Initial end-to-end testing of the ExacTrac dynamic deep inspiration breath hold workflow using a breath hold breast phantom

被引:1
|
作者
Goodall, Simon K. [1 ,2 ]
Rampant, Peter L. [2 ]
机构
[1] Univ Western Australia, Fac Engn & Math Sci, Sch Phys Math & Comp, Crawley, WA 6009, Australia
[2] GenesisCare, 24 Salvado Rd, Wembley, WA 6014, Australia
关键词
Deep inspiration breath hold; ExacTrac Dynamic; Surface guided Radiotherapy; Stereoscopic breast imagining; ACCURACY; GUIDANCE; RADIOTHERAPY;
D O I
10.1007/s13246-023-01291-y
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
ExacTrac Dynamic (ETD) provides a Deep Inspiration Breath Hold (DIBH) workflow for breast patients. Stereoscopic x-ray imaging combined with optical and thermal mapping allows localisation against simulation imaging, alongside surface guided breath hold monitoring. This work aimed to determine appropriate imaging parameters, the optimal Hounsfield Unit (HU) threshold for patient contour generation and workflow evaluation via end-to-end (E2E) positioning using a custom breast DIBH phantom. After localisation via existing Image Guidance (IG), stereoscopic imaging was performed with a range of parameters to determine best agreement. Similarly, residual errors in prepositioning were minimised using a range of HU threshold contours. E2E positioning was completed for clinical workflows allowing residual isocentre position error measurement and existing IG comparison. Parameters of 60 kV and 25mAs were determined appropriate for patient imaging and HU thresholds between -600 HU and -200 HU enabled adequate prepositioning. The average and standard deviation in residual isocentre position error was 1.0 & PLUSMN; 0.9 mm, 0.4 & PLUSMN; 1.0 mm and 0.1 & PLUSMN; 0.5 mm in the lateral, longitudinal and vertical directions, respectively. Errors measured using existing IG were -0.6 & PLUSMN; 1.1 mm, 0.5 & PLUSMN; 0.7 mm and 0.2 & PLUSMN; 0.4 mm in the lateral, longitudinal and vertical directions, and 0.0 & PLUSMN; 1.0(o), 0.5 & PLUSMN; 1.7(o) and -0.8 & PLUSMN; 1.8(o) for pitch roll and yaw. The use of bone weighted matching increased residual error, while simulated reduction of DIBH volume maintained isocentre positioning accuracy despite anatomical changes. This initial testing indicated suitability for clinical implementation during DIBH breast treatments.
引用
收藏
页码:1239 / 1247
页数:9
相关论文
共 50 条
  • [1] Initial end-to-end testing of the ExacTrac dynamic deep inspiration breath hold workflow using a breath hold breast phantom
    Simon K. Goodall
    Peter L. Rampant
    Physical and Engineering Sciences in Medicine, 2023, 46 : 1239 - 1247
  • [2] Radiotherapy in breast cancer with voluntary deep-inspiration breath-hold using BrainLab Exactrac
    Ippolito, E.
    D'Angelillo, R.
    Sicilia, A.
    Silipigni, S.
    Floreno, B.
    Molfese, E.
    Di Donato, A.
    Trecca, P.
    Gaudino, D.
    Stimato, G.
    Ramella, S.
    Trodella, L.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S817 - S817
  • [3] How Important Is a Reproducible Breath Hold for Deep Inspiration Breath Hold Breast Radiation Therapy?
    Wiant, David
    Wentworth, Stacy
    Liu, Han
    Sintay, Benjamin
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 93 (04): : 901 - 907
  • [4] Comparison of Deep Inspiration Breath Hold Left Breast Tangents With Active Breath Hold and Voluntary Breath Hold Using 3D Surface Image Guidance
    Zhao, B.
    Rahimi, A. S.
    Yan, Y.
    Spangler, A.
    Jiang, S. B.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S840 - S840
  • [5] A Simple Deep Inspiration Breath Hold System
    Rasmussen, B.
    Duelge, J.
    Baldwin, B.
    Valenti, M.
    Hunsader, A.
    Kaznowski, L.
    Blackburn, J.
    Chu, K.
    MEDICAL PHYSICS, 2014, 41 (06) : 92 - 92
  • [6] Heart Shift with Deep Inspiration Breath Hold
    Gergel, T. J.
    Veale, C. J.
    Jones, A. O.
    Treas, J.
    Vognetz, J. A.
    Christie, A. D.
    Yumen, O.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (02): : S246 - S247
  • [7] Deep inspiration breath-hold technique using an Arduino
    Gallego, P.
    Perez-Alija, J.
    Olivares, S.
    Loscos, S.
    Ambroa, E.
    Pedro, A.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S811 - S811
  • [8] First experience on deep inspiration breath hold radiotherapy in breast cancer
    Duma, M. -N
    Winkler, C.
    Berndt, J.
    Sill, E.
    Dodig, Pusic K.
    Luettich, K.
    Beierl, S.
    Molls, M.
    Oechsner, M.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2014, 190 : 26 - 26
  • [9] Deep inspiration breath hold with RapidArc in bilateral breast adjuvant radiotherapy
    Cima, S.
    Frapolli, M.
    Nicolini, G.
    Yordanov, K.
    Pesce, G.
    Azinwi, C.
    Martucci, F.
    Stelmes, J. J.
    Richetti, A.
    Valli, M. C.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2015, 191 (11) : 900 - 900
  • [10] Prone Positioning With Deep Inspiration Breath Hold for Left Breast Radiotherapy
    Goyal, Uma
    Saboda, Kathylynn
    Roe, Denise
    Gonzalez, Victor J.
    CLINICAL BREAST CANCER, 2021, 21 (04) : E295 - E301