath hold
    Sen, A.
    Michalski, A.
    Done, B.
    Windsor, A.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S102 - S102
  • [45] DIBH (Deep Inspiration Breath-Hold) in left breast radiotherapy -institutional experience
    Dzhugashvili, M.
    Montijano, M.
    Sevilla, B.
    Peraza, C.
    Fernandez, J.
    Gonsalves, D.
    Guijarro, M.
    Lopez, E.
    RADIOTHERAPY AND ONCOLOGY, 2023, 182 : S1818 - S1818
  • [46] Targeting accuracy of a deep inspiration breath-hold technique for locoregional breast radiotherapy
    Wells, D.
    Wai, E.
    Bendorffe, B.
    Bardutz, M.
    Rose, A.
    Norn, D.
    Peterson, R.
    McDonald, R.
    Cunningham, H.
    Sayers, D.
    Runkel, J.
    Cooper, S.
    Popescu, C.
    Beckham, W.
    Olivotto, I.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S15 - S15
  • [47] Visual coach to improve treatment of deep inspiration breath hold for left breast patients
    Zucca, Aurora
    Lambri, Nicola
    Galdieri, Carmela
    Dei, Damiano
    Franceschini, Davide
    Spoto, Ruggero
    Ilieva, Mariya
    Bertolini, Anna
    Dominici, Luca
    Franzese, Ciro
    Mancosu, Pietro
    Scorsetti, Marta
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S5607 - S5609
  • [48] Assessment of voluntary deep inspiration breath-hold with CINE imaging for breast radiotherapy
    Estoesta, Reuben Patrick
    Attwood, Lani
    Naehrig, Diana
    Claridge-Mackonis, Elizabeth
    Odgers, David
    Martin, Darren
    Pham, Melissa
    Toohey, Joanne
    Carroll, Susan
    JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2017, 61 (05) : 689 - 694
  • [49] Respiratory Traces Analysis in Left-Breast Irradiation with Deep Inspiration Breath Hold
    Langsen, M.
    Jermoumi, M.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [50] Patient compliance during Deep Inspiration Breath Hold for left breast cancer radiotherapy
    Marson, M.
    Ferretti, A.
    Angela, C.
    Vendrame, A.
    Cristina, C.
    Chiovatti, P.
    Vinante, L.
    RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S1067 - S1068